K070061 · Hayes Medical, Inc. · LWJ · Jan 31, 2007 · Orthopedic
Device Facts
Record ID
K070061
Device Name
36MM COCR FEMORAL HEAD AND 36MM ACETABULAR INSERT
Applicant
Hayes Medical, Inc.
Product Code
LWJ · Orthopedic
Decision Date
Jan 31, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3360
Device Class
Class 2
Attributes
Therapeutic
Intended Use
A) Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis. B) Revision of failed femoral head replacement, cup arthroplasty or other hip procedures. C) Proximal femoral fractures. D) Avascular necrosis of the femoral head. E) Non-union of proximal femoral neck fractures. F) Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, osteoporosis, pseudarthrosis conversion, and structural abnormalities Acetabular components are indicated for cemented and cementless use. Consensus femoral stems are indicated for cemented and cementless use. UniSyn femoral stems are indicated for cementless use only. HA coated implants are indicated for cementless use only.
Device Story
Consensus Hip System 36mm femoral head and acetabular insert; intended for total hip arthroplasty. Femoral head manufactured from CoCrMo alloy; acetabular insert manufactured from crosslinked UHMWPE. Insert includes titanium alloy (Ti 6Al-4V ELI) x-ray markers for hooded configurations. Device components interface with existing Consensus Hip System femoral stems and acetabular shells. Used by orthopedic surgeons in clinical/surgical settings to restore joint function in patients with degenerative or traumatic hip conditions. Provides mechanical articulation for hip joint replacement; design features (taper, locking mechanism) identical to previously cleared predicate components. Benefits include expanded sizing options for patient-specific anatomical fit.
Clinical Evidence
No clinical data. Substantial equivalence is based on design, material, and technological characteristics compared to legally marketed predicate devices.
Technological Characteristics
Femoral head: CoCrMo alloy (ASTM F799 or F1537). Acetabular insert: Crosslinked UHMWPE (ASTM F648). Hooded inserts include Ti 6Al-4V ELI (ASTM F136) x-ray markers. Mechanical articulation; semi-constrained hip prosthesis. Dimensions: 36mm femoral head; inserts sizes 54-68mm. No software or electronic components.
Indications for Use
Indicated for patients with significantly impaired hip joints due to rheumatoid, osteo, or post-traumatic arthritis; failed prior hip procedures; proximal femoral fractures; avascular necrosis; non-union of femoral neck fractures; or congenital/developmental/metabolic conditions requiring hip arthroplasty.
Regulatory Classification
Identification
A hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis is a device intended to be implanted to replace a portion of the hip joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum. This generic type of device includes designs which are intended to be fixed to the bone with bone cement (§ 888.3027) as well as designs which have large window-like holes in the stem of the device and which are intended for use without bone cement. However, in these latter designs, fixation of the device is not achieved by means of bone ingrowth.
Biomet Orthopedics, Inc., RingLoc® 36mm femoral head and liner (K032396)
Related Devices
K021310 — 36MM C-TAPER FEMORAL HEAD COMPONENTS (LFIT AND NON-LFIT) · Howmedica Osteonics Corp. · May 21, 2002
K183129 — Medtronic Total Hip Arthroplasty System · Medtronic Advanced Energy · Jul 26, 2019
K133228 — ICONACY I-HIP SYSTEM, GRADUAL TRANSITIONING (GT) FEMORAL SYSTEM · Iconacy Orthopedic Implants, LLC · Jan 17, 2014
K121034 — IOI TOTAL HIP · Iconacy Orthopedic Implants, LLC · Sep 4, 2012
K053536 — DUROM ACETABULAR COMPONENT AND METASUL LDH LARGE DIAMETER HEADS · Zimmer GmbH · Mar 16, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
K070061 Pg 1 of 2
## Special 510(k) Consensus Hip System - Addition of a 36mm CoCrMo Femoral Head and Acetabular Insert
# Revised Section 9 510(k) Summary of Safety and Effectiveness
Defined in 21 CFR 807 In accordance with 21 CFR 807.92 (Summary)
JAN 3 1 2007
| Applicant's Name: | Hayes Medical, Inc.<br>1115 Windfield Way, Suite 100<br>El Dorado Hills, CA 95682 |
|------------------------------|----------------------------------------------------------------------------------------|
| Contact Person: | Luke Rose |
| Trade Name: | 36mm femoral head, Consensus Hip |
| Common Name: | 36mm femoral head, Consensus Hip |
| Classification Name: | Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented (21 CFR<br>888.3360, LWJ) |
| | Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented (21 CFR<br>888.3350, JDI) |
| | Hip System acetabular Insert (21 CFR 888.3358, LPH) |
| Proposed Regulatory Class: | Class II |
| Device Classification Panel: | Orthopaedic |
| Substantially Equivalent To: | Hayes Medical Consensus Hip System (K922561) |
| | Hayes Medical Crosslinked UHMWPE Insert (K021466) |
| | DePuy 36mm UHMWPE highly crosslinked liner (K994415) |
| | Biomet Orthopedics, Inc., RingLoc® 36mm femoral head and liner<br>(K032396) |
#### Intended Use:
### Indications for use of the ConsensusTM Hip System or UNISYN™ Hip System
- A) Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis.
- B) Revision of failed femoral head replacement, cup arthroplasty or other hip procedures.
- C) Proximal femoral fractures.
- D) Avascular necrosis of the femoral head.
- E) Non-union of proximal femoral neck fractures.
- F) Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa vara, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, osteoporosis, pseudarthrosis conversion, and structural abnormalities
Acetabular components are indicated for cemented and cementless use. Consensus femoral stems are indicated for cemented and cementless use. UniSyn femoral stems are indicated for cementless use only. HA coated implants are indicated for cementless use only.
### Device Description:
The Consensus Hip 36mm CoCrMo (ASTM F799 or ASTM F1537) femoral head is offered with four different offsets, -5mm, neutral, +5mm and +10mm. Other than size, it incorporates no new design features. Female head taper is identical to the head taper design for the cleared devices.
{1}------------------------------------------------
K070061 Pg 2 of 2
### Special 510(k)
## Consensus Hip System - Addition of a 36mm CoCrMo Femoral Head and Acetabular Insert
The 36mm acetabular insert is offered in sizes 54 - 68 (2 mm increments), in neutral, 10 and 20 degree hooded configurations. Material for all 36mm insert sizes is crosslink UHMWPE (ASTM F648). Hooded configurations incorporate a titanium alloy x-ray marker (Ti 6Al-4V ELJ ASTM F136). Insert locking detail of the subject device is identical to the locking detail of the cleared devices.
### Comparison to Cleared Device:
The only change made to the previously cleared Consensus femoral heads and liners is the addition of a new size, the 36 mm femoral head and a series of compatible acetabular inserts, including 10 and 20 degree hooded configurations.
#### Substantial Equivalence Information
The intended use, material, design features and type of interface of the 36mm femoral head and inserts are substantially equivalent to predicate Hayes Medical and competitive devices. The safety and effectiveness of the subject 36mmm head and inserts are adequately supported by the substantial equivalence information and materials data provided within this Special 510(k) submission.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The image is in black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Hayes Medical, Inc. % Mr. Luke Rose Director, Quality Systems and Regulatory Affairs 115 Windfield Way, Suite 100 El Dorado Hills, California 95762-9623
JAN 3 1 2007
Re: K070061
Trade/Device Name: Consensus Hip System 36 mm CoCr Femoral Head Regulation Number: 21 CFR 888.3360 Regulation Name: Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis. Regulatory Class: II Product Code: JDI, LWJ, LPH Dated: January 4, 2007 Received: January 5, 2007
Dear Mr. Rose:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. Luke Rose
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807,97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Barbara Buchner
Mark N. Melkerson Director Division of General. Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Special 510(k)
## Consensus Hip System - Addition of a 36mm CoCrMo Femoral Head and Acetabular Insert
# Revised Section 7 Indications for Use
## 510(k) Number (if known): 10-7006 |
Device Name: Consensus Hip System
Indications For Use:
## Indications for use with the CONSENSUS® Hip System or UNISYN™ Hip Ssytem
- A > Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis.
- B) Revision of failed femoral head replacement, cup arthroplasty or other hip procedures.
- C) Proximal femoral fractures.
- D) Avascular necrosis of the femoral head.
- E) Non-union of proximal femoral neck fractures.
- F) Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, osteoporosis, pseudarthrosis conversion, and structural abnormalities
Acetabular components are indicated for cemented and cementless use.
Consensus femoral stems are indicated for cemented and cementless use.
UniSyn femoral stems are indicated for cementless use only.
HA coated implants are indicated for cementless use only.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Barbara Buelum
Page 1 of 1
Restorative. Division of Genera and Neurological Devices
**510(k) Number** Ko70061
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.