S-ROM ZIRCONIA CERAMIC FEMORAL HIP HEAD

K980081 · Johnson & Johnson Professionals, Inc. · LZO · Apr 9, 1998 · Orthopedic

Device Facts

Record IDK980081
Device NameS-ROM ZIRCONIA CERAMIC FEMORAL HIP HEAD
ApplicantJohnson & Johnson Professionals, Inc.
Product CodeLZO · Orthopedic
Decision DateApr 9, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3353
Device ClassClass 2
AttributesTherapeutic

Intended Use

The S-ROM Zirconia Ceramic Femoral Head is indicated for use in total hip replacement procedures for patients suffering severe pain and disability due to structural damage in the hip joint from rheumatoid arthritis, osteoarthritis, posttraumatic arthritis, collagen disorders, avascular necrosis, and non-union of femoral fractures. Use of the prosthesis is also indicated for revision of previous hip arthroplasty and for patients with congenital hip dysplasia, protrusio acetabuli, slipped femoral epiphysis, and disability due to previous fusion.

Device Story

S-ROM Zirconia Ceramic Femoral Head; prosthetic component for total hip replacement. Manufactured from Yttrium Stabilized Zirconium Oxide (PROZYR or ZYRANOX). Designed for use with 11/13 Morse-taper femoral stems composed of titanium or cobalt-chromium-molybdenum alloy. Used in surgical settings by orthopedic surgeons. Functions as articulating surface in hip arthroplasty; must be paired with UHMWPE or metal-backed UHMWPE acetabular cups. Provides structural replacement for damaged hip joints; restores mobility; reduces pain. Subject device extends compatibility of existing ceramic head design to include cobalt-chromium-molybdenum alloy stems.

Clinical Evidence

No clinical data provided. Substantial equivalence based on identical design and material specifications to previously cleared predicate devices.

Technological Characteristics

Material: Yttrium Stabilized Zirconium Oxide (PROZYR or ZYRANOX). Interface: 11/13 Morse-taper. Compatibility: Titanium or cobalt-chromium-molybdenum alloy femoral stems; UHMWPE or metal-backed UHMWPE acetabular cups. Form factor: Femoral head for hip prosthesis.

Indications for Use

Indicated for total hip replacement in patients with severe pain/disability due to structural hip joint damage (rheumatoid arthritis, osteoarthritis, posttraumatic arthritis, collagen disorders, avascular necrosis, non-union of femoral fractures, congenital hip dysplasia, protrusio acetabuli, slipped femoral epiphysis, or previous fusion) and for revision of previous hip arthroplasty. Contraindicated for use with acetabular components other than UHMWPE or metal-backed UHMWPE cups.

Regulatory Classification

Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Ka80081 Summary of Safety and Effectiveness Data for the S-ROM® Zirconia Ceramic Femoral Head Use with Cobalt-Chromium-Molybdenum Alloy 11/13 Morse-Taper Femoral Stems APR - 9 1998 # Johnson & Johnson Professional, Inc. 325 Paramount Drive Raynham, MA 02767-0350 #### Contact Person________________________________________________________________________________________________________________________________________________________________ Mary E. Gray Associate Regulatory Affairs Specialist Phone: (508) 828-3545 (508) 828-3212 Fax: ### Name of Device | Proprietary Name: | S-ROM Zirconia Ceramic Femoral Head | |----------------------|-----------------------------------------------------------------------------------------------| | Common Name: | Hip Prosthesis | | Classification Name: | Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis. | | Regulatory Class: | Class II by 21 CFR 888.3353 | | Product Code: | 87 LZO | | Owner/Operator No.: | 9001269 | #### Device Classification ________________________________________________________________________________________________________________________________________________________ This device has been placed in Class II for Hip joint metal/ ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis per 21 CFR & 888.3353. ### Statement of Substantial Equivalence__________________________________________________________________________________________________________________________________________ The S-ROM Zirconia Ceramic Femoral Head is identical in design (drawing specifications) to the S-ROM Zirconia Ceramic Femoral Head cleared for marketing under premarket notification K973307 (November 20, 1997) for use with titanium alloy femoral components. The S-ROM Zirconia Ceramic Femoral Head is identical in material (PROZYR® or ZYRANOX™) to the S-ROM Zirconia Ceramic Femoral Head cleared for marketing under premarket notifications K973307 and K921111. The subject device is to be utilized with the same 11/13 Morse-taper femoral components as the predicate device mentioned above (S-ROM Zirconia Ceramic Femoral Head). The difference between the femoral stems which will be utilized is the material composition of either titanium or cobalt-chromium alloy. {1}------------------------------------------------ ### Indications for Use __________________________________________________________________________________________________________________________________________________________ The S-ROM Zirconia Ceramic Femoral Head is indicated for use in total hip replacement procedures for patients suffering severe pain and disability due to structural damage in the hip joint from rheumatoid arthritis, osteoarthritis, posttraumatic arthritis, collagen disorders, avascular necrosis, and non-union of femoral fractures. Use of the prosthesis is also indicated for revision of previous hip arthroplasty and for patients with congenital hip dysplasia, protrusio acetabuli, slipped femoral epiphysis, and disability due to previous fusion. ### Physical Description__________________________________________________________________________________________________________________________________________________________ The S-ROM Zirconia Ceramic femoral head is manufactured of either Yttrium Stabilized Zirconium Oxide (Zirconia), also known as PROZYR® or ZYRANOX™ zirconia ceramic. The S-ROM ceramic femoral head is designed to be used with 11/13 Morse-taper femoral stems composed of either titanium alloy or cobalt-chromium alloy. The ceramic femoral heads are contraindicated for use with any acetabular components other than an UHMWPE cup or metal backed UHMWPE cup. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract design that resembles an eagle or a bird in flight. The design is composed of three curved lines that suggest the wings and body of the bird. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR - 9 1998 Ms. Marv E. Gray Associate Regulatory Affairs Specialist Johnson and Johnson Professional, Inc. 325 Paramount Drive Raynham, Massachusetts 02767-0350 K980081 Re : S-ROM® Zirconia Ceramic Femoral Hip Head Requlatory Class: II Product Code: LZO Dated: January 8, 1998 Received: January 9, 1998 Dear Ms. Gray: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the following limitation that the package insert must reflect that the S-ROM Zirconia Ceramic Femoral Heads are to be used only with cobalt-chrome or Ti-6Al-4V alloy hip stems with the Johnson and Johnson Professional, Inc. 11/13 Morse- taper trunnions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to {3}------------------------------------------------ #### Page 2 - Ms. Mary E. Gray comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does ents response of for for for might have under sections 531 through 542 of the Act for devices under the Electronic enrough Sadiation Control provisions, or other Federal Laws or Requlations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use for the S-ROM® Zirconia Ceramic Femoral Head for Use with Cobalt-Chromium-Molybdenum (Co-Cr-Mo) Alloy 11/13 Morse-Taper Femoral Stems ## Johnson & Johnson Professional, Inc. 325 Paramount Drive Raynham, MA 02767-0350 ## Indications of Use The S-ROM Zirconia Ceramic Femoral Head is indicated for use in total hip replacement procedures for patients suffering severe pain and disability due to structural damage in the hip joint from rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, collagen disorders, avascular necrosis, and non-union of femoral fractures. Use of the prosthesis is also indicated for revision of previous hip arthroplasty and for patients with congenital hip dysplasia, protrusio acetabuli, slipped femoral epiphysis, and disability due to previous fusion. **Prescription Use** (Per 21 CFR 801.109) (Division Sign-Off Division of General Re 510(k) Number 14
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...