K151785 · Innovasis, Inc. · MAX · Oct 14, 2015 · Orthopedic
Device Facts
Record ID
K151785
Device Name
Px HA PEEK IBF System
Applicant
Innovasis, Inc.
Product Code
MAX · Orthopedic
Decision Date
Oct 14, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Innovasis Px HA PEEK IBF System is an intervertebral body fusion device intended to stabilize a spinal segment to promote fusion using bone graft, in order to restrict motion and decrease pain. Users of these products are limited to physicians trained in orthopedic surgery. Clinical locations include hospitals and surgery sites equipped to perform spinal surgery. The Innovasis Px HA™ PEEK IBF System is an intervertebral body fusion device for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbar spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to a Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These implants are used to facilitate fusion in the lumbar spine and are placed via either a posterior (PLIF) or modified transforaminal (T-PLIF) approach. This device is intended to be used in pairs and with internal supplemental spinal fixation systems such as the Innovasis Excella® Spinal System. The interior of the Px HA implant is intended to be packed with autograft.
Device Story
Intervertebral body fusion device; used in pairs via posterior (PLIF) or modified transforaminal (T-PLIF) approach; stabilizes spinal segments to promote fusion. Implants feature open interior cavity for autograft packing; anti-migration teeth engage vertebral endplates; tapered leading edge aids insertion. Radiolucent PEEK-OPTIMA HA Enhanced material allows fusion assessment; tantalum spheres provide radiographic visualization. Used in hospitals/surgery centers by orthopedic surgeons; requires supplemental internal spinal fixation (e.g., Innovasis Excella). Benefits include motion restriction and pain reduction in DDD patients.
Clinical Evidence
Bench testing only. Performance evaluated per ASTM F2077-11 and F2267-04 standards, including static axial compression, dynamic axial compression, torsion, subsidence, and expulsion testing. Testing included simulated aging of devices prior to dynamic mechanical evaluation.
Technological Characteristics
Implants machined from Invibio PEEK-OPTIMA HA Enhanced (PEEK with integrated hydroxyapatite). Radiographic markers are unalloyed Tantalum (ASTM F560). Complies with ASTM F2026 (PEEK), ASTM F2077 (fusion device testing), and ASTM F2267 (subsidence). Sterilization via moist heat (ISO 17665-1). Form factor includes tapered leading edge, anti-migration teeth, and open cavity for bone graft.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous lumbar levels (L2-S1), including up to Grade 1 spondylolisthesis or retrolisthesis, who have failed at least six months of non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
K190684 — LxHA PEEK Lateral IBF System · Innovasis, Inc. · Jun 17, 2019
K160074 — Rampart D Lumbar Interbody Fusion Device · Spineology, Inc. · Oct 18, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are simple and abstract, and they are the most prominent element of the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 14, 2015
Innovasis. Incorporated Mr. Marshall McCarty Director QA/RA 614 East 3900 South Salt Lake City, Utah 84107
Re: K151785
Trade/Device Name: Px HATM PEEK IBF System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: September 1, 2015 Received: September 2, 2015
Dear Mr. McCarty:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2 - Mr. Marshall McCarty
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Innovasis. The logo is black and features the word "INNOVASIS" in all caps. Below the word is the phrase "INNOVATE / INVOLVE / INVENT" in a smaller font. The logo is simple and modern.
510(k)
June 26, 2015
K151785 Page 1 of 1
#### 4.0 Indications for Use Statement
510(k) Number: K151785
Device Name: Px HATM PEEK IBF System
INDICATIONS FOR USE are as follows:
## Indications for Use:
The Innovasis Px HA™ PEEK IBF System is an intervertebral body fusion device for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbar spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to a Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These implants are used to facilitate fusion in the lumbar spine and are placed via either a posterior (PLIF) or modified transforaminal (T-PLIF) approach.
This device is intended to be used in pairs and with internal supplemental spinal fixation systems such as the Innovasis Excella® Spinal System. The interior of the Px HA implant is intended to be packed with autograft.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart D)
OR Over-The-Counter-Use (21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Innovasis. The logo is black and features the word "INNOVASIS" in all caps. Below the word is the phrase "INNOVATE / INVOLVE / INVENT" in a smaller font. The letter "V" in Innovasis is stylized to look like a check mark.
#### 510(k) Summary Report: 5.0
# Px HA™ PEEK IBF System
Company: Innovasis, Inc. 614 E. 3900 South Salt Lake City, UT 84107
- Contact: Marshall C. McCarty Phone: (801) 261-2236 mmccarty@innovasis.com
- Px HA™ PEEK IBF System Trade Name:
- Common Name: Intervertebral fusion device with bone graft
| Classification: | Regulation No.: 21CFR 888.3080 | | |
|-----------------|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | Class 2 | | |
| | Product Code: MAX | | |
| | Review Panel: Orthopedic ASDB | | |
| | Applicable Standards: | | |
| | • ASTM F560-13 | Standard Specification for Unalloyed Tantalum for<br>Surgical Implant Applications | |
| | • ASTM F983-86 | (Reapproved 2013) Standard Practice for<br>Permanent Marking of Orthopedic Implant<br>Components | |
| | • ASTM F2026-14 | Standard Specification for Polyetheretherketone<br>(PEEK) Polymers for Surgical Implant<br>Applications | |
| | • ASTM F2077-11 | Test Methods for Intervertebral Body Fusion<br>Devices | |
| | • ASTM F2267-04 | (Reapproved 2011) Standard Test Method for<br>Measuring Load Induced Subsidence of<br>Intervertebral Body Fusion Device Under Static<br>Axial Compression | |
| | • ISO 17665-1:2006 | (Reapproved 2013) Sterilization of Healthcare<br>Products - Moist Heat – Part 1 Requirements for<br>the Development, Validation and Routine Control<br>of a Sterilization Process for Medical Devices | |
- K150500 Innovasis Px™ PEEK IBF System Predicate Device: This predicate has not been subject to a design-related recall.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Innovasis. The logo is black and features the word "INNOVASIS" in a bold, sans-serif font. Below the word "INNOVASIS" is the text "INNOVATE / INVOLVE / INVENT" in a smaller, sans-serif font. The logo is simple and modern, and the text below the logo suggests that the company is focused on innovation and creativity.
| Reference Devices: | P960025<br>K073177 | Jaguar IF / Brantigan Cage<br>Pioneer Bullet Tip IBF |
|--------------------|--------------------|------------------------------------------------------|
|--------------------|--------------------|------------------------------------------------------|
Device Description: The Innovasis Px HA™ PEEK IBF is designed for use in a posterior (PLIF) approach to the lumbar spine. Implants are manufactured by Innovasis from Invibio® PEEK-OPTIMA® HA Enhanced*. Hydroxyapatite (HA) is fully integrated into the PEEK-OPTIMA. The device is radiolucent allowing straightforward assessment of the fusion process, while tantalum spheres are located around the periphery of the device to allow implant visualization during and after surgery.
> The single use implant devices feature an open cavity in the interior geometry to accommodate bone graft and maximize bone in-growth, with anti-migration teeth to engage the vertebral endplates and prevent expulsion. The implants have a slightly convex profile and are offered in a variety of different sizes to fit the anatomical needs of a wide variety of patients. The implant has a tapered leading edge which aids in implant insertion due to limited anatomical space. Reusable instruments to support the PLIF surgery are provided with the implants in custom sterilization trays.
Performance Data: (Non-clinical)-Performance testing per ASTM F2077-11 and F2267-04 for Static Axial Compression, Dynamic Axial Compression, Subsidence and Expulsion indicates that the Px HA PEEK IBF is capable of performing in accordance with its intended use. Testing included simulated aging performed on Px HA PEEK IBF Devices, which then were subjected to additional dynamic compression and torsion tests in accordance with ASTM 2077.
- Materials: The implants are machined from Invibio PEEK-OPTIMA® HA Enhanced* polyetheretherketone with hydroxyapatite. The radiographic markers meet ASTM F560 for unalloyed Tantalum. The difference in the material vs. the predicate material is hydroxyapatite filled polyetheretherketone (HA PEEK). HA is a naturally occurring mineral in bone and is widely used in the orthopedic field.
*Invibio and PEEK-OPTIMA® are registered trademarks of Invibio Limited. All rights reserved.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Innovasis. The logo is in black and features the company name in a bold, sans-serif font. Below the name are the words "INNOVATE / INVOLVE / INVENT" in a smaller font. The logo has a modern and clean design.
Intended Use: The Innovasis Px HA PEEK IBF System is an intervertebral body fusion device intended to stabilize a spinal segment to promote fusion using bone graft, in order to restrict motion and decrease pain.
> Users of these products are limited to physicians trained in orthopedic surgery. Clinical locations include hospitals and surgery sites equipped to perform spinal surgery.
The Innovasis Px HA™ PEEK IBF System is an intervertebral Indications for Use: body fusion device for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbar spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition. these patients may have up to a Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These implants are used to facilitate fusion in the lumbar spine and are placed via either a posterior (PLIF) or modified transforaminal (T-PLIF) approach.
> This device is intended to be used in pairs and with internal supplemental spinal fixation systems such as the Innovasis Excella® Spinal System. The interior of the Px HA implant is intended to be packed with autograft.
### Basis for Substantial Equivalence:
The Px HA PEEK IBF System has been subjected to risk analysis, engineering analysis and testing to recognized standards on new and artificially aged devices and has been shown to be substantially equivalent to the predicate device, K150500 Innovasis Px PEEK IBF System.
The technological characteristics were found to be substantially equivalent in terms of design, sizes, materials (biocompatibility profile and processing), and mechanical strength.
- Conclusion: The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe and as effective, and performs as well as or better than the legally marketed predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.