Arthrex SwiveLock Anchor

K192532 · Arthrex, Inc. · MAI · Oct 19, 2019 · Orthopedic

Device Facts

Record IDK192532
Device NameArthrex SwiveLock Anchor
ApplicantArthrex, Inc.
Product CodeMAI · Orthopedic
Decision DateOct 19, 2019
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Arthrex SwiveLock Anchor is intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures: • Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction • Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus reconstruction, Mid-foot reconstruction, Metatarsal Ligament Repair, Tendon Repair, Bunionectomy • Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis • Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction • Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis Repair • Hip: Capsular Repair, Acetabular labral repair

Device Story

Sterile two-component suture anchor; consists of eyelet and hollow anchor body; pre-mounted on driver with components separated on shaft. Used by surgeons in clinical settings for soft tissue-to-bone fixation. Operates by securing suture to bone to facilitate tissue healing. Output is mechanical fixation of soft tissue. Benefits include stable reattachment of tendons/ligaments to bone, supporting surgical repair of various orthopedic injuries.

Clinical Evidence

Bench testing only. Mechanical testing evaluated pull-out (tensile) strength against predicate criteria. Geometric analysis and load vs. degradation time comparisons performed. Bacterial endotoxin testing conducted per USP <161>, USP <85>, and EP 2.6.14 to ensure compliance with pyrogen limit specifications.

Technological Characteristics

Two-component suture anchor (eyelet and hollow body). Materials are BioComposite, consistent with predicate. Mechanical fixation principle. Sterile, single-use device. No software or electronic components.

Indications for Use

Indicated for fixation of soft tissue to bone in shoulder, foot/ankle, knee, hand/wrist, elbow, and hip procedures in patients requiring ligament or tendon repair/reconstruction.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Arthrex Inc. Ivette Galmez Senior Regulatory Affairs Specialist 1370 Creekside Boulevard Naples, Florida 34108-1945 Re: K192532 Trade/Device Name: Arthrex SwiveLock Anchor Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: MAI, HWC Dated: September 12, 2019 Received: September 16, 2019 Dear Ms. Galmez: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Laurence D. Coyne, Ph.D. Acting Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) Device Name Arthrex SwiveLock Anchor #### Indications for Use (Describe) The Arthrex SwiveLock Anchor is intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in the following procedures: • Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction · Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus reconstruction, Mid-foot reconstruction, Metatarsal Ligament Repair, Tendon Repair, Bunionectomy • Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis - · Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction - · Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondy.itis Repair · Hip: Capsular Repair, Acetabular labral repair Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) | ☑ | |----------------------------------------------|---| | Over-The-Counter Use (21 CFR 801 Subpart C) | ☐ | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) Summary 1. | Date Prepared | October 8, 2019 | |-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter | Arthrex Inc.<br>1370 Creekside Boulevard<br>Naples, FL 34108-1945 | | Contact Person | Ivette Galmez<br>Senior Regulatory Affairs Specialist<br>1-239-643-5553, ext. 71263<br>ivette.galmez@arthrex.com | | Name of Device | Arthrex SwiveLock Anchor | | Common Name | Suture Anchor | | Product Code | MAI, HWC | | Classification Name | 21 CFR 888.3030: Single/multiple component metallic bone fixation appliances and<br>accessories<br>21 CFR 888.3040: Smooth or threaded metallic bone fixation fastener | | Regulatory Class | Class II | | Predicate Device | K101823: Arthrex SwiveLock Anchors | | Reference Device | K173788: Arthrex Corkscrew FT<br>K180594: Knotless SutureTak Anchors<br>K082810: BioComposite Suture Anchors (Hip)<br>K101679: Arthrex PushLock | | Purpose of Submission | This Special 510(k) premarket notification is submitted to obtain clearance for the<br>3.9mm BioComposite SwiveLock suture anchor made of the same materials as the<br>predicate. | | Device Description | The Arthrex SwiveLock Anchor is a sterile two-component suture anchor comprised of<br>an eyelet and a hollow anchor body. The Arthrex SwiveLock Anchor is pre-mounted on<br>a driver with the anchor body and eyelet physically separated on the driver shaft.<br>Arthrex 510(k) cleared suture may also be provided with the device. | | Indications for Use | The Arthrex SwiveLock Anchor is intended for fixation of suture (soft tissue) to bone in<br>the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip the following procedures:<br>• Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps<br>Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or<br>Capsulolabral Reconstruction<br>• Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux<br>Valgus Reconstruction, Mid-foot Reconstruction, Metatarsal Ligament<br>Repair/Tendon Repair, Bunionectomy<br>• Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar<br>Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis<br>• Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral<br>Ligament reconstruction.<br>• Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament<br>Reconstruction, Lateral Epicondylitis Repair (Tennis elbow)<br>• Hip: Capsular Repair, Acetabular labral repair | | Performance Data | Mechanical testing demonstrated that the pull-out (tensile) strength of the proposed<br>Arthrex SwiveLock Anchor met the criteria established by the predicate device and that<br>the anchor was successfully inserted into the bone substitute. Geometric analysis and<br>Load vs Degradation Time comparison was conducted to demonstrate that the<br>expected degradation and pull out strength of the proposed anchor is not a new worst<br>case over the predicate and referenced predicate devices.<br>Bacterial endotoxin per USP <161>, USP <85>, FDA Guidance for Industry Pyrogen &<br>Endotoxin Testing and EP 2.6.14 was conducted to demonstrate that the device meets<br>pyrogen limit specifications. | | Conclusion | The Arthrex SwiveLock Anchor is substantially equivalent to the predicate device in<br>which the basic design features and intended uses are the same. Any differences<br>between the proposed device and the predicate device is considered minor and do not<br>raise questions concerning safety or effectiveness. | | | Based on the indications for use, technological characteristics, and the summary of data<br>submitted, Arthrex Inc. has determined that the proposed device is substantially<br>equivalent to the currently marketed predicate devices. | {4}------------------------------------------------
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...