K120024 · Medacta International S.A. · MAX · Feb 28, 2012 · Orthopedic
Device Facts
Record ID
K120024
Device Name
MECTALIF TRANSFORAMINAL
Applicant
Medacta International S.A.
Product Code
MAX · Orthopedic
Decision Date
Feb 28, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The MectaLIF implants in combination with supplemental fixation are indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2 – S1 whose condition requires the use of interbody fusion. Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). Patients must be skeletally mature. Patients should have received 6 months of non-operative treatment prior to treatment with the devices.
Device Story
MectaLIF Transforaminal is an intervertebral body fusion device designed for TLIF procedures. Input: surgical placement into the intervertebral space to replace a degenerative disc. Device features a titanium gear mechanism allowing surgeons to adjust the implant angle in 15° increments in situ, facilitating repositioning without changing instrumentation. Output: restored spinal column height and stabilization. Used in clinical settings by surgeons; requires supplemental posterior fixation (e.g., pedicle screws) and autogenous bone graft. Benefits patients by providing structural support for fusion in DDD cases.
Clinical Evidence
Bench testing only. Performance evaluated via static and dynamic axial compression, static and dynamic compression/shear, and subsidence resistance testing according to ASTM F2077 and ASTM F2267 standards.
Technological Characteristics
Implant: PEEK-OPTIMA LT1 (ASTM F 2026). Gear: Ti6Al4V ELI titanium (ISO 5832-3/ASTM F 136). Markers: Tantalum (ISO 13782/ASTM F 560). Dimensions: 30-34mm length, 8-15mm height, 12-14mm width, 5° lordosis. Features an integrated titanium gear for in situ angle adjustment.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2–S1 requiring interbody fusion. DDD defined as discogenic pain with degeneration confirmed by history and radiography. Includes patients with prior non-fusion spinal surgery. Requires 6 months of prior non-operative treatment. Used with autogenous bone graft and supplemental posterior fixation.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
K212831 — MectaLIF Extension · Medacta International S.A. · Oct 29, 2021
K251016 — MectaLIF 3D Metal · Medacta International S.A. · Jul 17, 2025
K200551 — MectaLIF Transforaminal TiPEEK · Medacta Inernational SA · Feb 12, 2021
K181970 — MectaLIF Posterior Extension · Medacta International S.A. · Dec 4, 2018
K183705 — IdentiTi Porous Ti Interbody System · Alphatec Spine, Inc. · Mar 1, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
K120024
FEB 2 8 2012
Image /page/0/Picture/2 description: The image shows the logo for Medacta International. The logo consists of the word "Medacta" in a sans-serif font, with two triangles to the left of the word. Below the word is a horizontal line, and below that is the word "International" in a smaller font. To the right of the word "International" is a plus sign.
# 510(k) Summary
Applicant/Sponsor: Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
Contact Person: Mr. Adam Gross Director of Regulatory and Quality Medacta USA 4725 Calle Quetzal, Unit B Camarillo, CA, 93012 Phone: (805)437-7085 Fax: (805)437-7553 Email: AGross@medacta.us.com
Date Prepared: December 30, 2011
### DEVICE INFORMATION
Trade/Proprietary Name: MectaLIF Transforaminal Common Name: Intervertebral Body Fusion Device Classification Name: Intervertebral Fusion Device with Bone Graft, Lumbar
> 21 CFR 888.3080 Class II Device Product Codes: MAX
Predicate Devices: K110927 MectaLIF Oblique (Medacta Intl) K040536 Verte-Stack Boomerang (Medtronic) K072791 OPAL Spacer (Synthes) K081888 Dynamik (Spineart) K081917 Devex/Leopard (Depuy Spine) P960025 Lumbar IF Cage (Depuy Acromed) K103034 Apache TLIF (Genesys Spine)
MectaLIF Transforaminal 510(k)
Pg 1 of 3
Section 5 - Page 2 of 4
{1}------------------------------------------------
K12C024
# Product Description
MectaLIF Transforaminal lumbar intervertebral body fusion device is characterized by different sizes of PEEK-OPTIMA LT1 (Polyetheretherketone) implants that can be applied with a TLIF procedure (Transforaminal Lumbar Intervertebral Fusion). Mectal IF Transforaminal is used to replace a degenerative disc in order to restore the height of the spinal column structure. They are made of PEEK-OPTIMA LT1 and contain Tantalum Markers as well as a titanium gear which enables the surgeon to alter the angle of the MectaLIF Transforaminal in situ in 15° increments and to reposition during surgery without switching instrumentation. MectaLIF Transforaminal is intended to be used in combination with posterior fixation (e.g. Pedicle Screw System) as well as an autogenous bone graft. The dimensions of MectaLIF Transforaminal are within the following range: Length 30-34mm; Height 8-15mm; Width 12-14mm; Lordosis 5°. The materials of the components of MectaLIF Transforaminal are as follows: Implant: PEEK-OPTIMA LT1: Implant Grade Polyetheretherketone (ASTM F 2026), Gear: Titanium: Ti6Al4V ELI (ISO 5832-3/ASTM F 136), and Marker: Tantalum (ISO 13782 / ASTM F 560).
# Indications for Use
The MectaLIF implants in combination with supplemental fixation are indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2 – S1 whose condition requires the use of interbody fusion. Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). Patients must be skeletally mature. Patients should have received 6 months of non-operative treatment prior to treatment with the devices.
# Comparison to Predicate Devices
MectaLIF Transforaminal is substantially equivalent in terms of indications for use, material, design, and performance characteristics as the previously cleared K110927 MectaLIF Oblique (Medacta Intl), K040536 Verte-Stack Boomerang (Medtronic), K072791 OPAL Spacer (Synthes), K081888 Dynamik (Spineart), K081917 Devex/Leopard (Depuy Spine), P960025 Lumbar IF Cage (Depuy Acromed), and K103034 Apache TLIF (Genesys Spine).
{2}------------------------------------------------
K120024
# Performance Testing
MectaLIF Transforaminal has similar performance testing as the predicates in terms of:
Static Axial Compression - ASTM F2077 Dynamic Axial Compression - ASTM F2077 Static Compression/Shear - ASTM F2077 Dynamic Compression/Shear - ASTM F2077 Subsidence Resistance - ASTM F2267
Conclusion:
Based on the above information, MectaLIF Transforaminal can be considered as substantially equivalent to its predicate devices in terms of indications for use, material, design, and performance characteristics
Section 5 - Page 4 of 4
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Medacta International SA % Medacta USA Mr. Adam Gross Director of Regulatory and Quality 4725 Calle Quetzal, Unit B Camarillo, California 93012
FEB 2 8 2012
Re: K120024
Trade/Device Name: MectaLJF Transforaminal Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: January 31, 2012 Received: February 01, 2012
Dear Mr. Gross:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
### Page 2 - Mr. Adam Gross
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 70 http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
f.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm
Sincerely yours,
Mark N. Melkerso Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K120024
# Indications for Use
· 510(k) Number (if known):
Device Name: MectaLIF Transforaminal
Indications for Use:
The MectaLIF implants in combination with supplemental fixation are indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2 - S1 whose condition requires the use of interbody fusion. Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). Patients must be skeletally mature. Patients should have received 6 months of non-operative treatment prior to treatment with the devices.
| Prescription Use | x |
|------------------------|---|
| (21 CFR 801 Subpart D) | |
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) | |
|--------------------------------------------------------------------------|-------------------------|
| Concurrence of CDRH/Office of Device Evaluation (ODE) | |
| (Division Sign-Off) | |
| Division of Surgical, Orthopedic, and Restorative Devices | |
| 510(k) Number | K120024 |
| MectaLIF Transforaminal 510(k) | |
| December 30, 2011 | Section 4 - Page 2 of 2 |
| | Pg 1 of 1 |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.