← Product Code [MAX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX) · K150954

# Elite L Expandable Lumbar Fusion System (K150954)

_Innova Spinal Technologies, LLC · MAX · May 7, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K150954

## Device Facts

- **Applicant:** Innova Spinal Technologies, LLC
- **Product Code:** [MAX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX.md)
- **Decision Date:** May 7, 2015
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The ELITE™ L Expandable Lumbar Fusion System is indicated for use as intervertebral body fusion devices in sketetally mature patients with Degenerative Disc Disease (defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine from L2-L5. Patients should have received a minimum of six months of non-operative treatment prior to surgery. These patients may also have up to grade I spondylolisthesis or retrolisthesis at the involved level(s). These implants are intended for use with autogenous bone graft and supplemental internal fixation systems such as a pedicle screw or anterior plating system.

## Device Story

Titanium alloy (Ti6Al4V-ELI) intervertebral body fusion device; designed for lateral surgical approach; provides mechanical support to intradiscal space during fusion. Implant inserted in unexpanded state; expanded in situ to accommodate patient anatomy; features locking mechanism to maintain expanded position. Available in various footprints, heights, and parallel/lordotic angles; bi-convex end plates optimize vertebral body contact. Used by surgeons in clinical settings; requires supplemental internal fixation (e.g., pedicle screws or anterior plating). Benefits patient by restoring disc height and facilitating spinal fusion in degenerative disc disease.

## Clinical Evidence

Bench testing only. Performance evaluated via Finite Element Analysis (FEA) and mechanical testing per ASTM F2077 and ASTM F2267, including static and dynamic axial compression, subsidence, and expulsion testing.

## Technological Characteristics

Material: Medical grade titanium alloy (Ti6Al4V-ELI). Design: Expandable interbody fusion device with locking mechanism. Geometry: Various footprints, heights, and angles; bi-convex end plates. Standards: ASTM F2077, ASTM F2267. Energy source: None (mechanical). Connectivity: None.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Elite IBF by Innova Spine ([K133459](/device/K133459.md))
- Clydesdale Spinal System from Medtronic ([K100175](/device/K100175.md))
- Coroent XL Keeled System by Nuvasive ([K081611](/device/K081611.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three faces in profile, stacked on top of each other. The faces are oriented to the right. The logo is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in all caps.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Innova Spinal Technologies, LLC % Rich Jansen, Pharm.D. President Silver Pine Consulting, LLC 11821 Bramble Cove Drive Fort Myers, Florida 33905

May 7, 2015

Re: K150954

Trade/Device Name: ELITE™ L Expandable Lumbar Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: April 7, 2015 Received: April 9, 2015

Dear Dr. Jansen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{1}------------------------------------------------

Page 2 - Rich Jansen, Pharm.D.

(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K150954

### Device Name

ELITE™ L Expandable Lumbar Fusion System

#### Indications for Use (Describe)

The ELITE™ L Expandable Lumbar Fusion System is indicated for use as intervertebral body fusion devices in sketetally mature patients with Degenerative Disc Disease (defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine from L2-L5. Patients should have received a minimum of six months of non-operative treatment prior to surgery. These patients may also have up to grade I spondylolisthesis or retrolisthesis at the involved level(s). These implants are intended for use with autogenous bone graft and supplemental internal fixation systems such as a pedicle screw or anterior plating system.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) SUMMARY

## Elite L Expandable Lumbar Fusion System

| Date:               | April 7, 2015                                                                                                             |
|---------------------|---------------------------------------------------------------------------------------------------------------------------|
| Submitter:          | Armando Varela<br>Innova Spinal Technologies, Inc.<br>22324 Boyaca Ave<br>Boca Raton, FL 33433<br>Telephone: 561-504-5431 |
| Regulatory Contact: | Rich Jansen, Pharm. D.<br>Silver Pine Consulting, LLC<br>612-281-5505                                                     |
| Product             |                                                                                                                           |
| Trade Name:         | Elite TM L Expandable Lumbar Fusion System                                                                                |
| Common Name:        | Intervertebral Body Fusion Device                                                                                         |
| Product Class:      | Class II                                                                                                                  |
| Classification:     | 21 CFR 888.3080                                                                                                           |
| Product Code:       | MAX                                                                                                                       |
| Panel Code:         | 87                                                                                                                        |

## Purpose:

The purpose of this submission is to add devices to be implanted from the lateral surgical approach.

## Device Description

The Elite™ L is a titanium alloy (Ti6Al4V-ELI) implant that is designed to provide mechanical support of the intradiscal space while biologic fusion occurs. The device is available in a variety of footprints, all of which are inserted in the unexpanded state and are expanded in situ. There are various heights to accommodate varied patients' anatomy. The implants are available in parallel and lordotic angles, with the implant end plates being bi-convex to optimize contact with the vertebral body end plates. All implants are made from medical grade titanium alloy. Implants have a locking mechanism to hold the implant in the expanded position once deployed.

## Predicate Device

The predicate devices include the Primary Predicate which is the Elite IBF by Innova Spine (K133459). Additional predicates include the Clydesdale Spinal System from Medtronic (K100175) and the Coroent XL Keeled System by Nuvasive (K081611).

### Intended Use / Indications for Use

The ELITE™ L Expandable Lumbar Fusion System is indicated for use as intervertebral body fusion devices in skeletally mature patients with Degenerative Disc Disease (defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine from L2-L5. Patients should have received a minimum of six months of non-operative treatment prior to surgery. These patients may also have up to grade I spondylolisthesis or retrolisthesis at the involved level(s). These implants are intended for use with autogenous bone graft and supplemental internal fixation systems such as a pedicle screw or anterior plating system.

{4}------------------------------------------------

### Performance Testing

The Primary Predicate Device, the Elite Expandable Interbody Fusion Device was tested according to ASTM F2077 and ASTM F2267. Testing included static and dynamic axial compression, subsidence and expulsion. The Elite™ L device was compared to the Primary Predicate using FEA and demonstrated greater strength and durability to static and dynamic forces compared to the Primary Predicate. A complete analysis is provided.

### Performance Characteristics:

The Elite™ L is substantially equivalent to the predicate devices in regards to:

- . Indications for Use
- . Materials
- . Dimensions
- . Function
- Mechanical testing .

### Summary:

The Elite™ L demonstrates substantial equivalence to the predicate devices and raises no new questions of safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K150954](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K150954)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
