MEDACTA TOTAL HIP PROSTHESIS SYSTEM - AMISTEM C & QUADRA C SHORT NECK

K103189 · Medacta International S.A. · LZO · Dec 20, 2010 · Orthopedic

Device Facts

Record IDK103189
Device NameMEDACTA TOTAL HIP PROSTHESIS SYSTEM - AMISTEM C & QUADRA C SHORT NECK
ApplicantMedacta International S.A.
Product CodeLZO · Orthopedic
Decision DateDec 20, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3353
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Medacta Total Hip Prosthesis System – Quadra C Femoral Stems are intended for cemented use in total or partial hip arthroplasty and in primary or revision surgery. Hip replacement is indicated in the following cases: • Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia • Avascular necrosis of the femoral head • Acute traumatic fracture of the femoral head or neck • Failure of previous hip surgery, joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement. Size 0 implants should not be implanted in patients with a mass of 65 kg or greater.

Device Story

Femoral stems for total or partial hip arthroplasty; line extension of Medacta Total Hip Prosthesis System. Components include AMIStem C (decreased length, reduced shoulder) and Quadra C Short Neck (SN) (5mm shorter neck). Forged stainless steel (ISO 5832-9); collarless; straight quadratic triple-taper design. Proximal 12/14 taper connects to metallic or ceramic femoral heads. Available in multiple sizes with 135° (standard) or 127° (lateralized) CCD angles. Used by orthopedic surgeons in clinical settings for cemented hip replacement. Provides additional anatomical sizing options for patient-specific fit.

Clinical Evidence

Bench testing only. No clinical data provided. Performance verified through endurance testing of neck and stem regions and range of motion analysis based on engineering analysis of worst-case configurations.

Technological Characteristics

Material: Forged stainless steel (ISO 5832-9). Design: Collarless, straight quadratic stem with triple taper. Connection: Standard 12/14 taper for femoral head attachment. CCD angles: 135° (standard offset) and 127° (lateralized offset). Cemented fixation.

Indications for Use

Indicated for patients requiring total or partial hip arthroplasty (primary or revision) due to painful/disabled joints from arthritis, traumatic arthritis, rheumatoid polyarthritis, congenital hip dysplasia, avascular necrosis, or acute traumatic fracture of the femoral head/neck, or failure of previous hip surgery. Contraindication: Size 0 implants must not be used in patients weighing 65 kg or more.

Regulatory Classification

Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/13 description: The image shows the logo for Medacta International. The logo consists of two black triangles pointing upwards, followed by the word "Medacta" in black, sans-serif font. Below the word "Medacta" is a horizontal line, and below that is the word "International" in a smaller, sans-serif font, followed by a plus sign. # 510(k) Summary Manufacturer: Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 FAX (+41) 91 696 60 66 - Contact Person: Ms. Natalie J. Kennel Consultant NJK & Associates. Inc. 13721 Via Tres Vista San Diego, CA 92129 USA Phone: (858) 705-0350 Fax: (858) 764-9739 email: NKennel@njkconsulting.com Date Prepared: October 25, 2010 ### DEVICE INFORMATION Trade/Proprietary Name: AMIStem C & Quadra C SN femoral stems Common Name /Classification Name: Hip joint, femoral metal/ ceramic, polvmer semiconstrained metallic cemented or uncemented prosthesis > 21 CFR 888.3360 Class II Device Product Code: LZO. JDI Predicate Devices: Medacta Total Hip Prosthesis system - Quadra C. Medacta International cleared under K083558. #### Product Description: This modification to the original Medacta Total Hip Prosthesis System -Quadra C femoral stems is a line extension to include the AMIStem C femoral stems and Quadra C Short Neck (SN) femoral stems. AMIStem C femoral Section 5 - Page 1 of 3 K103189(pg1/3) DEC 2 0 2010 {1}------------------------------------------------ stems are based on the design of the Quadra C femoral stems in the original submission. The AMIStem C femoral stems differ from the Quadra C femoral stems by a decreased length of 15% and a reduced shoulder. The Quadra C short neck femoral stems are the same design femoral stems as the Quadra C femoral stems in the original submission but the length of the neck that is 5 mm shorter. Like the Quadra C femoral stems, the AMIStem C femoral stems and the Quadra C SN femoral stems are highly polished, collarless femoral components manufactured from forged stainless steel according to ISO 5832-9. The AMIStem C femoral stems and the Quadra C SN femoral stems are straight quadratic stems with a triple taper shape. The proximal portion of the stem has a standard 12/14 taper for mechanical attachment to cleared Medacta International metallic or ceramic femoral heads. Like the Quadra C femoral stems, two different caput-collum-diaphyseal angles (CCD) are available for the AMIStem C femoral stems and the Quadra C Short neck femoral stems: 135° for the standard offset and 127° for the lateralized offset. The Quadra C short neck femoral stems come in eight sizes with a standard offset and seven sizes with a lateral offset. The AMIStem C femoral stems come in eight sizes with a standard offset and seven sizes with a lateral offset. There are no short neck versions for the AMIStem C femoral stems. The AMIStem C femoral stems and the Quadra C short neck femoral stems provide additional femoral stem options to the surgeon for use with the Medacta Total Hip Prosthesis System. Indications for Use: The Medacta Total Hip Prosthesis System - Quadra C Femoral Stems are intended for cemented use in total or partial hip arthroplasty and in primary or revision surgery. Hip replacement is indicated in the following cases: • Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia - · Avascular necrosis of the femoral head - Acute traumatic fracture of the femoral head or neck · Failure of previous hip surgery, joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement. Size 0 implants should not be implanted in patients with a mass of 65 kg or greater. Comparison to Predicate Devices: {2}------------------------------------------------ The Medacta Total Hip Prosthesis System- AMIStem C femoral stems and Quadra C short neck femoral stems are substantially equivalent to the Medacta Total Hip Prosthesis system - Quadra C, cleared under K083558. The AMIStem C femoral stems and the Quadra C short neck femoral stems are made of the same materials as the Quadra C femoral stems in the predicate device. They have the same general design as the Quadra C femoral stems. They have the same compatibilities as the Quadra C femoral stems. ## Performance Testing No performance standards applicable to this device have been adopted under Section 514 of the Food, Drug and Cosmetic Act. Performance testing of the AMIStem C and Quadra C Short neck femoral stems was conducted in accordance with various standards and FDA guidance documents. A detailed list is provided in Section 18 Performance Testing - Bench. The modifications to the device system to include the addition of AMIStem C and Quadra C short neck femoral stems were evaluated by risk analysis to identify any new risks associated with these changes. Based on the risk analysis, design verification was conducted to written protocols with predefined acceptance criteria. The protocols and pre-defined acceptance criteria were based on the standards and FDA guidance. The testing was conducted on the worst case based on engineering analysis. The testing included endurance testing of the neck and stem regions of the femoral components as well as range of motion analysis. The design verification activities met all acceptance criteria and verified the performance of the Medacta Total Hip Prosthesis System - AMIStem C femoral stems and Quadra C short neck femoral stems. #### Conclusion: The results from the design control activities and the information provided in this submission support the conclusion that the Medacta Total Hip Prosthesis system - AMIStem C femoral stems and Quadra C short neck femoral stems . are substantially equivalent to its predicate device, Medacta Total Hip Prosthesis Quadra C femoral stems with respect to indications for use and technological characteristics. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Medacta International SA % NJK & Associates, Inc Ms. Natalie J. Kennel 13721 Via Tres Vista San Diego. CA 92129 DEC 2 0 2010 Re: K0103189 Trade/Device Name: Medacta Total Hip Prosthesis – AMIStem C, Quadra C SN Regulation Number: 21 CFR 888.3360 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Code: LZO, JDI Dated: November 30, 2010 Received: December 1, 2010 Dear Ms. Kennel: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is elabilitional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean I hat FDA has made a determination that your device complies with other requirements of the Act that 1 27 Thus Intatutes and regulations administered by other Federal agencies. You must or any I ederal state Act's requirements, including, but not limited to: registration and listing (21 {4}------------------------------------------------ Page 2 – Ms. Natalie J. Kennel comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/wcm115809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, AK B. R. h Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use Statement 510(k) Number (if known): K103189 (pg 1/1) Device Name: AMIStem C & Quadra C SN Indications for Use: The Medacta Total Hip Prosthesis System – Quadra C Femoral Stems are intended for cemented use in total or partial hip arthroplasty and in primary or revision surgery. Hip replacement is indicated in the following cases: - Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia - · Avascular necrosis of the femoral head - · Acute traumatic fracture of the femoral head or neck · Failure of previous hip surgery, joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement. Size 0 implants should not be implanted in patients with a mass of 65 kg or greater. Prescription Use × (Part 21 CFR 801 Subpart D) Over-The-Counter Use AND/OR (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) 510(k) Number Concurrence of CDRH, Office of Device Evaluation (ODE) for M. Melkerm (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices Page of
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...