← Product Code [MAX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX) · K132907

# PLIF LITE AND TLIF LITE (K132907)

_Sintea Plustek · MAX · Jul 24, 2014 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K132907

## Device Facts

- **Applicant:** Sintea Plustek
- **Product Code:** [MAX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX.md)
- **Decision Date:** Jul 24, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The PLIF Lite and TLIF Lite Lumbar Interbody Fusion Devices are indicated for intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. The PLIF Lite and TLIF Lite Lumbar Interbody Fusion Devices are designed for use with autograft bone to facilitate fusion and are intended for use with supplemental fixation systems cleared for use in the lumbar spine.

## Device Story

Lumbar interbody fusion device; PEEK implant; available in lordotic and non-lordotic configurations; includes titanium alloy (ASTM F-136) radiographic markers for visualization. Used by surgeons in clinical settings to facilitate intervertebral body fusion; requires concurrent use of autograft bone and supplemental lumbar fixation systems. Implanted via PLIF or TLIF surgical techniques to address discogenic back pain; provides structural support to the disc space to promote fusion.

## Clinical Evidence

Bench testing only. No clinical data provided. Mechanical performance validated via static and dynamic axial compression, compression shear, torsion, subsidence, and expulsion testing per ASTM F2077 and ASTM F2267 standards.

## Technological Characteristics

Material: PEEK; Radiographic markers: Titanium alloy (ASTM F-136). Form factor: Lumbar interbody cages (PLIF/TLIF). Energy source: None (mechanical). Sterilization: Provided sterile. Connectivity: None. Software: None.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Stryker (formerly Surgical Dynamics) Ray TFC Cage (P950019a)
- Depuy Spine Lumbar I/F Cage Implant System ([P960025](/device/P960025.md))
- Zimmer (formerly Spinal Concept) Fidji PEEK Lumbar Cage ([K042714](/device/K042714.md))
- R Tree Innovations Epicage ([K092901](/device/K092901.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) SUMMARY Sintea Plustek's PLIF Lite and TLIF Lite Interbody Fusion system

# JUL 2 4 2014

Date: ·

September 16, 2013

Contact:

Sintea Plustek Guido Zorzoli 305-673-6226 407 Lincoln Road Suite 10/L Miami, FL 33139

PLIF Lite and TLIF Lite (PLIF Lite and TLIF Lite Interbody Fusion system) Trade Name: Common Name: Intervertebral Body Fusion Device Product Class: Class II Classification: 888.3080 Product Code: MAX Panel Code: 87

# Name of Sponsor

- Sintea Plustek, LLC 407 Lincoln Road Suite 10/L Miami, FL 33139 305-673-6226
## Device Description

The PLIF Lite and TLIF Lite Lumbar Interbody Fusion Devices are made from PEEK. The PLIF Lite implants are available in lordotic and non-lordotic form, while the TLIF implants are provided with 7 degrees of lordosis only. The implants are available in various heights and lengths to accommodate patients' anatomy. The implants are provided sterile, and a set of instrument for implantation is provided to facilitate implantation. Three radiographic markers made of titanium alloy (ASTM F-136) are included in each implant to allow radiographic visualization.

# Predicate Device

The PLIF Lite and TLIF Lite Interbody Fusion system is substantially equivalent to legally marketed predicate devices. The predicate devices are the Stryker (formerly Surgical Dynamics) Ray TFC Cage (P950019a), the Depuy Spine Lumbar I/F Cage Implant System (P960025), the Zimmer (formerly Spinal Concept) Fidji PEEK Lumbar Cage (K042714) and the R Tree Innovations Epicage (K092901).

# Intended Use / Indications for Use

The PLIF Lite and TLIF Lite Lumbar Interbody Fusion Devices are indicated for intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. The PLIF Lite and TLIF Lite Lumbar Interbody Fusion Devices are designed for use with autograft bone to facilitate fusion

{1}------------------------------------------------

and are intended for use with supplemental fixation systems cleared for use in the lumbar spine.

## Performance Testing

The PLIF Lite and TLIF Lite Lumbar Interbody devices were tested according to ASTM F2077 and ASTM F2267. Testing included static and dynamic axial compression, static and dynamic compression shear, static torsion, subsidence and expulsion. Test results demonstrate that the PLIF Lite and TLIF Lite devices are substantially equivalent to the predicate devices.

#### Summary:

The PLF Lite and TLIF Lite Interbody Fusion system is substantially equivalent to the predicate devices in regards to:

- Indications for Use .
- Materials .
- . Dimensions
- Function .
- . Mechanical testing

There are no significant differences in technological characteristics compared to the predicate device.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 24, 2014

Sintea Plustek, LLC % Rich Jansen, Pharm.D. Silver Pine Consulting. LLC 11821 Bramble Cove Drive Fort Myers, Florida 33905

Re: K132907

Trade/Device Name: PLIF Lite and TLIF Lite Interbody Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: June 23, 2014 Received: June 26, 2014

Dear Dr. Jansen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.

If vour device is classified (sec above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Rich Jansen, Pharm.D.

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Ronald P. Jean -S for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known) K132907

### Device Name

PLIF Lite and TLIF Lite Interbody Fusion System

#### Indications for Use (Describe)

The PLIF Lite and TLIF Lite Lumbar Interbody Fusion Devices are indicated for intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. The PLIF Lite and TLIF Lite Lumbar Interbody Fusion Devices are designed for use with autograft bone to facilitate fusion and are intended for use with supplemental fixation systems cleared for use in the lumbar spine.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

# FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

# Katherine D. Kavlock, PhD

# Division of Orthopedic Devices

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K132907](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K132907)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
