BIOLOX DELTA CERAMIC FEMORAL HEAD

K082991 · Aesculap Implant Systems, Inc. · LZO · Nov 20, 2008 · Orthopedic

Device Facts

Record IDK082991
Device NameBIOLOX DELTA CERAMIC FEMORAL HEAD
ApplicantAesculap Implant Systems, Inc.
Product CodeLZO · Orthopedic
Decision DateNov 20, 2008
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3353
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Excia Hip System is intended to replace a hip joint. The device is intended for: Patients suffering from severe hip and disability due to rheumatoid arthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head and nonunion of previous fractures of the femur. Patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital femoral epiphysis Patients suffering from disability due previous fusion Patients with acute femoral neck fractures The Excia Hip System is available with two femoral stems. One is manufactured from CoCrMo and is intended for cemented fixation. The other femoral stem is for uncemented fixation and manufactured from Ti with Plasmapore with or without u-CaP®. The Metha® Hip System (uncemented, press-fit fixation) is intended to replace a hip joint. The device is intended for: skeletally mature individuals undergoing primary surgery for total hip replacement patients suffering from severe hip pain and disability due to rheumatoid arthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head and nonunion of previous fractures of the femur. patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital femoral epiphysis patients suffering from disability due to previous fusion patients with acute femoral neck fractures

Device Story

BIOLOX® delta Ceramic Femoral Head; alumina matrix composite ceramic component for total hip arthroplasty. Used as modular head option for Excia and Metha Hip Systems. Available in 28, 32, 36 mm diameters with various neck lengths. Implanted by orthopedic surgeons in clinical/hospital settings. Provides alternative to metal or alumina ceramic heads; improves wear characteristics and biocompatibility. Surgeon selects appropriate head size/neck length during surgery to restore joint function and stability.

Clinical Evidence

Bench testing only. Testing performed per FDA guidance documents for orthopedic implants, including metallic plasma sprayed coatings, non-articulating modular components, acetabular cup prostheses, femoral stem prostheses, ceramic ball hip systems, and UHMWPE data requirements.

Technological Characteristics

Material: Alumina matrix composite. Form factor: Modular femoral head, 28/32/36 mm diameters. Application: Total hip arthroplasty. Connectivity: None (mechanical implant). Sterilization: Not specified.

Indications for Use

Indicated for skeletally mature patients undergoing primary total hip replacement due to rheumatoid arthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis, nonunion of femoral fractures, congenital hip dysplasia, protrusion acetabuli, slipped capital femoral epiphysis, previous fusion, or acute femoral neck fractures.

Regulatory Classification

Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Page 1 of 2 # 6082991 (pg112) #### 510(k) SUMMARY (as required by 21 CFR 807.92) B. ## BIOLOX® delta Ceramic Femoral Head NOV 20 20 October 6, 2008 - Aesculap Implant Systems, Inc. COMPANY: 3773 Corporate Parkway Center Valley, PA 18034 Establishment Registration Number: 3005673311 - Kathy A. Racosky CONTACT: 610-984-9291 (phone) 610-791-6882 (fax) kathy.racosky@aesculap.com (email) - BIOLOX® delta Ceramic Femoral Head TRADE NAME: - Ceramic Femoral Head COMMON NAME: - CLASSIFICATION NAME: Hip joint Metal/Ceramic/Polymer Semi-Constrained Cemented or Non-Porous Uncemented Prosthesis Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calcium-Phosphate Prosthesis. Hip. Hemi-. Femoral. Metal/Polymer. Cemented or Uncemented - REGULATION NUMBER: 888,3353,888,3360,888,3353,888,3390 PRODUCT CODE: LZO, LWJ. MEH, KWY ## SUBSTANTIAL EQUIVALENCE Aesculab Implant Systems, Inc. believes that the BIOLOX® delta Ceramic Femoral Head is a line extension of Aesculap Implant Systems Excia (K042344, K060918, and K062684) - Hip Systems and Metha Short Stem Hip System (K080584) that were previously cleared. It is also substantially equivalent to the Zimmer BIOLOX detta Ceramic Femoral Head (K071535). ## DEVICE DESCRIPTION The Aesculap Implant Systems BIOLOX® delta Ceramic Femoral Heads are manufactured from an alumina matrix composite. The ceramic femoral head is offered in thee diameters of 28, 32, and 36 mm with a range of neck lengths. The BIOLOX® delta Ceramic head provides the surgeon another option to both the metal and alumina ceramic femoral heads for use in total hip arthroplasty. {1}------------------------------------------------ # K082991 (pg 212) Page 2 of 2 ## INDICATIONS FOR USE The Excia Hip System is intended to replace a hip joint. The device is intended for: - o is intents suffering from severe hip and disability due to rheumatoid arthritis, traumatic arthritis, ● r alle of the really on disorders, avascular necrosis of the femoral head and nonunion of previous fractures of the femur. - Patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital femoral epiphysis ● - Patients suffering from disability due previous fusion ◆ - . Patients with acute femoral neck fractures The Excia Hip System is available with two femoral stems. One is manufactured from CoCrMo and is intended for cemented Tixation. The other femoral stern is for uncemented fixation and manufactured from Ti with Plasmapore with or without u-CaP®. #### The Metha® Hip System (uncemented, press-fit fixation) is intended to replace a hip joint. The device is intended for: - skeletally mature individuals undergoing primary surgery for total hip replacement - patients suffering from severe hip pain and disability due to meumatoid arthritis, traumatic arthritis, . polyarthritis, collagen disorders, avascular necrosis of the femoral head and nonunion of previous fractures of the femur. - patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital femoral epiphysis . - patients suffering from disability due to previous fusion . - . patients with acute femoral neck fractures ## TECHNOLIGICAL CHARACTERISTICS(compared to Predicate(s)} The Aesculap Implant Systems BIOLOX® delta Ceramic Femoral Heads are offered in similar shapes and sizes as the predicate devices. The material used for the Aesculap Implant Systems device is the same as that used to manufacture the predicate devices. ### PERFORMANCE DATA All required testing per "Draft Guidance for the Preparation of Premarket Notifications (510(k)s) Applications for Orthopedic Devices-The Basic Elements" were done where applicable. In addition, testing per the: - "Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces . Apposing Bone or Bone Cement", - "Guidance for Industry on the Testing of Metallic Plasma Sprayed Coatings on . Orthopedic Implants to Support Reconsideration of Postmarket Surveillance Requirements", - "Guidance Document for Testing Non-articulating, "Mechanically Locked" Modular . Implant Components", - "Draft Guidance Document for Testing Acetabular Cup Prostheses", . - "Points to Consider for Femoral Stem Prostheses", . - "Guidance Document for the Preparation of Premarket Notifications for Ceramic Ball . Hip Systems" and - "Data Requirements for Ultrahigh Molecular Weight Polethevlene (UHMWPE) Used in . Orthopedic Devices" was completed where applicable. {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Image /page/2/Picture/2 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features an abstract eagle design, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" encircling the eagle. The text is arranged in a circular pattern around the eagle, emphasizing the department's name and national affiliation. The eagle is depicted in a stylized manner, with flowing lines suggesting movement and dynamism. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Aesculap Implant Systems, Inc. % Ms. Kathy A. Racosky Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034 NOV 20 2008 Re: K082991 Trade/Device Name: BIOLOX® DELTA Ccramic Fcmoral Head Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class; II Product Code: LZO, LWJ, MEH, KWY Dated: November 7, 2008 Received: November 10, 2008 Dear Ms. Racosky: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Ms. Kathy A. Racosky This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRII's Office of Surveillance and Biometric's (OSB's) Division of Prestmarket, Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events. (Medical Device Reporting (MDR)), please contact the Division of Surveillance Syrems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Mark M. Millman Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page 1 of 1 #### A. INDICATIONS FOR USE STATEMENT 510(k) Number: h08299) (pq ili) Device Name: BIOLOX® delta Ceramic Femoral Head For use with the Aesculap Implant Systems Excia and Metha Hip System ### Indications for Use: The Excia Hip System is intended to replace a hip joint. The device is intended for: Patients suffering from severe hip and disability due to rheumatoid arthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head and nonunion of previous fractures of the femur. - Patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital femoral epiphysis - Patients suffering from disability due previous fusion - Patients with acute femoral neck fractures ● The Excia Hip System is available with two femoral stems. One is manufactured from CoCMo and is intended for cemented fixation. The other femoral stem is for uncemented fixation and manufactured from Ti with Plasmapore with or without u-CaP®. The Metha® Hip System (uncemented, press-fit fixation) is intended to replace a hip (oint. The device is intended for: - skeletally mature individuals undergoing primary surgery for total hip replacement patients suffering from severe hip pain and disability due to rheumatoid arthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head and nonunion of previous fractures of the femur. - patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital femoral epiphysis - patients suffering from disability due to previous fusion - patients with acute femoral neck fractures . (Division Sign-Up) Division of General, Restorative and Neurologic **510(k) Number** k082441 Prescription Use × and/or Over-the-Counter Use (per 21 CFR 801.109) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...