ZIRCONIA CERAMIC FEMORAL HEAD - 12/14 TAPER

K013989 · Smith & Nephew, Inc. · LZO · Dec 19, 2001 · Orthopedic

Device Facts

Record IDK013989
Device NameZIRCONIA CERAMIC FEMORAL HEAD - 12/14 TAPER
ApplicantSmith & Nephew, Inc.
Product CodeLZO · Orthopedic
Decision DateDec 19, 2001
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3353
Device ClassClass 2
AttributesTherapeutic

Intended Use

Total hip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomyelitis with an extended drainage-free period; nonunion; femoral neck fracture and trochanteric fractures of the proximal femur with heads involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. The Zirconia Ceramic Femoral Head is designed for single use only.

Device Story

Zirconia Ceramic Femoral Head; prosthetic component for total hip replacement. Designed for use with titanium or cobalt chromium alloy femoral stems via 12/14 taper interface. Implanted by orthopedic surgeons during primary or revision hip surgery. Functions as load-bearing articular surface to restore hip joint function. Single-use device. Benefits patients by replacing damaged femoral head, reducing pain, and restoring mobility in cases of severe joint degeneration or trauma.

Clinical Evidence

Bench testing only. Mechanical testing performed per ceramic femoral head draft guidance requirements demonstrated equivalence to predicate devices and ability to withstand in vivo loading.

Technological Characteristics

Material: Zirconia ceramic. Interface: 12/14 taper. Compatibility: Titanium and cobalt chromium alloy femoral components. Single-use. Mechanical testing performed per ceramic femoral head draft guidance.

Indications for Use

Indicated for individuals undergoing primary and revision hip surgery for trauma, inflammatory joint disease (e.g., rheumatoid arthritis), or noninflammatory degenerative joint disease (e.g., osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, pelvic fracture, osteomyelitis, nonunion, femoral neck/trochanteric fractures, femoral osteotomy, Girdlestone resection, fracture dislocation, or deformity correction).

Regulatory Classification

Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K 0/3989 ge loft Summary of Safety and Effectiveness Smith & Nephew, Inc. Zirconia Ceramic Femoral Head # Contact Person and Address DEC 1 9 2001 Janet Johnson Green Manager, Clinical and Regulatory Affairs Smith & Nephew, Inc. Orthopaedic Division 1450 East Brooks Road Memphis, TN 38116 (901) 396-2121 #### Device Description The Zirconia Ceramic Femoral Head designed for use with both titanium and cobalt chromium alloy femoral components with a 12/14 taper. ### Device Classification Name Device Classification wetal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Class II #### Indications for Use Total hip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomyelitis with an extended drainage-free period; nonunion; femoral neck fracture and trochanteric fractures of the proximal femur with heads involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. The Zirconia Ceramic Femoral Head is designed for single use only. ### Mechanical and Clinical Data Mechanical testing was performed according to the requirements in the ceramic femoral head draft guidance document. All of the test results indicate that the Zirconia Ceramic Femoral Head is equivalent to devices currently on the market and capable of withstanding in vivo loading without failure. ## Substantial Equivalence Information The Zirconia Ceramic Femoral Head with a 12/14 taper is similar to the Zirconia Ceramic Femoral Head with a 14/16 taper distributed by Smith & Nephew, Inc. Both heads are indicated for total hip replacement and are similar in design. CONFIDENTIAL {1}------------------------------------------------ Image /page/1/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 1 9 2001 Ms. Janet Johnson Green Manager, Clinical and Regulatory Affairs Smith & Nephew, Inc. 1450 East Brooks Road Memphis, Tennessee 38116 Re: K013989 Trade/Device Name: Zirconia Ceramic Femoral Head Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-constrained Cemented or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO Dated: November 26, 2001 Received: December 4, 2001 Dear Ms. Green: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ # Page 2 - Ms. Janet Johnson Green This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ K013989 page 1 of 1 # Zirconia Ceramic Femoral Head Indications Statement Total hip components are indicated for individuals undergoing primary and revision Total mp Confoonchs are indicated in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses nonimalinitis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; Such as Useballunines, avascular incostrophic variant; old, remote osteomyelitis with an iused inip, fracture of the period; nonunion; femoral neck fracture and trochanteric fractures of the proximal femur with heads involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. Signature (Division Sign-Off) (Division of General, Restorative and Neurological Devices 510(k) Number K013989
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...