The CAPSTONE® Spinal System is indicated for interbody fusion with autogenous bone graft in patients with degenerative disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implanted via an open or a minimally invasive posterior approach. Alternatively, these implanted via an anterior and/or transforaminal approach. These implants are to be used with autogenous bone graft. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the FDA for use in the lumbar spine.
Device Story
CAPSTONE® Spinal System consists of PEEK and titanium alloy intervertebral body fusion cages; designed for lumbar or lumbosacral interbody fusion; hollow geometry allows packing with autogenous bone graft; implanted via open, minimally invasive posterior, anterior, or transforaminal approaches; used in conjunction with FDA-cleared supplemental spinal fixation instrumentation; provides structural support and correction during fusion surgery; intended for use by surgeons in clinical settings.
Clinical Evidence
Bench testing only.
Technological Characteristics
Intervertebral body fusion cages; materials: PEEK and titanium alloy; geometry: hollow for bone graft containment; dimensions: various widths and heights; intended for use with supplemental fixation; non-active device.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1, including those with up to Grade 1 Spondylolisthesis or retrolisthesis, who have failed six months of non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
K133650 — CAPSTONE SPINAL SYSTEM · Medtronic Sofamor Danek USA, Inc. · Dec 20, 2013
K082732 — MODIFICATION TO: CAPSTONE SPINAL SYSTEM · Medtronic Sofamor Danek USA, Inc. · Oct 16, 2008
K103731 — CAPSTONE R SPINAL SYSTEM · Medtronic Sofamor Danek · Jul 18, 2011
K123027 — CAPSTONE(R) SPINAL SYSTEM · Medtronic Sofamor Danek, Inc. · Jul 25, 2013
K121760 — CAPSTONE SPINAL SYSTEM · Medtronic Sofamor Danek USA, Inc. · Aug 29, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
# CAPSTONE® Spinal System 510(k) Summarv - K082342 September 2008
| I. | Company: | Medtronic Sofamor Danek<br>1800 Pyramid Place<br>Memphis, TN 38132<br>(901) 396-3133 | SEP 1 2 2008 |
|-----|-------------------------------------------------|--------------------------------------------------------------------------------------|--------------|
| | Contact: | Chris McKee<br>Sr. Regulatory Affairs Specialist | |
| II. | Proprietary Trade Name: CAPSTONE® Spinal System | | |
III. Classification Name: Intervertebral Body Fusion Device (21 CFR 888,3080)
### IV. Product Code: MAX
### V. Product Description
The CAPSTONE® Spinal System consists of PEEK cages and titanium alloy cages of various widths and heights, which can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. The hollow geometry of the implants allows them to be packed with autogenous bone graft,
### VI. Indications
The CAPSTONE® Spinal System is indicated for interbody fusion with autogenous bone graft in patients with degenerative disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via an open or a minimally invasive posterior approach. Alternatively, these implants may also be implanted via an anterior and/or transforaminal approach. These implants are to be used with autogenous bone graft. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the FDA for use in the lumbar spine.
#### VII. Substantial Equivalence
Documentation, including mechanical test result, was provided which demonstrated that the subject CAPSTONE® Spinal System devices are substantially equivalent to the predicate CAPSTONE® Spinal System devices (K073291 SE 04/24/08) as well as the VERTE-STACK® Spinal System (K043566 SE 01/07/05, K043561 SE 12/29/04).
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA", which is arranged in a circular fashion around the logo.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Medtronic Sofamor Danek % Mr. Chris McKee Sr. Regulatory Affairs Specialist 1800 Pyramid Place Memphis, Tennessee 38132
Re: K082342
> Trade/Device Name: CAPSTONE® Spinal System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: August 13, 2008 Received: August 15, 2008
Dear Mr. McKee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
SEP 1 2 2008
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 - Mr. Chris McKee
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark W. Millerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known): K082342
Device Name: CAPSTONE® Spinal System
Indications for Use:
The CAPSTONE® Spinal System is indicated for interbody fusion with autogenous bone graft in patients with degenerative disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These implanted via an open or a minimally invasive posterior approach. Alternatively, these implanted via an anterior and/or transforaminal approach. These implants are to be used with autogenous bone graft. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the FDA for use in the lumbar spine.
Prescription Use x x OR Over-The-Counter Use
Per 21 CFR 801.109
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE
Mark M. Milliken
K082342
( Division Sign-on) Restorative, and Neurological Dev
**510(k) Number**
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.