DIANA FUSION CAGE

K091122 · Signus Medical · MAX · Sep 18, 2009 · Orthopedic

Device Facts

Record IDK091122
Device NameDIANA FUSION CAGE
ApplicantSignus Medical
Product CodeMAX · Orthopedic
Decision DateSep 18, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The DIANA™ Fusion Cage is intended for interbody fusion procedures and is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. The device is intended to be used with supplemental fixation. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be used with autograft bone.

Device Story

DIANA™ Fusion Cage is an interbody fusion implant used in spinal surgery. It consists of titanium alloy (Ti6Al4V-ELI) cages of various sizes to match patient anatomy. The device is implanted by a surgeon during interbody fusion procedures to stabilize the spine at L2-S1 levels. It functions as a structural spacer to facilitate fusion when used with autograft bone and supplemental fixation. The device benefits patients by providing mechanical support for spinal segments affected by degenerative disc disease, potentially reducing discogenic back pain.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Material: Medical grade titanium alloy (Ti6Al4V-ELI). Form factor: Interbody fusion cage with various sizes. Energy source: None (mechanical implant). Connectivity: None. Sterilization: Not specified.

Indications for Use

Indicated for skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1, including those with up to Grade I spondylolisthesis or retrolisthesis. Requires 6 months of prior non-operative treatment. Used with supplemental fixation and autograft bone.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K091122 ## SEP 1 8 2009 # 4. 510(k) Summary according to 807.92(c) Contact: Thomas Hoghaug Signus Medical 18888 Lake Drive East Chanhassen, MN 55317 952-294-8700 Trade Name: Product Class: Classification: Product Codes: Panel Code: DIANATM Fusion Cage Class II 21 CFR §888.3080 Orthosis, intervertebral fusion MAX 87 #### Indications for Use: The DIANA™ Fusion Cage is intended for interbody fusion procedures and is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the device is intended to be used with supplemental fixation. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be used with autograft bone. #### Device Description: The DIANA™ Fusion Cage is comprised of a variety of implant sizes to accommodate various patient anatomy and pathology, and associated instrumentation. All implantable components are manufactured from medical grade titanium alloy (Ti6Al4V-ELI). #### Predicate Device(s): The DIANA™ Fusion Cage was shown to be substantially equivalent to previously cleared devices and has the same indications for use, design, function, and materials used. . The four formerly Class III predicate devices include the BAK Interbody Fusion Device (Spine-Tech, P950002), Inter Fix Threaded Fusion Device (Sofamor Danek, P970015), the Ray Threaded Fusion Cage (Surgical Dynamics, P950019) and the Brantigan Cage (P960025). #### Performance Testing: The pre-clinical testing performed indicated that the DIANA™ Fusion Cage is substantially equivalent to the predicate devices and is adequate for the intended use. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping human figures or abstract shapes. The symbol is positioned to the right of the text. Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002 ### SEP 1 8 2009 Signus Medical % MCRAPPD Medical Device Rich Jansen, Pharm.D. 1331 H Street Northwest, Suite 1200 Washington, District of Columbia 20005 Re: K091122 Trade/Device Name: The DIANA™ Fusion Cage Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: July 27, 2009 Received: August 5, 2009 Dear Mr. Jansen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {2}------------------------------------------------ Page 2 - Rich Jansen, Pharm.D. forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrl/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Harbare Buckins Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## 3. Statement of Indications for Use 510(k) Number (if known): _KO91122 Indications for Use: The DIANA™ Fusion Cage is intended for interbody fusion procedures and is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. The device is intended to be used with supplemental fixation. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be used with autograft bone. 1 Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Kareem S. Roney for MXM (Division Sign-Off) Division of Stagical, Orthopedic, and Restorative Devices 510(k) Number K091122 Concurrence of CDRH, Office of Device Evaluation (ODE)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...