← Product Code [LZO](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LZO) · K233724

# 28mm/38D MDM X3 Insert for MDM Liner (K233724)

_Howmedica Osteonics Corp., Dba Stryker Orthopaedics · LZO · Mar 11, 2024 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LZO/K233724

## Device Facts

- **Applicant:** Howmedica Osteonics Corp., Dba Stryker Orthopaedics
- **Product Code:** [LZO](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LZO.md)
- **Decision Date:** Mar 11, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3353
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The indications for use for total hip arthroplasty include: - 1. Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis; - 2. Rheumatoid arthritis: - 3. Correction of functional deformity; - 4. Revision procedures where other treatments or devices have failed; and, - 5. Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques. 6. Dislocation risks MDM Liners are intended for cementless use only.

## Device Story

The 28mm/38D MDM X3 Insert is a sterile, single-use orthopedic component for total hip arthroplasty. It functions as a modular dual mobility bearing, articulating within a highly polished cobalt chrome MDM liner and pairing with 28mm femoral heads. The device is implanted by orthopedic surgeons in a clinical setting. It serves as a line extension to existing Stryker dual mobility systems, utilizing the same X3 (sequentially crosslinked and annealed) polyethylene material. The insert provides a bearing surface to restore joint function and stability, reducing dislocation risk in patients with degenerative joint disease, rheumatoid arthritis, or femoral fractures. Clinical decision-making relies on the surgeon's assessment of the patient's joint pathology and the compatibility of the insert with existing Stryker femoral stems and acetabular components.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical bench testing, including material characterization (ASTM F648), biocompatibility (ISO 10993), sterilization validation (ISO 11135), bacterial endotoxin testing, and mechanical performance testing (axial pull-out, cam-out strength, wear, axial/rim fatigue, range of motion, dislocation resistance, and impingement FEA).

## Technological Characteristics

Material: Sequentially crosslinked and annealed UHMWPE (Type 1 GUR 1020, ASTM F648). Form factor: 28mm inner diameter, 38mm outer diameter acetabular insert. Principle: Modular dual mobility bearing articulation. Sterilization: Ethylene oxide (ISO 11135). Connectivity: None (mechanical implant).

## Regulatory Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

## Predicate Devices

- Trident X3 Acetabular Inserts, ADM and MDM X3 Acetabular Inserts, Accolade II Hip System ([K182468](/device/K182468.md))
- G7 Dual Mobility System, Active Articulation System ([K161190](/device/K161190.md))

## Reference Devices

- Exeter X3 RimFit Cup ([K213701](/device/K213701.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 11, 2024

Howmedica Osteonics Corp. dba Stryker Orthopaedics Meenakshi Verma Staff Regulatory Affairs Specialist 325 Corporate Drive Mahwah, New Jersey 07430

Re: K233724

Trade/Device Name: 28mm/38D MDM X3 Insert for MDM Liner Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, LZO, MEH Dated: November 21, 2023 Received: November 21, 2023

Dear Meenakshi Verma:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Limin Sun -S

Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

Submission Number (if known)

K233724

Device Name

28mm/38D MDM X3 Insert for MDM Liner

Indications for Use (Describe)

The indications for use for total hip arthroplasty include:

- 1. Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
- 2. Rheumatoid arthritis:
- 3. Correction of functional deformity;
- 4. Revision procedures where other treatments or devices have failed; and,
- 5. Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.

6. Dislocation risks

MDM Liners are intended for cementless use only.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

#### 510(k) Summary

| Sponsor              | Stryker Orthopaedics<br>325 Corporate Drive<br>Mahwah, NJ 07430                                                                                                                |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person       | Meenakshi Verma<br>Staff Regulatory Affairs Specialist<br>Howmedica Osteonics Corp<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>+911244850600<br>meenakshi.verma4@stryker.com |
| Alternate Contact    | Gregg Ritter<br>Principal Regulatory Affairs Specialist<br>Howmedica Osteonics Corp<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>+1-201-831-5665<br>gregg.ritter@stryker.com  |
| Date Prepared:       | 1-March-2024                                                                                                                                                                   |
| Proprietary Name:    | 28mm/38D MDM X3 Insert for MDM Liner                                                                                                                                           |
| Common Name:         | Total Hip Joint Replacement                                                                                                                                                    |
| Classification Name: | Hip joint metal/polymer/metal semi-constrained porous-coated<br>uncemented prosthesis (21 CFR 888.3358)                                                                        |
| Product Codes:       | LPH, LZO, MEH                                                                                                                                                                  |

#### Primary Predicate Device to Which Substantial Equivalence is Claimed:

- Trident X3 Acetabular Inserts, ADM and MDM X3 Acetabular Inserts, Accolade II Hip ● System (K182468)
## Additional Predicate Devices Used to Support Substantial Equivalence:

- G7 Dual Mobility System, Active Articulation System (K161190) .
## Reference Device Used to Support Substantial Equivalence :

- Exeter X3 RimFit Cup (K213701) .

{5}------------------------------------------------

#### Reason for 510(k) Submission:

The purpose of this submission is to introduce a line extension to the MDM X3 Modular Dual Mobility Acetabular System, specifically a 28mm/38D MDM X3 Insert for MDM Liner.

#### Device Description:

The subject device is an MDM X3 Insert with an inner diameter of 28mm and outer diameter of 38mm. It is a sterile, single-use device that will articulate within the inner surface of the highly polished cobalt chrome MDM Liner and will be compatible with 28mm Howmedica Osteonics femoral heads. The subject device is compatible with the same liners, cups, and femoral stems as the previously cleared Stryker dual mobility hip systems. The polyethylene MDM X3 Insert and femoral head combine to create the dual mobility bearing. The subject MDM X3 Insert is manufactured from sequentially crosslinked and annealed polyethylene, known as X3. The resin used is Type 1 (GUR 1020) which meets the requirements of ASTM F648.

#### Intended Use:

The subject device has the same intended use as that specified in the 510(k) clearance for the predicate devices.

#### Indications:

The indications for use for total hip arthroplasty include:

- . Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
- Rheumatoid arthritis; ●
- Correction of functional deformity;
- Revision procedures where other treatments or devices have failed; and,
- Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur ● with head involvement that are unmanageable using other techniques.
- Dislocation risks ●

MDM Liners are intended for cementless use only.

## Summary of Technological Characteristics:

The subject X3 acetabular insert has the same intended use, indications for use, material and operational principles as those of the predicate devices. The design of the subject device is similar to its predicate devices. The predicate MDM X3 Inserts have an inner diameter of 22.2mm and outer diameter of 36mm and 38mm. Restoration ADM X3 Inserts and Restoration ADM/MDM X3 Inserts have an inner diameter of 28mm and outer diameter ranging from 40mm to 58mm in 2mm increments. The predicate acetabular inserts are compatible with 22.2mm or 28mm Howmedica Osteonics femoral heads. respectively. The subject MDM X3 Insert has an inner diameter of 28mm and outer diameter of 38mm. It will articulate within the inner surface of the highly polished cobalt chrome MDM Liner and will be compatible with 28mm Howmedica Osteonics femoral heads.

{6}------------------------------------------------

The subject device 28mm/38D MDM X3 Insert is thinner than the Stryker predicates ADM and MDM X3 acetabular inserts but similar in size/dimensions as compared to the additional predicate 28/38 Dual Mobility heads (K161190).

## Non-Clinical Testing:

The following non-clinical laboratory testing was performed, or engineering analysis was conducted to determine substantial equivalence:

- Material testing per FDA's guidance titled, "Characterization of Ultrahigh Molecular ● Weight Polyethylene (UHMWPE) Used in Orthopedic Devices," issued April- 26-2019.
- Biocompatibility evaluation per ISO 10993-1:2018, ISO 10993-7:2008 and FDA's ● guidance titled, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," issued September 4, 2020.
- Ethylene oxide sterilization validation per ISO 11135:2014
- Bacterial endotoxin testing (BET)
- Axial pull out resistance
- Cam-out strength
- Wear testing
- Axial Fatigue
- Rim Fatigue
- Range of Motion
- Dislocation Resistance ●
- Impingement FEA ●
- MRI Safety Testing

## Clinical Testing:

Clinical testing was not required as a basis to demonstrate substantial equivalence.

## Conclusion:

Based upon a comparison of the intended use, indications for use, design, materials, technical and performance characteristics, and operational principles, the 28mm/38D MDM X3 Insert for MDM Liner is substantially equivalent to the predicate devices identified in this premarket notification.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LZO/K233724](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LZO/K233724)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
