PATIENT SPECIFIC IMPLANT HIP FEMORAL COMPONENT

K970346 · Kairos Orthopaedics · LZO · Feb 19, 1997 · Orthopedic

Device Facts

Record IDK970346
Device NamePATIENT SPECIFIC IMPLANT HIP FEMORAL COMPONENT
ApplicantKairos Orthopaedics
Product CodeLZO · Orthopedic
Decision DateFeb 19, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3353
Device ClassClass 2
AttributesTherapeutic

Intended Use

The PSI Hip Femoral Component with HA Coating is indicated for uncemented use as the femoral component in total hip arthroplasty (THA) for replacing the hip joint of patients whose hip joint has been damaged by degenerative joint disease, fracture or the failure of previous arthroplasty. Each PSI stem is specifically designed for clinical cases where the patient would see added benefit from a femoral component designed to match the natural bone geometry of the femur.

Device Story

Device is a patient-specific femoral hip stem designed to match individual femoral canal geometry. Input data consists of patient CT or X-ray images used to custom-manufacture the implant within defined dimensional envelopes. Device is intended for uncemented use in total hip arthroplasty. Operated by orthopedic surgeons in clinical settings. Output is a physical prosthetic implant that replaces the damaged hip joint. Benefit is improved fit and fill of the femoral canal compared to standard off-the-shelf stems. Fatigue performance is validated via engineering stress analysis and mechanical testing of representative stems.

Clinical Evidence

Bench testing only. Engineering stress analysis performed on design parameter envelopes to ensure fatigue performance comparable to clinically proven standard hip stems. Mechanical testing on representative PSI stems verified stress analysis calculations.

Technological Characteristics

Materials: Wrought or forged Titanium (Ti-6Al-4V). Surface treatment: Hydroxylapatite (HA) coating. Design: Patient-specific geometry within defined dimensional envelopes (90-300mm length, 8-20mm distal diameter). Features: Collar, distal flutes. Sterilization: Sterile and non-sterile options.

Indications for Use

Indicated for patients requiring total hip arthroplasty (THA) due to degenerative joint disease, fracture, or failed previous arthroplasty. Intended for uncemented use in cases where patient-specific geometry matching provides clinical benefit.

Regulatory Classification

Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} K970346 # Premarket Notification ## 510(k) Summary FEB 19 1997 In accordance with the Food and Drug Administration Interim Rule to implement provisions of the Safe Medical Devices Act (SMDA) of 1990 and in conformance with 21 CFR 807, this is to serve as a 510(k) Summary for the Kairos™ Orthopaedics Patient Specific Implant (PSI) Hip Femoral Component with HA Coating. | Submitter: | Kairos™ Orthopaedics 819 Striker Avenue, Suite 10 Sacramento, California 95834 Telephone: (916) 646-5403 Fax: (916) 646-5432 | | --- | --- | Date: January 28, 1997 Contact Person: Brian Cleary<br/>Director, Regulatory Affairs Classification Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis (21 CFR 888.3353) Common Name: Patient Matched Hip Stem Trade/Proprietary Name: Patient Specific Implant (PSI) Hip Femoral Component with HA Coating Product Code: 87 LZO ## Device Description: The Kairos™ Orthopaedics Patient Specific Implant (PSI) Hip Femoral Component with HA Coating is designed using the patient's natural bone geometry to construct an implant to better fit and fill the patient's femoral canal relative to other commercially available standard hip stems. Since each stem is matched to a particular patient, a precise device description cannot be provided, as the device design will vary from patient to patient depending upon their unique, naturally occurring bone geometry. Kairos™ Orthopaedics specifies eight design parameters that define an "envelope" of dimensional limits within which all PSI stems are defined. The PSI Hip Femoral Component with HA Coating is indicated for uncemented use as the femoral component in total hip arthroplasty (THA) for replacing the hip joint of patients whose hip joint has been damaged by degenerative joint disease, fracture or the failure of previous arthroplasty. Each PSI stem is specifically designed for clinical cases where the patient would see added benefit from a femoral component designed to match the natural bone geometry of the femur. 34 {1} 35 # Substantial Equivalence: Several commercially available standard line hip products are substantially equivalent to the Kairos™ Orthopaedics' PSI Hip Femoral Component with HA Coating with respect to overall design, materials, intended use, and sterilization methods. Additionally, the manufacturing methods and applied surface treatments are substantially the same as those used in the production of a wide range of currently marketed standard line hip products. It is our understanding that other anatomically based hip stems have received 510(k) clearance in the past, including the following: 1. Patient Matched Implant (PMI) Hip Femoral Component Biomet, Inc. (Warsaw, Indiana) 2. Techmedica Continuum Hip System Techmedica, Inc. (Camarillo, California) 3. Biopro DRG Option Hip Replacement System – Femoral Stem Biopro, Inc. (Port Huron, Michigan) Of these, Biomet's Patient Matched Implant (PMI) is the most similar and is still presently marketed. Biomet's PMI hip stem is also a patient matched hip stem which uses patient data to design and manufacture a hip stem to match a particular patient's naturally occurring bone geometry. Due to the similarities in the general device design concept, Kairos™ Orthopaedics believes that the Biomet PMI stem is ideal for comparison to demonstrate the substantial equivalence of Kairos™ Orthopaedics' PSI Hip Femoral Component with HA Coating. The table on the following page summarizes these feature comparisons. Additionally, an engineering stress analysis was performed on the envelope of design parameters to ensure fatigue performance of the PSI Hip Femoral Component with HA Coating comparable to a clinically proven standard hip stem. Mechanical testing performed by Kairos™ Orthopaedics on a representative PSI stem verified these stress analysis calculations. {2} Feature comparisons between the Kairos™ Orthopaedics PSI Hip Femoral Component with HA Coating and Biomet, Inc. PMI hip femoral component. Differences are indicated in bold print. | Characteristics | Subject Device Kairos™ Orthopaedics Patient Specific Implant (PSI) with HA Coating | Predicate Device Biomet, Inc. Patient Matched Implant (PMI) | | --- | --- | --- | | Manufacturer | Kairos™ Orthopaedics | Biomet, Inc. | | 510(k) Approved | - | Yes | | Materials | Wrought or Forged Titanium (Ti-6Al-4V) | Wrought Cobalt Chromium (Co-Cr-Mo) or Wrought Titanium (Ti-6Al-4V) | | Intended Use | Uncemented | Cemented/Uncemented | | Patient Matched | Yes | Yes | | Collar | Yes | Yes | | Distal Flutes | Yes | Yes | | Surface Treatments | • Hydroxylapetite (HA) Coating | • Smooth • Grit Blasted • Plasma Sprayed Coating • Ion Nitride (NTS) | | Stem Length from Resection Level | 90mm - 300mm | unknown | | Distal Diameter | 8mm - 20mm | unknown | | Patient Information | CT or X-Ray | CT or X-Ray | | Sterilization Status | Sterile & Non-sterile | Sterile |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...