← Product Code [LZO](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LZO) · K193050

# Alloclassic Zweymuller SL Femoral Stem, Alloclassic Zweymuller SL Offset Femoral Stem (K193050)

_Zimmer GmbH · LZO · Dec 26, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LZO/K193050

## Device Facts

- **Applicant:** Zimmer GmbH
- **Product Code:** [LZO](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LZO.md)
- **Decision Date:** Dec 26, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3353
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

Alloclassic® Zweymüller® SL/ SL Offset Femoral Stems: • Noninflammatory degenerative joint disease (NIDJD), e.g., avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g., rheumatoid arthritis. · Failed previous surgery where pain, deformity, or dysfunction persists. • Revision of previously failed hip arthroplasty

## Device Story

Alloclassic Zweymüller SL and SL Offset Femoral Stems are orthopedic implants for total hip arthroplasty; designed as conical, straight stems with rectangular cross-sections for rotational stability. Stems are implanted by surgeons in clinical settings to replace the femoral head/neck; provide primary stability in poor bone quality; anchor distally via edges and proximally against bone cortex. SL Offset version provides additional 6.25 mm offset and 121° CCD/neck angle (vs 131° for SL) to assist in soft tissue balancing, leg length restoration, and joint stability. Devices are manufactured from wrought Ti-Al-Nb titanium alloy with grit-blasted surfaces to enhance bone/prosthesis interface. Output is a mechanical structural support for the hip joint; clinical decision-making relies on surgeon assessment of patient anatomy and joint pathology. Benefits include restoration of hip function and reduction of pain.

## Clinical Evidence

No clinical data was required or provided for this submission. Substantial equivalence is supported by non-clinical performance testing, including packaging integrity verification per ISO 11607-1:2006 and ISO 11607-2:2006, and mechanical integrity/resistance verification for instruments.

## Technological Characteristics

Materials: Wrought Ti-Al-Nb titanium alloy (Protasul-100™, ISO 5832-11 / ASTM 1295-16). Design: Conical straight stem, rectangular cross-section, grit-blasted surface. SL Offset variant: 121° CCD/neck angle, 6.25 mm additional offset. Sterilization: Not specified (implied same as predicate). Connectivity: None (mechanical implant).

## Regulatory Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

## Predicate Devices

- Alloclassic® Zweymüller® SL Femoral Stem ([K962101](/device/K962101.md), [K030373](/device/K030373.md))
- Alloclassic® Zweymüller® SL Offset Femoral Stem ([K033664](/device/K033664.md))

## Reference Devices

- Avenir Complete Hip System ([K182048](/device/K182048.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

December 26, 2019

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.

Zimmer GmbH Gaelle Georges Regulatory Affairs Senior Specialist Sulzerallee 8 Winterthur, 8404 Ch

### Re: K193050

Trade/Device Name: Alloclassic Zweymuller SL Femoral Stem, Alloclassic Zweymuller SL Offset Femoral Stem Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO, JDI, KWL, KWY Dated: October 29, 2019 Received: November 1, 2019

Dear Gaelle Georges:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

FOR Vesa Vuniqi Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### K193050

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

Device Name

Alloclassic® Zweymüller® SL / SL Offset Femoral Stems

Indications for Use (Describe)

Alloclassic® Zweymüller® SL/ SL Offset Femoral Stems:

• Noninflammatory degenerative joint disease (NIDJD), e.g., avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g., rheumatoid arthritis.

· Failed previous surgery where pain, deformity, or dysfunction persists.

• Revision of previously failed hip arthroplasty

Type of Use (Select one or both, as applicable)

| <div> <span> <span style="text-decoration: overline;"> </span>      Prescription Use (Part 21 CFR 801 Subpart D)     </span> </div> | <span>     Over-The-Counter Use (21 CFR 801 Subpart C)    </span> |
|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|

**CONTINUE ON A SEPARATE PAGE IF NEEDED.**

 This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

 The burden time for this collection of information is estimated to average 79 hours per response, including the
 time to review instructions, search existing data sources, gather and maintain the data needed and complete
 and review the collection of information. Send comments regarding this burden estimate or any other aspect
 of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services
   

 Food and Drug Administration
   

 Office of Chief Information Officer
   

 Paperwork Reduction Act (PRA) Staff
   

*PRAStaff@fda.hhs.gov*

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

==============================================================================================================================================================================

{3}------------------------------------------------

K193050

ZIMMER BIOMET

# 510(k) SUMMARY

| Sponsor:                         | Zimmer GmbH<br>Sulzerallee 8, P.O. Box<br>8404 Winterthur, Switzerland                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                  | Gaelle Georges<br>Regulatory Affairs Senior Specialist<br>Telephone: +41 58 854 8136<br>Fax: +41 52 244 86 58                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Date:                            | October 29, 2019                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Trade Name:                      | Alloclassic® Zweymüller® SL Femoral Stem<br>Alloclassic® Zweymüller® SL Offset Femoral Stem                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Common Name                      | Hip Prosthesis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Classification Product Code :    | LZO, JDI, KWL, KWY                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Device Classification Name:      | Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or<br>uncemented                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Regulation Number / Description: | 21 CFR § 888.3353 - Hip joint metal/ceramic/polymer semi-<br>constrained cemented or nonporous uncemented prosthesis.<br>21 CFR § 888.3350 - Hip joint metal/polymer semi-constrained<br>cemented prosthesis.<br>21 CFR § 888.3360 - Hip joint femoral (hemi-hip) metallic<br>cemented or uncemented prosthesis.<br>21 CFR § 888.3390 - Hip joint femoral (hemi-hip) metal/polymer<br>cemented or uncemented prosthesis.                                                                                                                                                                                                                                              |
| Predicate Device:                | Alloclassic® Zweymüller® SL Femoral Stem, manufactured by<br>Zimmer GmbH, K962101, first cleared August 19, 1996,<br>subsequently modified by a design change and cleared March 6,<br>2003 under K030373<br>Alloclassic® Zweymüller® SL Offset Femoral Stem, manufactured<br>by Zimmer GmbH, K033664, cleared December 17, 2003                                                                                                                                                                                                                                                                                                                                       |
| Device Description:              | The Alloclassic® Zweymüller® SL Femoral Stem is designed as a<br>conical straight stem with a rectangular cross section to ensure<br>rotational stability. The general design of the Alloclassic<br>Zweymüller SL Femoral Stem comprises a slim neck and a short<br>taper. Alloclassic Zweymüller SL Femoral Stems are available in<br>14 sizes and can be implanted to achieve primary stability in<br>cases of unexpected poor bone quality. Distally the stem is<br>anchored primarily by its edges; proximally the surfaces anchor<br>against the bone cortex. The Alloclassic Zweymüller SL Femoral<br>Stem is manufactured from wrought Ti-Al-Nb titanium alloy |

{4}------------------------------------------------

(Protasul-100™, ISO 5832-11 / ASTM 1295-16). The entire stem below the neck of the hip stem is grit basted to provide enhanced bone/prosthesis interface.

The design of Alloclassic® Zweymüller® SL Offset Femoral Stem is very similar to the previously cleared Alloclassic Zweymüller SL Femoral Stems (K962101, K030373), and identical to the previously cleared Alloclassic Zweymüller SL Offset Femoral Stem (K033664). Similar to the Alloclassic Zweymüller SL Femoral Stem, the Alloclassic Zweymüller SL Offset Femoral Stem is a conically shaped straight stem with a rectangular cross section for rotational stability. The entire stem below the neck of the hip stem is grit blasted to provide enhanced bone/prosthesis interface. The stem is available in 14 sizes and is manufactured from the wrought Ti-Al-Nb titanium alloy (Protasul-100™, ISO 5832-11 / ASTM 1295-16). In contrast to the Alloclassic Zweymüller SL Femoral Stem, the Alloclassic Zweymüller SL Offset Femoral Stem provides an additional 6.25 mm of offset (per size) and has a CCD/neck angle of 121° (vs 131° for SL stem). Overall the Alloclassic Zweymüller SL Offset Femoral Stem design provides a higher offset sometimes needed for correcting insufficient soft tissue balancing, restoration of leg length, and sufficient joint stability following total hip arthroplasty.

Indications For Use:

Alloclassic® Zweymüller® SL/ SL Offset Femoral Stems:

- Noninflammatory degenerative joint disease (NIDJD), e.g., avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g., rheumatoid arthritis.
- Failed previous surgery where pain, deformity, or dysfunction persists
- . Revision of previously failed hip arthroplasty
- Comparison to Predicate Device: The intended use and indications for use of the modified devices, as described in its labeling, have not changed as a result of the modifications proposed in the present submission. Further, the materials, design features and sterilization method of the subject devices remain identical to the predicate devices.

Minor modifications have been implemented since the last clearances.

Modification of packaging configuration for the subject implant devices is proposed. The new packaging configuration and the methods to support the package integrity have been previously described in K182048 (Avenir Complete Hip System). The proposed changes do not alter the fundamental scientific technology shared by both the subject devices and predicate devices. In addition, the Instructions For Use (IFU) of the implant devices is proposed to be revised without any changes to the intended use, indications or contraindications: the IFU is being converted to a brand specific IFU and the content is adapted to the subject devices. Zimmer GmbH is furthermore seeking clearance for certain system-specific Class II instruments - these instruments that have previously considered Class I exempt and

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Zimmer Biomet. The logo consists of a blue circle with a white "Z" inside of it, followed by the text "ZIMMER BIOMET" in a sans-serif font. The text is gray and is aligned to the right of the circle.

correction of classification to Class II is proposed within present submission.

#### Performance Data (Nonclinical and/or Clinical):

correction of classification to Class II is proposed within present
submission.

Non-Clinical Performance and Conclusions:

Packaging configuration change:

Packaging performance testing was performed to verify that packaging configuration maintains integrity of the sterile barrier system up to the point of use and provides adequate protection to the product through the hazards of sterilization, handling, distribution and storage according to ISO 11607-1:2006 and ISO 11607-2:2006. Packaging Configuration testing was conducted by representative worst-case products.

Correction of instrument classification from Class I to Class II:

Amendment of Design Controls with verification of mechanical integrity and resistance.

Clinical Performance and Conclusions:

Clinical data and conclusions were not needed for this device.

Conclusion: The subject devices have the same intended use and same indications for use as the predicate devices. The subject devices use the same operating principle, incorporate the same basic design and labeling and are manufactured and sterilized using the same materials and processes as the predicate devices.

> Except for the modifications described in this submission the subject devices are identical to the predicate devices, and the performance data and analyses demonstrate that:

- · any differences do not raise new questions of safety and effectiveness as established with performance testing; and
- the subject devices are at least as safe and effective as the legally marketed predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LZO/K193050](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LZO/K193050)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).