K103034 · Genesys Spine · MAX · Dec 10, 2010 · Orthopedic
Device Facts
Record ID
K103034
Device Name
GENESYS SPINE INTERBODY FUSION SYSTEM
Applicant
Genesys Spine
Product Code
MAX · Orthopedic
Decision Date
Dec 10, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Intended Use
When used as a cervical intervertebral body fusion device, the Genesys Spine Interbody Fusion System is indicated for intervertebral body fusion in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from the C2-C3 disc to the C7-TI disc. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. Patients should have at least six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage. When used as a lumbar intervertebral body fusion device, the Genesys Spine Interbody Fusion System is indicated for intervertebral body fusion of the lumbar spine, from L2 to Sl, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. When used as a vertebral body replacement device, the Genesys Spine System is indicated for use to replace a vertebral body that has been resected or excised (i.e. partial or total vertebrectomy) due to tumor or trauma/fracture. The device system is intended for use in the thoracolumbar spine (from T1 to L5) and is intended for use with supplemental fixation. These devices are designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period. The device system is intended to be used with autograft or allograft bone.
Device Story
System comprises PEEK (ASTM F2026) interbody fusion and vertebral body replacement (VBR) implants with tantalum markers (ASTM F560) for radiographic visualization. Implants configured for various surgical approaches: anterior (cervical/lumbar), posterior (PLIF), or transforaminal (TLIF). Used by surgeons in clinical settings to restore spinal biomechanical integrity; facilitate fusion via autograft/allograft; replace resected vertebral bodies. Implantation performed using stainless steel (ASTM F899) instruments. Output is physical structural support for spinal column; aids clinical decision-making by providing stability in degenerative or traumatic conditions.
Clinical Evidence
Bench testing only. No clinical data provided. Mechanical testing included static and dynamic compression and torsion (ASTM F2077-03), subsidence (ASTM F2267-04), and expulsion (ASTM Draft Standard F-04.25.02.02).
Technological Characteristics
Materials: PEEK Optima LT1 (ASTM F2026) and tantalum markers (ASTM F560). Instruments: Stainless steel (ASTM F899). Form factor: Various configurations for cervical, lumbar, and VBR applications. No software or electronic components.
Indications for Use
Indicated for skeletally mature patients requiring intervertebral fusion for DDD (cervical C2-T1, lumbar L2-S1) or vertebral body replacement (thoracolumbar T1-L5) due to tumor or trauma/fracture. Cervical DDD requires 6 weeks non-operative treatment; lumbar DDD requires 6 months. Contraindications not explicitly listed.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
K150521 — Interbody Fusion (IBF)/Vertebral Body Replacement(VBR) System · Rti Surgical, Inc. Dba Rti Biologics · May 15, 2015
K110659 — MEDICAL CONCEPTS INTERBODY FUSION SYSTEM · Medical Concepts, Inc. · Sep 13, 2011
K170643 — Interbody Fusion (IBF)/Vertebral Body Replacement (VBR) System · Rti Surgical, Inc. Dba Rti Biologics · Apr 14, 2017
K151140 — LnK Lumbar Interbody Fusion Cage System · L&K BIOMED Co., Ltd. · Aug 18, 2015
K101812 — CONSTRUX MINI PEEK SPACER SYSTEM · Orthofix, Inc. · Sep 27, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
DEC 1 0 2010
# Genesys Spine Interbody Fusion System
# Premarket Notification
| SUBMITTED BY | Genesys Spine<br>1250 Capital of Texas Highway South<br>Building Three, Suite 600<br>Austin, TX 78746 |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ESTABLISHMENT<br>REGISTRATION NUMBER | Pending |
| OWNER/OPERATOR<br>NUMBER | Pending |
| CONTACT PERSON | Josh Kaufmann<br>Principal<br>Genesys Spine<br>Phone: 512-560-3446<br>Fax: 800-817-4938 |
| SUBMISSION PREPARED BY | Lisa Peterson<br>QA Consulting, Inc.<br>Phone: 512-507-0746 |
| DATE PREPARED | October 11, 2010 |
| CLASSIFICATION NAME | Intervertebral Fusion Device with Bone Graft, Lumbar<br>Intervertebral Fusion Device with Bone Graft, Cervical<br>Spinal Intervertebral Body Fixation Orthosis |
| DEVICE CLASS | Class II |
| REGULATION NUMBER | 888.3080 (Product Code: MAX)<br>888.3080 (Product Code: ODP)<br>888.3060 (Product Code: MQP) |
| COMMON NAME | Intervertebral Body Fusion Device (MAX, ODP)<br>Spinal Vertebral Body Replacement Device (MQP) |
| PROPRIETARY NAME | Genesys Spine Interbody Fusion System |
| IDENTIFICATION OF PREDICATE<br>DEVICE(S) | Predicate devices include various cleared interbody fusion<br>and VBR systems:<br>RAY® Threaded Fusion Cage (P950019)<br>Lumbar I/F Cage (P960025)<br>InterFix (P970015)<br>BAK/C (P980048)<br>Affinity Cage System (P000028)<br>MC+ (K043479, K091088) |
- MC+ (K043479, K091088)
Eminent Spine System (K090064) -
{1}------------------------------------------------
### DEVICE DESCRIPTION
The Genesvs Spine Interbody Fusion System will be offered in various device configurations based on surgical approach and patient anatomy, and consist of:
- 1) Genesys Spine cervical interbody fusion device(s), which may be implanted as a single device via an anterior approach.
- 2) Genesys Spine lumbar interbody fusion device(s), which may be implanted
- bi-laterally via a posterior (PLIF) approach:
- as a single device via a transforaminal (TLIF) approach; or ،
- as as a single device via an anterior/anterolateral (ALIF) approach. ﺑ
- Genesys Spine vertebral body replacement device(s), which may be implanted in the thoracic 3) and/or thoracolumbar spine (T1-L5).
The Genesys Spine Interbody Fusion System implant components are made of polyether ether ketone (PEEK Optima LT1) that conforms to ASTM F2026. Additionally, the devices contain tantalum markers (ASTM F560) to assist the surgeon with proper placement of the device.
The Genesys Spine Interbody Fusion System is implanted using a combination of device specific and universal class I instruments manufactured from stainless steel materials that conform to ASTM F899.
#### INDICATIONS
When used as a cervical intervertebral body fusion device, the Genesys Spine Interbody Fusion System is indicated for intervertebral body fusion in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from the C2-C3 disc to the C7-TI disc. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. Patients should have at least six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage.
When used as a lumbar intervertebral body fusion device, the Genesys Spine Interbody Fusion System is indicated for intervertebral body fusion of the lumbar spine, from L2 to SI, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion.
When used as a vertebral body replacement device, the Genesys Spine System is indicated for use to replace a vertebral body that has been resected or excised (i.e. partial or total vertebrectomy) due to tumor or trauma/fracture. The device system is intended for use in the thoracolumbar spine (from T1 to L5) and is intended for use with supplemental fixation. These devices are designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period. The device system is intended to be used with autograft or allograft bone.
{2}------------------------------------------------
#### SUMMARY OF TECHNOLOGICAL CHARACTERISTICS
The purpose of this premarket notification is to obtain clearance to market the Genesys Spine Interbody Fusion System. The Genesys Spine Interbody Fusion System is comprised of various device configurations designed to accommodate patient anatomy and provide the surgeon with different surgical approach options.
The Genesys Spine Interbody Fusion System implant components are made of polyether ether ketone (PEEK Optima LT1) that conforms to ASTM F2026. Additionally, the devices contain tantalum markers (ASTM F560) to assist the surgeon with proper placement of the device.
The subject system has similar technological characteristics as the predicate devices identified above. Specifically, the following characteristics support this conclusion:
- Intended for use at one level from the C2-C3 disc to the C7-TI disc for the treatment । degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at the involved level.
- -Intended for use at either one level or two contiquous levels from L2-S1 for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis at the involved level(s).
- -When used as a VBR, the implants are designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column (T1-L5) even in the absence of fusion for a prolonged period.
- Substantially equivalent results of non-clinical testing relative to static and dynamic testing (per ASTM F2077-03), subsidence (per ASTM F2267-04), and expulsion (per ASTM Draft Standard F-04.25.02.02)
#### DISCUSSION OF NON-CLINICAL TESTING
The following non-clinical tests were conducted:
- Static and dynamic compression testing, conducted in accordance with ASTM F2077-03 -
- Static and dynamic torsion testing, conducted in accordance with ASTM F2077-03 ﺘ
- Subsidence testing, conducted in accordance with ASTM F2267-04 ﻬﺎ ﻓﻬﻮ: ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘ
- Expulsion testing, conducted in accordance with ASTM Draft Standard F-04.25.02.02 -
#### CONCLUSIONS
The subiect and predicate device(s) share the same intended use, primary implant design and equivalent material of manufacture. The non-clinical mechanical test results demonstrate that any minor differences do not impact device performance as compared to the predicates and demonstrate that the Genesys Spine Interbody Fusion System is substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Genesys Spine % Mr. Josh Kaufmann 1250 Capital of Texas Highway South Building Three, Suite 600 Austin, Texas 78746
DEC 1 0 2010
Re: K103034
Trade/Device Name: Genesys Spine Interbody Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX, ODP, MQP Dated: October 11, 2010 Received: October 13, 2010
Dear Mr. Kaufmann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
Page 2 – Mr. Josh Kaufmann
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. __
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Arj. P. h. ta
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### INDICATIONS FOR USE
510(k) Number (if known):
#### Genesys Spine Interbody Fusion System Device Name:
Indications for Use:
When used as a cervical intervertebral body fusion device, the Genesys Spine Interbody Fusion System is indicated for intervertebral body fusion in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from the C2-C3 disc to the C7-TI disc. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. Patients should have at least six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage.
When used as a lumbar intervertebral body fusion device, the Genesys Spine Interbody Fusion System is indicated for intervertebral body fusion of the lumbar spine, from L2 to Sl, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion.
When used as a vertebral body replacement device, the Genesys Spine System is indicated for use to replace a vertebral body that has been resected or excised (i.e. partial or total vertebrectomy) due to tumor or trauma/fracture. The device system is intended for use in the thoracolumbar spine (from T1 to L5) and is intended for use with supplemental fixation. These devices are designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column even in the absence of fusion for a prolonged period. The device system is intended to be used with autograft or allograft bone.
× Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Genesys Spine
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
| | 510(k) Number | K 103034<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
|--|---------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | |
1250 Capital of Texas Highway South Building Three, Suite 600 Austin, Texas 78746
Phone: 512-560-3446 Fax: 800-817-4938
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.