K152539 · Spine View, Inc. · MAX · Feb 24, 2016 · Orthopedic
Device Facts
Record ID
K152539
Device Name
SpineView X-Pac Expandable Lumbar Cage System
Applicant
Spine View, Inc.
Product Code
MAX · Orthopedic
Decision Date
Feb 24, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The SpineView X-Pac Expandable Lumbar Cage System is indicated for use as an intervertebral body fusion device in skeletally mature patients. The system is designed for use with autogenous bone graft material to facilitate spinal fusion. The system is intended for use in patients with degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine (L2-S1). Patients should have at least 6 months of non-operative treatment prior to surgery. In addition, these patients may have up to Grade 1 spondylolisthesis at the involved level(s). These implants are used to facilitate fusion in the lumbar spine and are placed via transforaminal approach using autogenous bone graft material. The system is intended for use with supplemental fixation systems cleared for use by the FDA for use in the thoracolumbar spine.
Device Story
SpineView X-Pac Expandable Lumbar Cage System is an intervertebral body fusion device; provides structural stability following discectomy. Implant features expandable design to accommodate patient anatomy and restore lordosis; superior/inferior surface protrusions grip vertebral endplates to resist expulsion. Used in lumbar spine (L2-S1) via transforaminal approach; requires autogenous bone graft and supplemental fixation. Operated by surgeons in clinical/OR settings. Benefits patients by facilitating spinal fusion in degenerative disc disease cases.
Clinical Evidence
Bench testing only. Mechanical testing conducted per Class II Special Controls Guidance Document, ASTM F2077, and ASTM F2267, including static and dynamic compression, static and dynamic compression-shear, and subsidence testing.
Technological Characteristics
Expandable lumbar interbody cage; manufactured from medical grade titanium alloy (ASTM F136, ASTM F1295). Features surface protrusions for endplate fixation. Non-sterile; requires steam sterilization. Mechanical performance validated per ASTM F2077 and ASTM F2267.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (discogenic back pain confirmed by history/radiography) at one or two contiguous lumbar levels (L2-S1), including up to Grade 1 spondylolisthesis or retrolisthesis. Requires 6 months of prior non-operative treatment. Used with autogenous bone graft via transforaminal approach and supplemental fixation.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
K201145 — X-Pac Expandable Lumbar Cage System · Expanding Innovations, Inc. · Sep 10, 2020
K203802 — X-Pac Expandable Lumbar Cage System · Expanding Innovations, Inc. · Jun 7, 2021
K223174 — X-Pac Expandable LLIF Cage System · Expanding Innovations, Inc. · Feb 14, 2023
K153105 — MLX - Medial Lateral Expandable Lumbar Interbody System · Nu Vasive, Incorporated · Jul 11, 2016
K190398 — Republic Spine Restore Intervertebral Body Fusion System · Republic Spine, LLC · Sep 12, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an image of three human profiles facing to the right, with the profiles overlapping each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 24, 2016
Spine View, Incorporated % Ms. Sevrina Ciucci Regulatory Affairs Consultant Lince Consulting 8 Crow Canyon Court, #205 San Ramon, California 94583
Re: K152539
Trade/Device Name: SpineView X-Pac Expandable Lumbar Cage System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: January 22, 2016 Received: January 27, 2016
Dear Ms. Ciucci:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K152539
Device Name
SpineView X-Pac Expandable Lumbar Cage System
Indications for Use (Describe)
The SpineView X-Pac Expandable Lumbar Cage System is indicated for use as an intervertebral body fusion device in skeletally mature patients. The system is designed for use with autogenous bone graft material to facilitate spinal fusion.
The system is intended for use in patients with degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine (L2-S1). Patients should have at least 6 months of non-operative treatment prior to surgery. In addition, these patients may have up to Grade 1 spondylolisthesis at the involved level(s). These implants are used to facilitate fusion in the lumbar spine and are placed via transforaminal approach using autogenous bone graft material. The system is intended for use with supplemental fixation systems cleared for use by the FDA for use in the thoracolumbar spine.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### A. Name, Address, Phone, and Fax Number of Applicant
Spine View. Inc. 48810 Kato Road, Suite 100E Fremont, CA 94538 Phone: (510) 743-5090 Fax: (510) 490-1753
- Contact Person B. Sevrina Ciucci Requlatory Affairs Consultant
#### C. Date Prepared
August 31, 2015
#### D. Device Name and Classification
| Trade Name: | SpineView X-Pac Expandable Lumbar Cage System |
|----------------------|-----------------------------------------------|
| Common Name: | Intervertebral Body Fusion Device |
| CFR Classification: | 21 CFR§888.3080 |
| Classification Name: | Intervertebral Body Fusion Device |
| Product Code: | MAX |
#### E. Predicate Device
The SpineView X-Pac Expandable Lumbar Cage System is substantially equivalent to the Globus Medical, Inc., Caliber Spacers, K123231.
#### F. Device Description
The SpineView X-Pac Expandable Lumbar Cage System is a lumbar interbody fusion system comprised of a caqe implant and instruments. The implants are used to provide structural stability in skeletally mature individuals following discectomy and are placed via the transforaminal approach. The device is available in various sizes to accommodate varying anatomy and can expand to the desired lordosis. The implants are designed for use with autogenous bone graft material. Protrusions on the superior and inferior surfaces of each implant grip the endplates of the adjacent vertebrae to resist expulsion. The implants are manufactured from medical grade titanium alloy per ASTM F136 and ASTM F1295.
The SpineView X-Pac Expandable Lumbar Cage System implants are singleuse, provided non-sterile, and are intended to be cleaned and steam sterilized
{4}------------------------------------------------
before use. The instruments are re-usable, provided non-sterile, and are intended to be cleaned and steam sterilized before use.
#### G. Indications for Use / Intended Use
The SpineView X-Pac Expandable Lumbar Cage System is indicated for use as an intervertebral body fusion device in skeletally mature patients. The system is designed for use with autogenous bone graft material to facilitate spinal fusion.
The system is intended for use in patients with degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine (L2-S1). Patients should have at least 6 months of non-operative treatment prior to surgery. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These implants are used to facilitate fusion in the lumbar spine and are placed via transforaminal approach using autogenous bone graft material. The system is intended for use with supplemental fixation systems cleared for use by the FDA for use in the thoracolumbar spine.
#### H. Technological Comparison
The technological characteristics of the SpineView X-Pac Expandable Lumbar Cage System implants are similar to the predicate in terms of design, dimensions, intended use, materials, and performance characteristics.
#### l. Performance Data
Mechanical testing consisting of static and dynamic compression, static and dynamic compression-shear, and subsidence was conducted in accordance with "Class II Special Controls Guidance Document: Intervertebral Fusion Device", June 12, 2007, ASTM F2077, and ASTM F2267 to demonstrate substantial equivalence to the predicate巨盆�] ˘ |• ã } Æ· ¡ Å æ ¡ å [ å ` & ˆ å æ å å æ æ animal usability study.
#### J. Basis for Substantial Equivalence
The SpineView X-Pac Expandable Lumbar Cage System is similar to the predicate with respect to technical characteristics, performance and intended use. The information provided supports the substantial equivalence of the X-Pac Expandable Lumbar Cage System to the legally marketed predicate.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.