K132700 · Medtronic Sofamor Danek, Inc. · MAX · Sep 26, 2013 · Orthopedic
Device Facts
Record ID
K132700
Device Name
PERIMETER INTERBODY FUSION DEVICE
Applicant
Medtronic Sofamor Danek, Inc.
Product Code
MAX · Orthopedic
Decision Date
Sep 26, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The PERIMETER® Interbody Fusion Device is indicated for interbody fusion with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implanted via a variety of open or minimally invasive approaches. These approaches include anterior, lateral and oblique. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared for use in the lumbar spine.
Device Story
PERIMETER® Interbody Fusion Device consists of hollow cages inserted between lumbar or lumbosacral vertebral bodies to provide support and correction during fusion surgery. Implants are packed with autogenous bone graft; designed for use with supplemental fixation instrumentation. Available in Titanium Alloy or PEEK Optima LT1; features teeth on superior/inferior surfaces for expulsion resistance. Titanium version includes lateral windows for graft visibility. Used by surgeons in open or minimally invasive (anterior, lateral, oblique) lumbar spine procedures. Device provides structural support to facilitate fusion, potentially reducing discogenic back pain.
Clinical Evidence
No clinical data. Bench testing and cadaveric validation performed to confirm labeling modifications.
Technological Characteristics
Materials: Titanium Alloy (Ti-6Al-4V ELI) or PEEK Optima LT1. Dimensions: 8-20mm height, 21-28mm length, 19-38mm width; 4-15 degree lordosis options. Features: Hollow geometry for bone graft, surface teeth for fixation. Sterilization: Sterile (PEEK) or non-sterile options.
Indications for Use
Indicated for interbody fusion with autogenous bone graft in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1, including those with up to Grade 1 Spondylolisthesis or retrolisthesis. Requires six months of prior non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
K131669 — PERIMETER INTERBODY FUSION DEVICE · Medtronic Sofamor Danek USA, Inc. · Nov 1, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
K132700 Page 1 of 2
# PERIMETER® Interbody Fusion Device 510(k) Summary August 2013
## I. COMPANY:
II. CONTACT:
Medtronic Sofamor Danek USA. Inc 1800 Pyramid Place Memphis, Tennessee 38132
Ankit K. Shah Regulatory Affairs Specialist Telephone: (901) 344-1272 Fax: (901) 346-9738
PERIMETER® Interbody Fusion
Intervertebral Fusion with Bone
SEP 2 6 2013
Graft.
## III. PROPRIETARY TRADE NAME:
IV. CLASSIFICATION NAMES:
COMMON NAME:
CLASS:
PRODUCT CODE:
MAX (21 CFR 888.3080)
Interbody Fusion Device
## V. PRODUCT DESCRIPTION:
The PERIMETER® Interbody Fusion Device consists of cages of various widths and heights which can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. The hollow geometry of the implants allows them to be packed with autogenous bone graft. The PERIMETER® Interbody Device is to be used with supplemental instrumentation.
l l
Device
Lumbar
The device is offered in Titanium Alloy (Titanium-6Aluminum-4Vanadium ELI) or PEEK Optima LT1 (Polyetheretherketone). This interbody device is offered in sterile (PEEK) or non-sterile (PEEK and Titanium Alloy) forms.
The PERIMETER® Interbody Fusion Device is offered in a variety of sizes ranging from 8mm to 20mm in height, 21mm to 28mm in length and between 19mm and 38mm in width. An array of lordosis options are provided for this device spanning from 4 degrees to 15 degrees of angulation. Both the PEEK and Titanium Alloy (Titanium-6Aluminum-4 Vanadium EL.I) devices are designed with teeth across both the superior and inferior surfaces to allow the implant to grip the superior and inferior end plates, thus providing
{1}------------------------------------------------
expulsion resistance. Additionally, the Titanium Alloy (Titanium-6Aluminum-4 Vanadium ELI) version of this device offers lateral windows for visibility of the autogenous bone graft.
### · VI. INDICATIONS FOR USE:
The PERIMETER® Interbody Fusion Device is indicated for interbody fusion with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via a variety of open or minimally invasive approaches. These approaches include anterior, lateral and oblique. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared for use in the lumbar spine.
#### VII. Summary of the Technological Characteristics:
The subject PERIMETER® Interbody Fusion Device has the same indications, intended use, fundamental scientific technology, design and material as the previously FDA cleared predicates; PERIMETER® Interbody Fusion Device K111525 (S.E. 08/24/2011) and K090353 (S.E. 09/29/2009). The only change in the application is a modification to the labeling.
## VIII. Identification of Legally Marketed Devices:
The fundamental scientific technology, design features and indications for use for the subject PERIMETER® Interbody Fusion Device are identical to the predicate PERIMETER® Interbody Fusion Devices K111525 (S.E. 08/24/2011) and K090353 (S.E. 09/29/2009).
#### IX. Discussion of Non-Clinical Testing:
This modified labeling has been confirmed for the subject device by surgeons performing the procedure on cadavers. Data from this confirmatory validation supports the labeling modification. The intended use has not changed as the result of this labeling modification. Medtronic believes that the subject device is substantially equivalent to the predicate device.
## X. Conclusion:
Validation and risk analysis were completed for the labeling change. Based on the validation, risk analysis and additional supporting documentation provided in this premarket notification, Medtronic believes the subject system demonstrates substantial equivalence to listed predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features an eagle-like symbol with three curved lines representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Ficalth Service
September 26, 2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G6(b9 Silver Spring, MD 20993-0002
Medtronic Sofamor Danek USA, Incorporated Ankit K. Shah Regulatory Affairs Specialist 1800 Pyramid Place Memphis. Tennessee 38132
Re: K132700
Trade/Device Name: PERIMETER® Interbody Fusion Device Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: August 28, 2013 Received: August 29, 2013
Dear Ankit K. Shah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that I DA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 – Ankit K. Shah
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the requlation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## 510(k) Number (if known): Not yet assigned
## Device Name: PERIMETER® Interbody Fusion Device
## Indications for Use:
The PERIMETER® Interbody Fusion Device is indicated for interbody fusion with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implanted via a variety of open or minimally invasive approaches. These approaches include anterior, lateral and oblique. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared for use in the lumbar spine.
Prescription Use X (Part 2) CFR 80) Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anton E. Dmitriev, PhD Division of Orthopedic Devices
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.