SeaSpine Spacer System- Hollywood NanoMetalene, Hollywood VI NanoMetalene, Ventura NanoMetalene; SeaSpine Cambria NanoMetalene
K162715 · SeaSpine Orthopedics Corporation · MAX · Oct 28, 2016 · Orthopedic
Device Facts
Record ID
K162715
Device Name
SeaSpine Spacer System- Hollywood NanoMetalene, Hollywood VI NanoMetalene, Ventura NanoMetalene; SeaSpine Cambria NanoMetalene
Applicant
SeaSpine Orthopedics Corporation
Product Code
MAX · Orthopedic
Decision Date
Oct 28, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
SeaSpine® Spacer System - Hollywood NanoMetalene®, Hollywood VI NanoMetalene®, and Ventura NanoMetalene®: When used as an intervertebral body fusion device, the system is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of non-operative treatment. The device is intended to be used with autogenous bone graft (autograft) and supplemental fixation. Cambria NanoMetalene®: Cambria is intended to be used as an adjunct to spinal fusion procedures at one level (C3-C7) in skeletally mature patients with degenerative disc disease (defined as neck pain with discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Patients should have received at least six weeks of non-operative treatment prior to treatment with the device. Devices are intended to be implanted via an open, anterior approach and used with autogenous bone and supplemental fixation, such as an anterior plating system.
Device Story
Intervertebral body fusion devices; PEEK-OPTIMA core with 1-micron CP titanium surface coating; tantalum or Ti-6Al-4V ELI markers for radiographic visualization. Central canal holds autogenous bone graft to promote fusion. Implanted by surgeons via open approach; used with supplemental fixation (e.g., anterior plating). Provides structural support to disc space; facilitates spinal fusion in DDD patients. Offered in various geometries/sizes to match patient anatomy. Gamma sterilized; provided in sterile packaging.
Clinical Evidence
No clinical data. Substantial equivalence based on non-clinical bench testing, including packaging, shipping, and sterilization validation (SAL 10^-6) and bacterial endotoxin testing per ANSI/AAMI ST-72:2011.
Technological Characteristics
Materials: PEEK-OPTIMA (ASTM F2026), CP Titanium surface (ASTM F67), Tantalum (ASTM F560) or Ti-6Al-4V ELI (ASTM F136) markers. Design: Interbody spacer with central graft channel. Sterilization: Gamma radiation. Connectivity: None.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) requiring spinal fusion. Lumbar/lumbosacral (L2-S1) use for Hollywood/Ventura systems (1-2 levels, DDD with/without Grade 1 spondylolisthesis/retrolisthesis, prior non-fusion surgery allowed, 6 months non-operative treatment). Cervical (C3-C7) use for Cambria system (1 level, 6 weeks non-operative treatment). Contraindications: none stated.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
SeaSpine® Spacer System (Hollywood NanoMetalene®) (K102026)
SeaSpine® Spacer System (Hollywood VI NanoMetalene®, Ventura NanoMetalene®, and Cambria NanoMetalene®) (K142488)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 28, 2016
SeaSpine® Orthopedics Corporation % Ms. Gina Flores Regulatory Specialist 5770 Armada Drive Carlsbad, California 92008
Re: K162715
Trade/Device Name: SeaSpine® Spacer System - Hollywood NanoMetalene®, Hollywood VI NanoMetalene®, Ventura NanoMetalene®; SeaSpine® Cambria NanoMetalene® Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX, ODP Dated: September 28, 2016 Received: September 29, 2016
Dear Ms. Flores:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K162715
#### Device Name
SeaSpine® Spacer System - Hollywood NanoMetalene®, and Ventura NanoMetalene®
Indications for Use (Describe)
SeaSpine® Spacer System - Hollywood NanoMetalene®, Hollywood VI NanoMetalene®, and Ventura NanoMetalene®:
When used as an intervertebral body fusion device, the system is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of non-operative treatment. The device is intended to be used with autogenous bone graft (autograft) and supplemental fixation.
Type of Use (*Select one or both, as applicable*)
| <input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162715
Device Name SeaSpine® Cambria NanoMetalene®
#### Indications for Use (Describe)
Cambria is intended to be used as an adjunct to spinal fusion procedures at one level (C3-C7) in skeletally mature patients with degenerative disc disease (defined as neck pain with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Patients should have received at least six weeks of non-operative treatment prior to treatment with the device. Devices are intended to be implanted via an open, and used with autogenous bone and supplemental fixation, such as an anterior plating system.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size:16px">✖</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|--------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
K162715 Page 1 of 4
#
# Contact Details
| Applicant Name: | SeaSpine® Orthopedics Corporation |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Address:<br>Phone number:<br>Fax number: | 5770 Armada Drive, Carlsbad CA<br>(760) 216-5136<br>(760) 683-6874 |
| Contact person:<br>Email address: | Gina Flores, Regulatory Specialist<br>gina.flores@SeaSpine.com |
| Date Prepared: | October 27, 2016 |
| Device Name | |
| Trade Name: | SeaSpine® Spacer System - Hollywood NanoMetalene®, Hollywood VI NanoMetalene®, Ventura NanoMetalene®<br>SeaSpine® Cambria NanoMetalene® |
| Common Name: | Intervertebral Body Fusion Device |
| Classification Name: | Intervertebral body fusion device<br>(21 CFR 888.3080) |
| Class: | II |
| Product Code: | MAX, ODP |
# Legally Marketed Predicate Devices
| 510(k) Number | Product Code | Trade Name | Manufacturer |
|--------------------------|--------------|----------------------------------------------------------------------------------------------------------------|--------------------------------------|
| Primary Predicate Device | | | |
| K102026 | MAX, MQP | SeaSpine® Spacer System-<br>(Hollywood NanoMetalene®) | SeaSpine® Orthopedics<br>Corporation |
| Predicate Device | | | |
| K142488 | MAX, ODP | SeaSpine® Spacer System<br>(Hollywood VI NanoMetalene®,<br>Ventura NanoMetalene®)<br>and Cambria NanoMetalene® | SeaSpine® Orthopedics<br>Corporation |
{5}------------------------------------------------
# Device Description
The SeaSpine® Spacer System (Hollywood NanoMetalene®, Hollywood VI NanoMetalene®, Ventura NanoMetalene®), and the Cambria NanoMetalene® are intervertebral fusion devices manufactured from polyetheretherketone (PEEK OPTIMA LT1 per ASTM F2026) with markers (tantalum per ASTM F560 or Ti-6AI-4V ELI per ASTM F136) for radiographic visualization. The implants have a one-micron thick surface coat of commercially pure (CP) titanium. The devices have a central canal for receiving autogenous bone graft and are offered in a variety of sizes and geometries to accommodate variations in pathology and patient anatomy. The purpose of this 510(k) is to receive clearance to offer the interbody devices of the noted systems gamma sterilized in individual packaging.
# Intended Use/Indications for use
The NanoMetalene® subject devices have substantially equivalent indications and intended use as the cited predicates:
# SeaSpine® Spacer System (Hollywood NanoMetalene®, Hollywood VI NanoMetalene®, Ventura NanoMetalene®):
When used as an intervertebral body fusion device, the system is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of non-operative treatment. The device is intended to be used with autogenous bone graft (autograft) and supplemental fixation.
# Cambria NanoMetalene®:
Cambria is intended to be used as an adjunct to spinal fusion procedures at one level (C3-C7) in skeletally mature patients with degenerative disc disease (defined as neck pain with discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Patients should have received at least six weeks of non-operative treatment prior to treatment with the device. Devices are intended to be implanted via an open, anterior approach and used with autogenous bone and supplemental fixation, such as an anterior plating system.
# Summary of Technological Characteristics
The SeaSpine® Spacer System (Hollywood NanoMetalene®, Hollywood VI NanoMetalene®, Ventura NanoMetalene®), and the Cambria NanoMetalene® devices are substantially equivalent to the cited predicate devices in areas including intended use/indications for use, technological characteristics (operating principle, design, materials, sterility, manufacturing, etc.) and performance (mechanical safety).
{6}------------------------------------------------
| | Subject Devices: | Predicate Devices: |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| | Hollywood NanoMetalene® System | Hollywood NanoMetalene® System |
| | Hollywood VI NanoMetalene® System | Hollywood VI NanoMetalene® System |
| | Ventura NanoMetalene® System | Ventura NanoMetalene® System |
| | Cambria NanoMetalene® System | Cambria NanoMetalene® System |
| Intended Use | Adjunct to spinal fusion procedures in<br>skeletally mature patients with<br>degenerative disc disease (DDD) | Adjunct to spinal fusion procedures in<br>skeletally mature patients with<br>degenerative disc disease (DDD) |
| Materials | PEEK-OPTIMA (ASTM 2026),<br>Titanium alloy (ASTM F136),<br>Tantalum, (ASTM F560),<br>CP Titanium Surface (ASTM F67) | PEEK-OPTIMA (ASTM 2026),<br>Titanium alloy (ASTM F136),<br>Tantalum, (ASTM F560),<br>CP Titanium Surface (ASTM F67) |
| Design | Varying footprints and heights with a<br>central channel to be filled with<br>autogenous bone graft to allow for<br>bone fusion. | Varying footprints and heights with a<br>central channel to be filled with<br>autogenous bone graft to allow for<br>bone fusion. |
| Performance<br>Testing | No new testing required | Analysis and static & dynamic<br>testing (ASTM F2027),<br>Wear Evaluation (ASTM F1877),<br>Subsidence (ASTM F2267) |
| Packaging | Implants will be double packaged in<br>PETG trays with Tyvek lids.<br><br>Components, accessories and<br>instruments are provided trays/caddies<br>intended for sterilization by the end<br>user | N/A - System implants, components,<br>accessories and instruments are<br>provided in trays/caddies intended for<br>sterilization by the end user |
| Sterilization | Implants will be provided in sterile<br>packaging validated to ensure a<br>SAL 10-6<br><br>Components, accessories and<br>instruments are sterilized by the end<br>user | System implants, components,<br>accessories and instruments are<br>sterilized by the end user |
## Non-Clinical Testing
Packaging, shipping and sterilization tests were performed to validate a Sterility Assurance Level (SAL) of 10° and ensure maintenance of a sterile barrier. Bacterial Endotoxin testing (BET) was conducted in accordance with ANSI/AAMI ST-72:2011.
## Clinical Testing
Not applicable; determination of substantial equivalence is not based on an assessment of clinical performance data.
{7}------------------------------------------------
## Conclusions
data demonstrate that the SeaSpine® Spacer System (Hollywood The submitted NanoMetalene®, Hollywood VI NanoMetalene®, Ventura NanoMetalene®), and Cambria NanoMetalene® devices are as safe, as effective, and perform at least as safely and effectively as the cited legally marketed predicates.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.