← Product Code [LZO](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LZO) · K070928

# CERAMIC  BALL HEADS, 28, 32 AND 36 MM (K070928)

_Plus Orthopedics AG · LZO · Aug 1, 2007 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LZO/K070928

## Device Facts

- **Applicant:** Plus Orthopedics AG
- **Product Code:** [LZO](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LZO.md)
- **Decision Date:** Aug 1, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3353
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Plus Orthopedics' Ceramic Ball Heads are intended for mechanical fixation to a mating hip stem and indicated for treatment of patients who are candidates for primary and revision total hip arthroplasty where the hip joint needs restructuring due to disease or trauma. These devices are intended to aid the surgeon in relieving the patient of hip pain and restoring hip motion. Please note that the patient should be skeletally mature and the patient's condition should be due to the following: 1. Osteoarthritis 2. Rheumatoid arthritis 3. Tumor conditions involving the upper third of the femur of the Acetabular 4. Ankylosing spondylitis 5. Psoriatic arthritis 6. Old osteomyelitis with a long infection-free period and a normal WBC. ESR and C-reactive protein 7. Non-union of femoral neck fracture or avascular necrosis of the femoral head 8. Post-traumatic fracture/dislocation of the hip 9. Revision of an unsuccessful arthrodesis with either poor positioning or pain in the hip, or where low back pain or knee pain is becoming disabling 10. Revision of an unsuccessful cemented or un-cemented hip replacement, providing sufficient bone stock is present 11. Revision of a previous unsuccessful osteotomy, Girdlestone resection, cup arthroplasty or hemi-arthroplasty. These ceramic ball heads are to be used only with hip stems manufactured by Plus Orthopedics AG, specifically SL-PLUS®/SLR-PLUS®, SL- PLUS® Lateral, Modular- PLUS® and IPM stem.

## Device Story

Ceramic ball heads for mechanical fixation to specific Plus Orthopedics hip stems (SL-PLUS, SLR-PLUS, SL-PLUS Lateral, Modular-PLUS, IPM). Used in primary and revision total hip arthroplasty to restore hip motion and relieve pain. Operated by surgeons in clinical settings. Device acts as the femoral head component of a hip prosthesis. Benefits include structural restoration of the hip joint in patients with disease or trauma.

## Clinical Evidence

Bench testing only. Compliance with ISO 6474:1994 (ceramic materials) and ISO 7206-10:2003 (hip-joint prostheses). Design verification and validation performed per 21 CFR 820.30.

## Technological Characteristics

Material: BIOLOX® forte (high purity aluminum oxide per ISO 6474 with magnesium). Sizes: 28, 32, 36 mm. Sterilization: Radiation per ISO 11137. Mechanical fixation to proprietary hip stems.

## Regulatory Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

## Predicate Devices

- Portland Ceramic (BIOLOX® forte) ([K061564](/device/K061564.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K070928

AUG - 1 2007

## Premarket Notification 510(k) Summary

#### 1.0 Submitter Information

| Company Name      | Plus Orthopedics AG                                              |
|-------------------|------------------------------------------------------------------|
| Company Address   | Erlenstraße 4a<br>CH-6343 Rotkreuz<br>Switzerland                |
| Contact Name      | Pamela J. Weagraff, Principal Consultant<br>Quintiles Consulting |
| Contact Address   | 18 Bridie Lane<br>Norfolk, MA 02056                              |
| Contact Telephone | 508-528-1745                                                     |
| Contact Facsimile | 978-752-1225                                                     |
| Contact E-mail    | pamela.weagraff@quintiles.com                                    |

#### 2.0 Date Prepared: June 8, 2007

#### 3.0 Name of Device

- 3.1 Device Name: Ceramic Ball Heads, various sizes
- 3.2 Common Name: Ceramic Ball Heads
- 3.3 Classification Name and Reference: Hip Joint Metal/Ceramic/Polymer Semi-constrained Cemented or Nonporous Uncemented Prosthesis, Title 21 CFR Part 888.3353

#### 4.0 Substantial Equivalence Claimed to Predicate Device

Portland Ceramic (BIOLOX® forte), K061564 Need to add CoCrMo taper predicate used with Portland

#### 5.0 Device Description

The Plus Orthopedics' Ceramic Ball Heads are intended for mechanical fixation to a mating hip stem and indicated for treatment of patients who are candidates for primary and revision total hip arthroplasty where the hip joint needs restructuring due to disease or trauma. These devices are intended to aid the surgeon in relieving the patient of hip pain and restoring hip motion.

These ceramic ball heads are to be used only with hip stems manufactured by Plus Orthopedics AG, specifically SL-PLUS®/SLR- PLUS®, SL- PLUS® Lateral, Modular- PLUS® and IPM stem. Manufactured by CeramTec AG, the Ceramic Ball Heads are made of BIOLOX® forte, a high purity aluminum oxide per ISO 6474, with a small amount of magnesium to prevent grain growth. The Ceramic Ball Heads are available in various sizes from 28, 32 up to 36 mm, in small medium and large for each.

{1}------------------------------------------------

#### 6.0 Indications for Use

The Plus Orthopedics AG Ceramic Ball Heads are intended for mechanical fixation to a mating hip stem and indicated for treatment of patients who are candidates for primary and revision total hip arthroplasty where the hip joint needs restructuring due to disease or trauma. These devices are intended to aid the surgeon in relieving the patient of hip pain and restoring hip motion. Please note that the patient should be skeletally mature and the patient's condition should be due to the following:

- Osteoarthritis 1.
- Rheumatoid arthritis 2.
- Tumor conditions involving the upper third of the femur of the Acetabular 3.
- 4. Ankylosing spondylitis
- 5. Psoriatic arthritis
- Old osteomyelitis -- with a long infection-free period and a normal WBC. ESR and C-reactive 6. protein
- 7. Non-union of femoral neck fracture or avascular necrosis of the femoral head
- 8. Post-traumatic fracture/dislocation of the hip
- ರ್ Revision of an unsuccessful arthrodesis with either poor positioning or pain in the hip, or where low back pain or knee pain is becoming disabling
- 10. Revision of an unsuccessful cemented or un-cemented hip replacement, providing sufficient bone stock is present
- 11. Revision of a previous unsuccessful osteotomy, Girdlestone resection, cup arthroplasty or hemi-arthroplasty.

These ceramic ball heads are to be used only with hip stems manufactured by Plus Orthopedics AG, specifically SL-PLUS®/SLR-PLUS®, SL- PLUS® Lateral, Modular- PLUS® and IPM stem.

### 7.0 Predicate Device Comparison of Indications for Use / Intended Use and Technical Characteristics

The comparison of the Ceramic Ball Heads was based on a review of the Design Control documentation, relevant aspects of which are included in the company's 510(k) Premarket Notification, and information concerning the predicate device that was obtained from the FDA web site. The comparison considered technical characteristics and the indications for use / intended use.

#### 8.0 Performance Data

- 8.1 Performance Standards: no performance standards applicable to this device, Hip Joint Metal/Ceramic/Polymer Semi-constrained Cemented or Nonporous Uncemented Prosthesis, Title 21 CFR Part 888.3353; Product Code: LZO, have been adopted under Section 514 of the Food Drug and Cosmetic Act.
- 8.2 The Ceramic Ball Heads comply with the following FDA recognized standards:
	- 8.2.1 ISO 6474:1994, Implants for surgery -- Ceramic materials based on high purity alumina
	- 8.2.2 ISO 7206-10, 2003, Implants for surgery -- Partial and total hip-joint prostheses --Part 10

{2}------------------------------------------------

- AAMI / ANSI / ISO 11137:1994, Sterilization of health care products Requirements 8.2.3 for validation and routine control -- radiation sterilization and ANSI/AAMI/ISO 11137:1994 (Amendment 1:2002).
- Performance Testing: Design verification and design validation, e.g., bench testing was 8.3 performed according to FDA's Design Control Requirements, Title 21 Code of Federal Regulations, Part 820.30.

#### 9.0 Conclusion

The information and data provided in this 510(k) Premarket Notification establish that the Ceramic Ball Heads, various sizes, are substantially equivalent to the afore-mentioned predicate device with respect to indications for use/intended use, and technical characteristics.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird with three human profiles incorporated into its design. The profiles are arranged in a way that they form the body and wings of the bird. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird image.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG - 1 2007

Plus Orthopedics AG % Ms. Pamela Weagraff Principal Consultant 18 Bridie Lane Norfolk, Massachusetts 02056

Re: K070928

Trade/Device Name: Ceramic Ball Heads, 28, 32 and 36 mm Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint, metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: II Product Code: LZO Dated: March 30, 2007 Received: April 3, 2007

Dear Ms. Weagraff:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Ms. Pamela Weagraff

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,
Mark N. Milkerson

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# ODE Indications for Use Statement

### 510(k) Number (if known): K070928

### Device Name: Ceramic Ball Heads, 28, 32 and 36 mm

### Indications for Use:

The Plus Orthopedics' Ceramic Ball Heads are intended for mechanical fixation to a mating hip stem and indicated for treatment of patients who are candidates for primary and revision total hip arthroplasty where the hip joint needs restructuring due to disease or trauma. These devices are intended to aid the surgeon in relieving the patient of hip pain and restoring hip motion. Please note that the patient should be skeletally mature and the patient's condition should be due to the following:

- 1. Osteoarthritis
- Rheumatoid arthritis 2.
- Tumor conditions involving the upper third of the femur of the Acetabular 3.
- 4. Ankylosing spondylitis
- Psoriatic arthritis 5.
- Old osteomyelitis with a long infection-free period and a normal WBC. ESR and C-6. reactive protein
- 7. Non-union of femoral neck fracture or avascular necrosis of the femoral head
- 8. Post-traumatic fracture/dislocation of the hip
- Revision of an unsuccessful arthrodesis with either poor positioning or pain in the hip, or 9. where low back pain or knee pain is becoming disabling
- · 10. Revision of an unsuccessful cemented or un-cemented hip replacement, providing sufficient bone stock is present
- 11. Revision of a previous unsuccessful osteotomy, Girdlestone resection, cup arthroplasty or hemi-arthroplasty.

These ceramic ball heads are to be used only hip stems manufactured by Plus Orthopedics AG, specifically SL-PLUS®/SLR- PLUS®, SL- PLUS® Lateral, Modular- PLUS® and IPM stem.

Prescription Use: ____________________________________________________________________________________________________________________________________________________________ AND/OR (21 CFR 801 Subpart D)

Over-the-Counter Use: (21 CFR 807 Subpart C)

## (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

**Concurrence of CDRH, Office of Device Evaluation (ODE)**

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

Indications for Use Statement

510(k) Number_

Page C-2 of C-2

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LZO/K070928](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LZO/K070928)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
