← Product Code [MAI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAI) · K191226

# Arthrex SwiveLock Suture Anchor (K191226)

_Arthrex, Inc. · MAI · Jun 11, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAI/K191226

## Device Facts

- **Applicant:** Arthrex, Inc.
- **Product Code:** [MAI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAI.md)
- **Decision Date:** Jun 11, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Arthrex SwiveLock Suture Anchors are intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ ankle, knee, hand/wrist, elbow, and hip in the following procedures: Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction. . Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair/Tendon Repair and Bunionectomy. . Knee: Anterior Cruciate Ligament Repair (4.75-5.5 SwiveLock Only), Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Illiotibial Band Tenodesis, and Quadriceps Tendon Repair. Secondary or adjunct fixation for ACL/PCL reconstruction or repair (4.75-5.5 SwiveLock only). - . Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction. . Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis repair. . Hip: Capsular repair, acetabular labral repair.

## Device Story

Arthrex SwiveLock Suture Anchor is a two-component, knotless bone fixation device consisting of an eyelet and a hollow anchor body. Premounted on a driver, the device secures soft tissue to bone during orthopedic procedures. Used by surgeons in clinical settings, the anchor facilitates tissue reattachment or ligament repair. The device is designed to provide secure fixation, potentially improving patient recovery outcomes in various joint surgeries. No software or electronic components are involved.

## Clinical Evidence

Bench testing only. Cyclic pull-out testing performed after 16 weeks of degradation demonstrated performance statistically equivalent to the predicate device (K061863). Bacterial endotoxin testing conducted per EP 2.6.14/USP <85> to confirm compliance with pyrogen limit specifications.

## Technological Characteristics

Two-component, knotless suture anchor system comprising an eyelet and hollow anchor body. Premounted on a driver. Materials are metallic bone fixation components. No software, electronics, or energy sources.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Arthrex SwiveLock Anchors ([K180768](/device/K180768.md))
- Arthrex Corkscrew ([K061863](/device/K061863.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a solid blue square and the agency name in a lighter blue. The text reads "FDA U.S. FOOD & DRUG ADMINISTRATION".

Arthrex Inc. David L. Rogers Manager, Regulatory Affairs 1370 Creekside Boulevard Naples, Florida 34108-1945

Re: K191226

Trade/Device Name: Arthrex SwiveLock Suture Anchor Regulation Number: 21 CFR 888,3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: MAI, MBI Dated: June 10, 2019 Received: June 11, 2019

Dear Mr. Rogers:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Laurence D. Coyne, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

#### K191226

Device Name Arthrex SwiveLock Suture Anchor

#### Indications for Use (Describe)

The Arthrex SwiveLock Suture Anchors are intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ ankle, knee, hand/wrist, elbow, and hip in the following procedures:

Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.

. Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair/Tendon Repair and Bunionectomy.

. Knee: Anterior Cruciate Ligament Repair (4.75-5.5 SwiveLock Only), Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Illiotibial Band Tenodesis, and Quadriceps Tendon Repair. Secondary or adjunct fixation for ACL/PCL reconstruction or repair (4.75-5.5 SwiveLock only).

- . Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction.
. Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis repair.

. Hip: Capsular repair, acetabular labral repair.

| Type of Use (Select one or both, as applicable)                                                        |                                             |
|--------------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: underline;"><b>X</b></span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary or 510(k) Statement

| Date Prepared            | May 3, 2019                                                                                                                                                                                                                                                                                                                                                              |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter                | Arthrex Inc.                                                                                                                                                                                                                                                                                                                                                             |
|                          | 1370 Creekside Boulevard                                                                                                                                                                                                                                                                                                                                                 |
|                          | Naples, FL 34108-1945                                                                                                                                                                                                                                                                                                                                                    |
| Contact Person           | David L Rogers                                                                                                                                                                                                                                                                                                                                                           |
|                          | Manager, Regulatory Affairs                                                                                                                                                                                                                                                                                                                                              |
|                          | 1-239-598-4302, ext. 71924                                                                                                                                                                                                                                                                                                                                               |
|                          | David.rogers@arthrex.com                                                                                                                                                                                                                                                                                                                                                 |
| Name of Device           | Arthrex SwiveLock Suture Anchor                                                                                                                                                                                                                                                                                                                                          |
| Common Name              | Suture Anchor                                                                                                                                                                                                                                                                                                                                                            |
| Product Code             | MAI, MBI                                                                                                                                                                                                                                                                                                                                                                 |
| Classification Name      | 21 CFR 888.3040: Smooth or threaded metallic bone fixation fastener                                                                                                                                                                                                                                                                                                      |
|                          | 21 CFR 888.3030: Single/multiple component metallic bone fixation appliances<br>and accessories                                                                                                                                                                                                                                                                          |
| Regulatory Class         | II                                                                                                                                                                                                                                                                                                                                                                       |
| Predicate Device         | K180768: Arthrex SwiveLock Anchors                                                                                                                                                                                                                                                                                                                                       |
|                          | K061863: Arthrex Corkscrew                                                                                                                                                                                                                                                                                                                                               |
| Purpose of<br>Submission | This traditional 510(k) premarket notification is submitted to obtain anterior<br>cruciate ligament repair indications for the Arthrex SwiveLock Anchors cleared<br>under K180768, K171141, and K151342.                                                                                                                                                                 |
| Device Description       | The Arthrex SwiveLock Anchor is a two-component, knotless suture anchor<br>comprised of an eyelet and a hollow anchor body. The SwiveLock Anchor is<br>premounted on a driver with the anchor body and eyelet physically separated on<br>the driver shaft. Arthrex 510(k) cleared suture may also be provided with the<br>device.                                        |
| Indications for Use      | The Arthrex SwiveLock Suture Anchors are intended for fixation of suture (soft<br>tissue) to bone in the shoulder, foot/ankle, knee, hand/wrist, elbow, and hip in<br>the following procedures:                                                                                                                                                                          |
|                          | Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair,<br>Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair,<br>Capsular Shift or Capsulolabral Reconstruction.                                                                                                                                                                         |
|                          | Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon<br>Repair, Hallux Valgus Reconstruction, Mid-foot Reconstruction,<br>Metatarsal Ligament Repair/Tendon Repair and Bunionectomy.                                                                                                                                                                 |
|                          | Knee: Anterior Cruciate Ligament Repair (4.75-5.5 SwiveLock Only),<br>Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair,<br>Patellar Tendon Repair, Posterior Oblique Ligament Repair, Illiotibial<br>Band Tenodesis, and Quadriceps Tendon Repair. Secondary or adjunct<br>fixation for ACL/PCL reconstruction or repair (4.75-5.5 SwiveLock only). |
|                          | Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial<br>Collateral Ligament Reconstruction.                                                                                                                                                                                                                                                                 |
|                          | Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or<br>Radial Collateral Ligament Reconstruction, Lateral Epicondylitis repair.                                                                                                                                                                                                                             |
|                          | Hip: Capsular repair, acetabular labral repair.                                                                                                                                                                                                                                                                                                                          |
| Performance Data         | Cyclic pull-out testing after 16 weeks of degradation was conducted on the<br>Arthrex SwiveLock Suture Anchor. Results demonstrate that the Arthrex                                                                                                                                                                                                                      |
|                          |                                                                                                                                                                                                                                                                                                                                                                          |
| Conclusion               | SwiveLock Anchor performs statistically equivalent to the predicate device<br>(K061863 – Arthrex Corkcrew).<br>Bacterial endotoxin per EP 2.6.14/USP <85> was conducted to demonstrate that<br>the device meets pyrogen limit specifications.                                                                                                                            |
|                          | The Arthrex SwiveLock Anchor is substantially equivalent to the predicate device<br>in which the basic design features and intended uses are the same. Any<br>differences between the proposed device and the predicate device are<br>considered minor and do not raise questions concerning safety or effectiveness.                                                    |
|                          | Based on the indications for use, technological characteristics, and the summary<br>of data submitted, Arthrex Inc. has determined that the proposed device is<br>substantially equivalent to the currently marketed predicate devices.                                                                                                                                  |

{4}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAI/K191226](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MAI/K191226)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
