K133371 · Spineology, Inc. · MAX · Jan 9, 2014 · Orthopedic
Device Facts
Record ID
K133371
Device Name
RAMPART-L
Applicant
Spineology, Inc.
Product Code
MAX · Orthopedic
Decision Date
Jan 9, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Intended Use
RampartTM L is an intervertebral body fusion device indicated for intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade I spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. RampartTM L is designed for use with autograft as an adjunct to fusion and is intended for use with supplemental fixation systems cleared by the FDA for use in the lumbar spine.
Device Story
Rampart-L is an intervertebral body fusion device used in lumbar spine surgery. It is implanted by a surgeon to facilitate fusion at one or two contiguous levels (L2-S1). The device is packed with autograft to promote bone growth and must be used in conjunction with FDA-cleared supplemental fixation systems. It serves as a structural spacer to maintain disc space height and stability during the fusion process. The device is provided sterile and is intended for single-patient use.
Clinical Evidence
Bench testing only. Mechanical performance was evaluated via static compression, static shear compression (ASTM F2077), and subsidence testing (ASTM F2267). Results demonstrated performance characteristics similar to the predicate devices.
Technological Characteristics
Material: PEEK-OPTIMA LT-1 with tantalum markers. Form factor: Intervertebral body fusion spacer available in various lengths and heights. Sterilization: Provided sterile. Testing standards: ASTM F2077, ASTM F2267.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) and up to Grade I spondylolisthesis at one or two contiguous lumbar levels (L2-S1) who have failed six months of non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
K153082 — Rampart A Lumbar Interbody Fusion Device · Spineology, Inc. · Feb 4, 2016
K171914 — Cortina Lumbar Cage System · Neurostructures, Inc. · Oct 25, 2017
K130396 — RAMPART O AND RAMPART P · Spineology, Inc. · Mar 25, 2013
K131216 — RAMPART O · Spineology, Inc. · May 30, 2013
K113030 — SPINEOLOGY PEEK BULLET LUMBAR INTERBODY FUSION DEVICE · Spineology, Inc. · Jan 26, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary
JAN - 9 2014 ・
.
| Applicant: | Spineology Inc.<br>7800 3rd Street N., Suite 600<br>Saint Paul, MN 55128<br>Phone: 651-256-8500<br>Fax: 651-256-8505 |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Bryan Becker |
| Date Prepared: | October 30, 2013 |
| Trade Name: | RampartTM-L |
| Product Classification<br>and Code: | Class II Medical Device, Product Code MAX, 21 CFR 888.3080 |
| Predicate Device(s): | Spineology PEEK Lumbar Interbody Fusion Devices/Rampart™-O<br>(K111880, K113030, K120293, K123652, K130396, K131216,<br>and K132053) |
| Device Description: | RampartTM L devices are designed for use with autograft as an adjunct to<br>fusion and are intended for use with supplemental fixation systems cleared<br>for use in the lumbar spine. The device is available in a range of lengths<br>and heights and is made of PEEK-OPTIMA LT-1 with tantalum markers.<br>The device is provided sterile and associated instruments are provided non-<br>sterile. |
| Intended Use: | RampartTM L is an intervertebral body fusion device indicated for<br>intervertebral body fusion at one level or two contiguous levels in the<br>lumbar spine from L2 to S1 in patients with degenerative disc disease<br>(DDD) with up to Grade I spondylolisthesis at the involved level(s). DDD<br>is defined as back pain of discogenic origin with degeneration of the disc<br>confirmed by patient history and radiographic studies. These patients<br>should be skeletally mature and have had six months of non-operative<br>treatment. RampartTM L is designed for use with autograft as an adjunct<br>to fusion and is intended for use with supplemental fixation systems cleared<br>by the FDA for use in the lumbar spine. |
| Purpose of this 510(k): | The purpose of this 510(k) is to extend the range of sizes of Rampart O to<br>create the Rampart-L devices. |
| Summary of<br>Technological<br>Characteristics: | The device is shown to be substantially equivalent to the intended use,<br>materials, configuration, and performance characteristics of the predicate<br>products. |
| Testing | Preclinical testing according to ASTM F2077 (static compression and<br>static sheer compression) and ASTM F2267 (subsidence), was used to<br>justify substantial equivalence. This testing demonstrated similar<br>performance characteristics to the identified predicate devices. |
| Conclusion: | The information submitted in this premarket notification supports a<br>determination that RampartTM L is substantially equivalent in<br>technological characteristics and intended use to the predicate device. |
{1}------------------------------------------------
#### K133371 Page 2 of 2
.
.
:
:
·
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the department's emblem, which features a stylized caduceus-like symbol with three wavy lines extending from a central point. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the emblem within the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 9, 2014
Spineology, Incorporated Mr. Bryan Becker Regulatory Affairs Manager and Compliance Officer 7800 Third Street North, Suite 600 Saint Paul, Minnesota 55128
Re: K133371
Trade/Device Name: Rampart™-L Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: December 13, 2013 · Received: December 16, 2013
Dear Mr. Becker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
#### Page 2 - Mr. Bryan Becker
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Lori A. Wigqins
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K133371
Device Name
Rampart-L
Indications for Use (Describe)
Rampar™ L is an intervertebral body fusion device indicated for intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade I spondyolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and
radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Rampart™ L is designed for use with autograft as an adjunct to fusion and is intended for use with supplemental fixation systems cleared by the FDA for use in the lumbar spine.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
: ::
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
... . FOR FDA USE ONLY : 14 : 14
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# STATE PARTY
**Division of Orthopedic Devices**
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
{5}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.