COUGAR LS 15 LATERAL CAGE SYSTEM

K122896 · Medos International SARL · MAX · Oct 18, 2012 · Orthopedic

Device Facts

Record IDK122896
Device NameCOUGAR LS 15 LATERAL CAGE SYSTEM
ApplicantMedos International SARL
Product CodeMAX · Orthopedic
Decision DateOct 18, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Cougar® LS Lateral Cage System is indicated for use in the thoracolumbar spine (i.e., T1 to L5) to replace a diseased vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. This system is also indicated for treating fractures of the thoracic and lumbar spine. The system is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period. When used as a vertebral body replacement device, this system is intended for use with DePuy Spine supplemental internal fixation. The Cougar® LS Lateral Cage System is also indicated for intervertebral body fusion with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via an open or a minimally invasive lateral approach. When used as an interbody fusion device, this system is intended for use with DePuy Spine supplemental internal fixation.

Device Story

Carbon Fiber Reinforced Polymer (CFRP) intervertebral cage system; used for vertebral body replacement or interbody fusion; implanted via open or minimally invasive lateral surgical approach. Cages feature teeth to resist migration/rotation and cavities for autogenous bone graft; radiolucent design includes tantalum X-ray markers for radiographic assessment. Used by surgeons in clinical settings to restore spinal height and biomechanical integrity; requires supplemental internal fixation. Benefits include spinal decompression and stabilization for patients with DDD, tumors, or fractures.

Clinical Evidence

Bench testing only. Performed static and dynamic compression, compressive shear, expulsion, and subsidence testing per ASTM F2077-11 and ASTM F2267-04 standards.

Technological Characteristics

Material: Carbon Fiber Filled PEEK-OPTIMA LT1 Compound (CFRP). Features: Tantalum X-ray markers, teeth for fixation. Dimensions: 18mm/21mm widths, 10-20mm heights, 45-60mm lengths. Lordotic configuration: 15°. Standards: ASTM F2077-11, ASTM F2267-04.

Indications for Use

Indicated for skeletally mature patients with thoracolumbar (T1-L5) tumors, vertebral body collapse, or thoracic/lumbar fractures requiring vertebral body replacement; also indicated for L2-S1 degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis or retrolisthesis, requiring intervertebral body fusion with autogenous bone graft after 6 months of failed non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 长122896 page 1 of 3 : # 510(K) SUMMARY OCT 18 2012 | A. | Submitter Information | | |----|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------| | | Manufacturer: | Medos International Sárl | | | | Chemin-Blanc 38 | | | | 2400 Le Locle, Switzerland | | | Submitter: | DePuy Spine, Inc. | | | | 325 Paramount Drive | | | | Raynham, MA 02767 | | | Contact Person: | Kirsten Lehmuller | | | | 325 Paramount Drive | | | | Raynham, MA 02767 | | | Telephone number: | 508-828-3291 | | | Fax number: | 508-828-3797 | | | Email: | klehmull@its.jnj.com | | B. | Date Prepared | September 20, 2012 | | C. | Device Name | | | | Trade/Proprietary Name: | Cougar® LS Lateral Cage System | | | Common/Usual Name: | Spinal Intervertebral Body Fixation Orthosis,<br>Intervertebral body fusion device | | | Classification Name: | Spinal Intervertebral Body Fixation Orthosis<br>per 21 CFR §888.3060<br>Intervertebral body fusion device<br>per 21 CFR §888.3080 | #### D. Predicate Device Name Trade name: DePuy Spine Cougar® LS Lateral Cage System (K082128, K110454) {1}------------------------------------------------ #### E. Device Description The Cougar® LS Lateral System implants are manufactured from Carbon Fiber Reinforced Polymer (CFRP) material. Cages are available with 15° lordotic configurations, widths of 18mm and 21mm, and heights ranging from 10mm to 20mm in 2mm increments. The cages' lengths are 45mm, 50mm, 55mm, and 60mm. The cage structure is radiolucent with tantalum X-ray wire so that healing can be assessed by normal radiographic methods. The cages have teeth that resist rotation and migration and have cavities to accept packing of autogenous bone graft. The implants may be utilized in either an open or minimally invasive surgical approach. The implants are placed using a lateral surgical approach. #### F. Intended Use The Cougar® LS Lateral Cage System is indicated for use in the thoracolumbar spine (i.e., T1 to L5) to replace a diseased vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. This system is also indicated for treating fractures of the thoracic and lumbar spine. The system is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period. When used as a vertebral body replacement device, this system is intended for use with DePuy Spine supplemental internal fixation. The Cougar® LS Lateral Cage System is also indicated for intervertebral body fusion with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six {2}------------------------------------------------ months of non-operative treatment. These implants may be implanted via an open or a minimally invasive lateral approach. When used as an interbody fusion device, this system is intended for use with DePuy Spine supplemental internal fixation. ### F. Summary of Similarities and Differences in Technological Characteristics, Performance and Intended Use The proposed modification to the DePuy Cougar LS Lateral System (i.e. increase in lordotic angle) results in a device that is substantially equivalent to the predicates. The footprint of the subject devices remains unchanged from the predicate Cougar LS devices. The increase in lordotic angle provides the same strength and performance as the predicate devices. The materials and technology remain identical to the predicate system. #### Materials G. The proposed cages are manufactured from Carbon Fiber Filled PEEK-OPTIMA LT1 Compound (CFRP). #### H. Performance Data Performance data per ASTM F2077-11 and ASTM F2267-04 was submitted to characterize the subject Cougar LS Lateral Cage implants addressed in this notification. This testing was comprised of static and dynamic compression, static and dynamic compressive shear, expulsion, and subsidence testing on the proposed device. #### l. Conclusion Both the Performance Testing and Substantial Equivalence Justification demonstrate that the device is as safe, as effective, and performs as well as the predicate device. {3}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular fashion around the eagle. ### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Medos International Sárl % Depuy Spine, a Johnson and Johnson Company Ms. Kristen Lehmuller Regulatory Affairs Associate 325 Paramount Drive Raynham, Massachusetts 02767 OCT 1 8 2012 Re: K122896 Trade/Device Name: Cougar® LS Lateral Cage Implants Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX, MQP Dated: September 20, 2012 Received: September 21, 2012 Dear Ms. Lehmuller: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {4}------------------------------------------------ ## Page 2 - Ms. Kristen Lehmuller CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Sincerely yours, Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # INDICATIONS FOR USE STATEMENT 510(k) Number (if known): ≤ (22796 Device Name: Cougar® LS Lateral Cage Implants ## Indications For Use: The Cougar® LS Lateral Cage System is indicated for use in the thoracolumbar spine (i.e., T1 to L5) to replace a diseased vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. This system is also indicated for treating fractures of the thoracic and lumbar spine. The system is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period. When used as a vertebral body replacement device, this system is intended for use with DePuy Spine supplemental internal fixation. The Cougar® LS Lateral Cage System is also indicated for intervertebral body fusion with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via an open or a minimally invasive lateral approach. When used as an interbody fusion device, this system is intended for use with DePuy Spine supplemental internal fixation. Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR Over-The-Counter Use_ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number k122896
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...