PIVOTEC LUMBAR INTERBODY FUSION DEVICE (LIFD)

K092017 · Captiva Spine, Inc. · MAX · Dec 1, 2009 · Orthopedic

Device Facts

Record IDK092017
Device NamePIVOTEC LUMBAR INTERBODY FUSION DEVICE (LIFD)
ApplicantCaptiva Spine, Inc.
Product CodeMAX · Orthopedic
Decision DateDec 1, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Pivotec LIFD implant is an intervertebral body fusion device indicated for intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. The Pivotec LIFD Implant is designed for use with autograft to facilitate fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine.

Device Story

Pivotec LIFD is an intervertebral body fusion device; implanted in lumbar disc space to stabilize spinal segments and promote fusion. Device features hollow core for autogenous bone graft; manufactured from PEEK-Optima with Tantalum markers. Implanted by surgeons using specialized insertion instruments for manipulation within disc space. Used in conjunction with supplemental fixation systems. Benefits include spinal stabilization and fusion facilitation in patients with DDD.

Clinical Evidence

Bench testing only. Static and dynamic compression, compression shear, and subsidence testing conducted per ASTM F2077-03 and ASTM F2267-04. Device met or exceeded all specifications.

Technological Characteristics

Material: PEEK-Optima with Tantalum markers. Form factor: Hollow core implant in various configurations/heights. Mechanical testing standards: ASTM F2077-03, ASTM F2267-04. Non-active, mechanical device.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) and up to Grade 1 spondylolisthesis at one or two contiguous lumbar levels (L2-S1) who have failed six months of non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ く ## 510(k) Summary 心 · . . ... ... | Submitter: | Captiva Spine, Inc. | |------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Mr. Dale Mitchell, President<br>Captiva Spine, Inc.<br>967 Alt. AlA #1<br>Jupiter, FL 33477<br>Phone: 877-772-5571<br>Fax: 866-318-3224 | | | DEC - 1 2009 | | Date Prepared: | July 2, 2009 | | Trade Name: | Pivotec Lumbar Interbody Fusion Device (LIFD) | | Classification<br>Name and Number: | Class II<br>Intervertebral Body Fusion Device<br>21 CFR 888.3080 | | Product Code: | MAX | | Predicate Device(s): | The subject device is substantially equivalent to the following<br>devices:<br>Lumbar I/F Cage (P960025)<br>Marketed and distributed by DePuy<br>Ray Threaded Fusion Cage (TFC)TM (P950019)<br>Marketed and distributed by Stryker (originally approved by<br>Surgical Dynamics)<br>AVS TL PEEK Spacer (K083661)<br>Marketed and distributed by Stryker | | Device Description: | The Pivotec LIFD is an intervertebral body fusion device for use<br>with autogenous bone graft in the intervertebral disc space to<br>stabilize spinal segments and promote fusion. The device is made<br>of PEEK-Optima with Tantalum markers and is provided in<br>various configurations and heights, containing a hollow core to<br>receive bone autograft. Placement is achieved with an insertion<br>instrument that allows for manipulation of the implant within the<br>intra-vertebral disc space. | 2 {2}------------------------------------------------ Intended Use: The Pivotec LIFD implant is an intervertebral body fusion device indicated for intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. The Pivotec LIFD Implant is designed for use with autograft to facilitate fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine. Mechanical testing of the subject device consisted of static and Functional and dynamic compression, static and dynamic compression shear and Safety Testing: subsidence. All testing was conducted in accordance with ASTM F2077-03 and ASTM F2267-04. The device performed as designed and met, or exceeded, all product specifications. Conclusion: The documentation provided demonstrates that the Pivotec LIFD is substantially equivalent to the predicate devices listed above. This conclusion is based on the devices' similarities in materials, design, indications for use, function and mechanical function. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with its wings spread, composed of three curved lines that suggest movement or flight. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002 Captiva Spine, Inc. % Sanacor, LLC Mr. Mike Ensign Director of Engineering P.O. Box 1196 Pleasant Grove, Utah 84062 DEC - 1 2009 Re: K092017 Trade/Device Name: Pivotec Lumbar Interbody Fusion Device (LIFD) Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: II Product Code: MAX Dated: July 2, 2009 Received: July 6, 2009 Dear Mr. Ensign: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ ## Page 2 - Mr. Mike Ensign If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Sincerely yours, Mark N. Mulkern Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known): NA Device Name: Pivotec Lumbar Interbody Fusion Device (LIFD) Indications for Use: The Pivotec LIFD implant is an intervertebral body fusion deviceindicated for intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. The Pivotec LIFD Implant is designed for use with autograft to facilitate fusion and is intended for use with supplemental fixation systems cleared for use in the lumbar spine. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Division Sign Off (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K092017
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...