← Product Code [MAX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX) · K120293

# SPINEOLOGY PEEK LUMBAR INTERBODY FUSION DRIVE (K120293)

_Spineology, Inc. · MAX · Feb 29, 2012 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K120293

## Device Facts

- **Applicant:** Spineology, Inc.
- **Product Code:** [MAX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX.md)
- **Decision Date:** Feb 29, 2012
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

Crescent: The Spineology PEEK Lumbar Interbody Fusion Device is an intervertebral body fusion device indicated for intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade I spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These devices may be implanted via an open or a minimally invasive transforaminal approach. The Spineology PEEK Lumbar Interbody Fusion Device is designed for use with autograft to facilitate fusion and is intended for use with supplemental fixation systems cleared by the FDA for use in the lumbar spine. Bullet: The Spineology PEEK Bullet Lumbar Interbody Fusion Device is an intervertebral body fusion device indicated for intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade I spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These devices may be implanted singly or in pairs via an open or a minimally invasive posterior or transforaminal approach. The Spineology PEEK Bullet Lumbar Interbody Fusion Device is designed for use with autograft to facilitate fusion and is intended for use with supplemental fixation systems cleared by the FDA for use in the lumbar spine.

## Device Story

Spineology PEEK Lumbar Interbody Fusion Devices (Crescent and Bullet models) are spinal implants used to facilitate intervertebral body fusion. The devices are implanted by surgeons in an OR setting via open or minimally invasive posterior/transforaminal approaches. They are designed to be used with autograft and supplemental fixation systems. The current modification adds tantalum markers to the PEEK devices to improve radiographic visibility. The device provides structural support to the disc space to promote fusion, benefiting patients with degenerative disc disease.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

PEEK (Polyetheretherketone) interbody fusion device with added tantalum markers for radiopacity. Available in various lengths and heights. Provided non-sterile.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Spineology PEEK Lumbar Interbody Fusion Devices ([K110933](/device/K110933.md))
- Spineology PEEK Lumbar Interbody Fusion Devices ([K111880](/device/K111880.md))
- Spineology PEEK Lumbar Interbody Fusion Devices ([K113030](/device/K113030.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

队 120293

page 1 of 2

# FEB 2 9 2012

## 510(k) Summary

| Applicant:                                      | Spineology Inc.<br>7800 3rd Street N., Suite 600<br>Saint Paul, MN 55128<br>651-256-8500                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                                 | Bryan Becker                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Date Prepared:                                  | January 18, 2012                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Trade Name:                                     | Spineology PEEK Lumbar Interbody Fusion Devices                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Product Classification<br>and Code:             | Class II Medical Device, Product Code MAX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Purpose of Application:                         | The purpose of this special 510(k) modification is to add tantalum<br>markers.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Predicate Device(s):                            | Spineology PEEK Lumbar Interbody Fusion Devices (K110933,<br>K111880, K113030)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Device Description:                             | The Spineology PEEK Lumbar Interbody Fusion Device is designed for<br>use with autograft to facilitate fusion and is intended for use with<br>supplemental fixation systems cleared for use in the lumbar spine. The<br>device is available in a range of lengths and heights. The device and<br>associated instruments are provided non-sterile.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Intended Use:                                   | Crescent: The Spineology PEEK Lumbar Interbody Fusion Device is an<br>intervertebral body fusion device indicated for intervertebral body fusion<br>at one level or two contiguous levels in the lumbar spine from L2 to S1 in<br>patients with degenerative disc disease (DDD) with up to Grade I<br>spondylolisthesis at the involved level(s). DDD is defined as back pain of<br>discogenic origin with degeneration of the disc confirmed by patient<br>history and radiographic studies. These patients should be skeletally<br>mature and have had six months of non-operative treatment.<br>These devices may be implanted via an open or a minimally invasive<br>transforaminal approach.<br>The Spineology PEEK Lumbar Interbody Fusion Device is designed for<br>use with autograft to facilitate fusion and is intended for use with<br>supplemental fixation systems cleared by the FDA for use in the lumbar<br>spine.<br>Bullet: The Spineology PEEK Bullet Lumbar Interbody Fusion Device is<br>an intervertebral body fusion device indicated for intervertebral body |
|                                                 | to S1 in patients with degenerative disc disease (DDD) with up to Grade I<br>spondylolisthesis at the involved level(s). DDD is defined as back pain of<br>discogenic origin with degeneration of the disc confirmed by patient<br>history and radiographic studies. These patients should be skeletally<br>mature and have had six months of non-operative treatment.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                                 | These devices may be implanted singly or in pairs via an open or a<br>minimally invasive posterior or transforaminal approach.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                                 | The Spineology PEEK Bullet Lumbar Interbody Fusion Device is<br>designed for use with autograft to facilitate fusion and is intended for use<br>with supplemental fixation systems cleared by the FDA for use in the<br>lumbar spine.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Summary of<br>Technological<br>Characteristics: | The device is shown to be substantially equivalent to the intended use,<br>materials, configuration, and performance characteristics of the predicate<br>products.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Testing                                         | No testing was required for this change. The results of a risk analysis<br>showed that the device was substantially equivalent to the identified<br>predicate devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Conclusion:                                     | The information submitted in this premarket notification supports a<br>determination that the Spineology PEEK Lumbar Interbody Fusion<br>Device is substantially equivalent in technological characteristics and<br>intended use to the predicate devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |

{1}------------------------------------------------

page 2 of 2.

.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a bold, sans-serif font. The text is black and the background is white. The text is centered in the image.

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of two main elements: the department's name encircling a symbol. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The symbol is a stylized representation of an abstract shape, resembling a person embracing or supporting another, with flowing lines suggesting movement or care.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Spineology Inc. % Mr. Bryan Becker Clinical and Regulatory Affairs Manager 7800 310 Street North, Suite 600 Saint Paul, Minnesota 55128

FEB 2 9 2012

Re: K120293

Trade/Device Name: PEEK Bullet Lumbar Interbody Fusion Device Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: January 25, 2012 Received: January 31, 2012

Dear Mr. Becker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

#### Page 2 - Mr. Bryan Becker

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark Haldane

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### Device Name: The Spineology PEEK Bullet Lumbar Interbody Fusion Device

#### Indications for Use:

K120293

Crescent: The Spineology PEEK Lumbar Interbody Fusion Device is an intervertebral body fusion device indicated for intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade I spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

These devices may be implanted via an open or a minimally invasive transforaminal approach.

The Spineology PEEK Lumbar Interbody Fusion Device is designed for use with autograft to facilitate fusion and is intended for use with supplemental fixation systems cleared by the FDA for use in the lumbar spine.

Bullets: The Spineology PEEK Bullet Lumbar Interbody Fusion Device is an intervertebral body fusion device indicated for intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade I spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

These devices may be implanted singly or in pairs via an open or a minimally invasive posterior or transforaminal approach.

The Spineology PEEK Bullet Lumbar Interbody Fusion Device is designed for use with autograft to facilitate fusion and is intended for use with supplemental fixation systems cleared by the FDA for use in the lumbar spine.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

| 510(k) Number | K120293 |
|---------------|---------|
|---------------|---------|

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K120293](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAX/K120293)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
