NUVASIVE COROENT TITANIUM SYSTEM

K120918 · Nuvasive, Inc. · MAX · Jun 28, 2012 · Orthopedic

Device Facts

Record IDK120918
Device NameNUVASIVE COROENT TITANIUM SYSTEM
ApplicantNuvasive, Inc.
Product CodeMAX · Orthopedic
Decision DateJun 28, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Intended Use

Intervertebral Body Fusion The NuVasive CoRoent Titanium System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft to facilitate fusion. The CoRoent Titanium System is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The devices are to be used in patients who have had at least six months of non-operative treatment. The System is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the lumbar spine. Partial Vertebral Body Replacement The NuVasive CoRoent Titanium System may also be used as a partial vertebral body replacement device indicated for use in the thoracolumbar spine (T1 to L5) to replace a diseased or damaged vertebral body caused by tumor or fracture, to restore height of a collapsed vertebral body, and to achieve decompression of the spinal cord and neural tissues. The System is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the thoracic and lumbar spine. Allograft or autograft material may be used at the surgeon's discretion.

Device Story

NuVasive CoRoent Titanium System comprises titanium alloy (Ti-6Al-4V) spinal implants; designed for intervertebral body fusion or partial vertebral body replacement. Implants available in various sizes to accommodate patient anatomy. Used in conjunction with FDA-cleared supplemental internal spinal fixation systems; requires autogenous bone graft (fusion) or allograft/autograft (replacement). Surgeon implants device during spinal surgery to restore vertebral height, facilitate fusion, or decompress neural tissues. Benefits include stabilization of diseased/damaged spinal segments and restoration of structural integrity. Performance validated via Finite Element Analysis (FEA) covering axial compression, compression shear, torsion, and subsidence.

Clinical Evidence

Bench testing only. No clinical data provided. Performance equivalence established via Finite Element Analysis (FEA) comparing subject device to predicate across axial compression, compression shear, torsion, and subsidence metrics.

Technological Characteristics

Material: Titanium alloy (Ti-6Al-4V) per ASTM F136 or ISO 5832-3. Form factor: Spinal implants in various sizes for lumbar/thoracolumbar application. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.

Indications for Use

Indicated for skeletally mature patients requiring intervertebral body fusion (lumbar L2-S1) for degenerative disc disease with up to Grade I spondylolisthesis after 6 months of failed non-operative treatment, or partial vertebral body replacement (thoracolumbar T1-L5) for tumor or fracture-related vertebral damage.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JUN 28 2012 Image /page/0/Picture/2 description: The image shows the logo for Nuvasive. The logo consists of a stylized leaf-like shape on the left, followed by the word "NUVASIVE" in capital letters. Below the word "NUVASIVE" is the tagline "Speed of Innovation" in a smaller font. 510(k) Premarket Notification CoRoent Titanium System # 510(k) Summary In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided: ### Submitted by: A. Sheila Bruschi Associate Manager, Regulatory Affairs NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (858) 320-4515 Fax: (858) 320-4615 Date Prepared: March 23, 2012 #### B. Device Name | Trade or Proprietary Name: | <i>NuVasive CoRoent</i> ® Titanium System | |----------------------------|--------------------------------------------------------------------------------| | Common or Usual Name: | Intervertebral Body Fusion Device,<br>Spinal Vertebral Body Replacement Device | | Classification Name: | Intervertebral Body Fusion Device,<br>Spinal Vertebral Body Replacement Device | | Device Class: | Class II | |-----------------|-------------------------------| | Classification: | 21 CFR § 888.3060, § 888.3080 | | Product Code: | MAX, MQP | #### C. Predicate Devices . The subject NuVasive CoRoent Titanium System is substantially equivalent to the following predicate device: NuVasive CoRoent System (K071795) #### D. Device Description The NuVasive CoRoent Titanium System is a lumbar interbody system manufactured from titanium alloy (Ti-6Al-4V) conforming to ASTM F136 or ISO 5832-3. The implants are available in a variety of sizes to suite the individual pathology and anatomical conditions of the patient. The device is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the lumbar spine. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the Nuvasive logo. The logo consists of a stylized letter "i" enclosed in a partial circle on the left, followed by the word "NUVASIVE" in capital letters. Below the word "NUVASIVE" is the tagline "Speed of Innovation" in a smaller font size. # 510{k} Premarket Notification CoRoent Titanium System #### E. Intended Use ## Intervertebral Body Fusion The NuVasive CoRoent Titanium System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft to facilitate fusion. The CoRoent Titanium System is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The devices are to be used in patients who have had at least six months of non-operative treatment. The System is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the lumbar spine. # Partial Vertebral Body Replacement The NuVasive CoRoent Titanium System may also be used as a partial vertebral body replacement device indicated for use in the thoracolumbar spine (T1 to L5) to replace a diseased or damaged vertebral body caused by tumor or fracture, to restore height of a collapsed vertebral body, and to achieve decompression of the spinal cord and neural tissues. The System is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the thoracic and lumbar spine. Allograft or autograft material may be used at the surgeon's discretion. #### F. Technological Characteristics As was established in this submission, the subject CoRoent Titanium System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have the same technological characteristics to its predicate devices through comparison in areas including design, intended use, material composition, and function. #### G. Performance Data Finite Element Analysis (FEA) was performed to demonstrate that the subject CoRoent Titanium System is substantially equivalent to other predicate devices for the following: - Axial Compression ● - Compression Shear ● - Torsion . ● - Subsidence . The results demonstrate that the subject CoRoent Titanium System presents no new worstcase for performance testing, and the subject device was therefore found to be substantially equivalent to the predicate. ### H. Conclusions Based on the indications for use, technological characteristics, and comparison to predicate devices, the subject CoRoent Titanium System has been shown to be substantially equivalent to legally marketed predicate devices, and safe and effective for its intended use. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol. The text is in all caps and appears to be in a sans-serif font. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 JUN 2 8 2012 NuVasive, Inc. % Ms. Sheila Bruschi Associate Manager, Regulatory Affairs 7475 Lusk Blvd. San Diego, California 92121 Re: K120918 > Trade/Device Name: NuVasive CoRoent® Titanium System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX, MQP Dated: June 6, 2012 Received: June 7, 2012 Dear Ms. Bruschi: We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Ms. Sheila Bruschi Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Sincerely yours, for Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): NuVasive® CoRoent® Titanium System Device Name: Indications For Use: # Intervertebral Body Fusion The NuVasive CoRoent Titanium System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft to facilitate fusion. The CoRoent Titanium System is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The devices are to be used in patients who have had at least six months of non-operative treatment. The System is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the lumbar spine. # Partial Vertebral Body Replacement The NuVasive CoRoent Titanium System may also be used as a partial vertebral body replacement device indicated for use in the thoracolumbar spine (T1 to L5) to replace a diseased or damaged vertebral body caused by tumor or fracture, to restore height of a collapsed vertebral body, and to achieve decompression of the spinal cord and neural tissues. The System is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the thoracic and lumbar spine. Allograft or autograft material may be used at the surgeon's discretion. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Diysson Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices K120918 510(k) Number_ Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...