← Product Code [MAX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX) · K151352

# CONCORDE Bullet Spinal System (K151352)

_Medos International SARL · MAX · Oct 8, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K151352

## Device Facts

- **Applicant:** Medos International SARL
- **Product Code:** [MAX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX.md)
- **Decision Date:** Oct 8, 2015
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The CONCORDE Bullet Spinal System is indicated for use as intervertebral body fusion devices in skeletally mature patients with degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine (L2-S1). Patients should have six months of non-operative treatment prior to surgery. These implants are used to facilitate fusion in the lumbar spine and are placed via either a PLIF or TLIF approach using autogenous bone. When used as interbody fusion devices, these implants are intended for use with DePuy Spine supplemental internal fixation products. The CONCORDE Bullet Spinal System is indicated for use in the thoracolumbar spine (i.e., T1-L5) to replace a diseased vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore height of a collapsed vertebral body. This system is also indicated for treating fractures of the thoracic and lumbar spine. This system is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of fusion for a prolonged period. When used as a vertebral body replacement device, this system is intended for the use with DePuy Spine supplemental internal fixation products.

## Device Story

System of intervertebral body fusion cages and implantation instrumentation; cages feature bulleted nose for insertion, ridges/teeth for stability, and cavities for bone graft. Used in lumbar/thoracolumbar spine surgery; implanted by surgeons via PLIF or TLIF approaches. Provides structural support for fusion or vertebral body replacement; restores spinal column biomechanical integrity. Subject device introduces titanium alloy (Ti-6Al-4V-ELI) as an alternative to existing polymer/carbon fiber composite cages. Used with DePuy Spine supplemental internal fixation products. Benefits include restoration of vertebral height and spinal stability.

## Clinical Evidence

Bench testing only. Performance characterized per ASTM F-2077. Dynamic axial compression testing and finite element analysis were performed to verify mechanical strength of the titanium implants compared to the predicate.

## Technological Characteristics

Intervertebral body fusion/vertebral body replacement cages. Material: Ti-6Al-4V-ELI titanium alloy conforming to ASTM F136. Features: Bulleted nose, ridges/teeth for anti-migration, cavities for bone graft. Dimensions: Varying configurations to match patient anatomy. Sterilization: Offered sterile or non-sterile.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Bengal System ([K140759](/device/K140759.md))
- Bengal Stackable System ([K140759](/device/K140759.md))
- Devex System ([K140759](/device/K140759.md))
- Ocelot Stackable Cage System ([K140759](/device/K140759.md))
- Concorde (Bullet and Inline) Lumbar Interbody System ([K140759](/device/K140759.md))
- Cougar System ([K140759](/device/K140759.md))
- Cougar LS Lateral Cage System ([K140759](/device/K140759.md))
- Leopard System ([K140759](/device/K140759.md))
- DePuy Spine BENGAL, CONCORDE, COUGAR, DEVEX, and LEOPARD Systems ([K081917](/device/K081917.md))
- DEVEX Mesh System ([K023835](/device/K023835.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human face in profile, composed of three overlapping silhouettes. The silhouettes are arranged in a way that suggests a sense of community and support. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 8, 2015

MEDOS International Sárl c/o Mr. Jeffrey Shiffman Senior Regulatory Affairs Specialist DePuy Spine, Incorporated 325 Paramount Drive Raynham, Massachusetts 02767

Re: K151352

Trade/Device Name: CONCORDE Bullet Spinal System Regulation Number: 21 CFR 800.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX, MQP Dated: September 9, 2015 Received: September 10, 2015

Dear Mr. Shiffman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K151352

## Device Name CONCORDE Bullet Spinal System

## Indications for Use (Describe)

The CONCORDE Bullet Spinal System is indicated for use as intervertebral body fusion devices in skeletally mature patients with degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine (L2-S1). Patients should have six months of non-operative treatment prior to surgery. These implants are used to facilitate fusion in the lumbar spine and are placed via either a PLIF or TLIF approach using autogenous bone. When used as interbody fusion devices, these implants are intended for use with DePuy Spine supplemental internal fixation products.

The CONCORDE Bullet Spinal System is indicated for use in the thoracolumbar spine (i.e., T1-L5) to replace a diseased vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore height of a collapsed vertebral body. This system is also indicated for treating fractures of the thoracic and lumbar spine. This system is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of fusion for a prolonged period. When used as a vertebral body replacement device, this system is intended for the use with DePuy Spine supplemental internal fixation products.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| Prescription Use (Part 21 CFR 801 Subpart D) |  |
|----------------------------------------------|--|
| Over-The-Counter Use (21 CFR 801 Subpart C)  |  |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Medos International. The word "medos" is in black font, while "international" is in gray. There is a red symbol to the left of the word "medos", and a gray line underneath the entire logo.

## 510(K) SUMMARY

| A. | Submitter Information       |                                                                                                                                                 |                               |
|----|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|
|    | Manufacturer:               | Medos International Sárl<br>Chemin-Blanc 38<br>2400 Le Locle, Switzerland                                                                       |                               |
|    | Submitter:                  | DePuy Spine, Inc.<br>325 Paramount Drive<br>Raynham, MA 02767                                                                                   |                               |
|    | Contact Person:             | Jeffrey Shiffman<br>325 Paramount Drive<br>Raynham, MA 02767                                                                                    |                               |
|    | Telephone number:           | 508-977-3857                                                                                                                                    |                               |
|    | Fax number:<br>Email:       | 908-828-3797<br>jshiffma@its.jnj.com                                                                                                            |                               |
| B. | Date Prepared               | May 8, 2015                                                                                                                                     |                               |
| C. | Device Name                 |                                                                                                                                                 |                               |
|    |                             | Trade/Proprietary Name:                                                                                                                         | CONCORDE Bullet Spinal System |
|    | Common/Usual Name:          | Intervertebral Body Fusion Device<br>Vertebral Body Replacement Device                                                                          |                               |
|    | Classification Name (Code): | Intervertebral body fusion device<br>per 21 CFR § 888.3080 (MAX)<br>Spinal intervertebral body fixation orthosis<br>per 21 CFR § 888.3060 (MQP) |                               |

### D. Predicate Device Name

- Trade name: Bengal System, Bengal Stackable System, Devex System, Ocelot Stackable Cage System, Concorde (Bullet and Inline) Lumbar Interbody System, Cougar System, Cougar LS Lateral Cage System, Leopard System (K140759) – Primary predicate

{4}------------------------------------------------

# DePuy Spine BENGAL, CONCORDE, COUGAR, DEVEX, and LEOPARD Systems (K081917) - Additional predicate DEVEX Mesh System (K023835) - Additional predicate

### Device Description E.

The CONCORDE Bullet Spinal System is a system of intervertebral body fusion devices consisting of cages and implantation instrumentation. The cages feature a bulleted nose to aid in implant insertion. The proposed cages will offer a titanium alternative at certain geometries to the primary predicate system, which is currently only available in polymer/carbon fiber composite. The cages are available in varying shape and size configurations to match patient anatomy. The cages have ridges or teeth that resist rotation and migration and have cavities to accept packing of bone graft. The proposed cages will be implanted with the same instrumentation currently used in the system. The proposed cages will be offered sterile or non-sterile.

### F. Intended Use

The CONCORDE Bullet Spinal System is indicated for use as intervertebral body fusion devices in skeletally mature patients with degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine (L2-S1). Patients should have six months of non-operative treatment prior to surgery. These implants are used to facilitate fusion in the lumbar spine and are placed via either a PLIF or TLIF approach using autogenous bone. When used as interbody fusion devices, these implants are intended for use with DePuy Spine supplemental internal fixation products.

The CONCORDE Bullet Spinal System is indicated for use in the thoracolumbar spine (i.e., T1-L5) to replace a diseased vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore height of a collapsed vertebral body. This system is also indicated for treating fractures of the thoracic and lumbar spine. This system is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of fusion for a prolonged period. When used as a vertebral body replacement device, this system is intended for the use with DePuy Spine supplemental internal fixation products.

{5}------------------------------------------------

### G. Summary of Similarities and Differences in Technological Characteristics, Performance and Intended Use

The proposed additions to the CONCORDE Bullet Spinal System are equivalent to the predicate and reference devices within the system (reference K140759 and K081917), except for a change in the material. Currently, the CONCORDE Bullet Spinal System cages are available in polymer/carbon fiber composite. The proposed cages will be available in titanium alloy, the same material as used in the DEVEX System (K023835).

### H. Materials

The proposed additions to the CONCORDE Bullet Spinal System cages will be manufactured from Ti-6A1-4V-Eli titanium alloy and conforms to the standard specifications for surgical implant applications defined in ASTM F136.

### I. Performance Data

A summary of performance data per ASTM F-2077 is included to characterize the subject CONCORDE Bullet Spinal System cages addressed in this notification. Dynamic axial compression testing verified the mechanical strength of the proposed titanium implants. A finite element analysis also confirmed the axial compressive strength of the material compared to the predicate. No other testing was required because the only modification that would affect performance compared to predicate and reference devices within the CONCORDE Bullet Spinal System (K140759, K081917) is the material, which is used for the same indications in the predicate DEVEX System (K023835).

#### J. Conclusion

Both the Performance Testing and Substantial Equivalence Justification demonstrate that the proposed device is substantially equivalent to the predicate devices. Based on the 510(k) flowchart and the review of performance data, the proposed titanium cages in the CONCORDE Bullet Spinal System are substantially equivalent to the predicate devices. The technological characteristics between the proposed device and its predicates are substantially equivalent and the performance data raises no new questions of safety or effectiveness. Additionally, the proposed device does not constitute a new intended use.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K151352](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K151352)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
