UNI-ROM TOTAL HIP SYSTEM FEMORAL STEM

K974331 · Johnson & Johnson Professionals, Inc. · LPH · Feb 6, 1998 · Orthopedic

Device Facts

Record IDK974331
Device NameUNI-ROM TOTAL HIP SYSTEM FEMORAL STEM
ApplicantJohnson & Johnson Professionals, Inc.
Product CodeLPH · Orthopedic
Decision DateFeb 6, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3358
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The UNI-ROM™ Total Hip System Femoral Stem is indicated for use in total hip replacement procedures for patients suffering severe pain and disability due to structural damage in the hip joint from rheumatoid arthritis, osteoarthritis, posttraumatic arthritis, collagen disorders, avascular necrosis, and nonunion of femoral fractures. Use of the prosthesis is also indicated for patients with congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis, and disability due to previous fusion, where bone stock is inadequate for other reconstructive techniques.

Device Story

One-piece porous-coated femoral hip stem; manufactured from Ti-6Al-4V titanium alloy; proximal circumferential area porous-coated with commercially pure titanium. Anatomically shaped with distal flutes and coronal slot; designed for uncemented fixation via compressive loading. Mates with femoral hip heads using 11/13 Morse-type taper; compatible with S-ROM Co-Cr or ceramic femoral heads. Used by orthopedic surgeons in total hip replacement procedures. Provides structural support for hip joint; facilitates bone ingrowth for long-term fixation; restores joint function; reduces pain and disability.

Clinical Evidence

Bench testing only.

Technological Characteristics

Material: Ti-6Al-4V titanium alloy substrate; commercially pure titanium porous coating. Design: One-piece, anatomically shaped, distal flutes, coronal slot, 11/13 Morse-type taper. Fixation: Uncemented, press-fit, proximal porous coating. Sterilization: Not specified.

Indications for Use

Indicated for patients requiring total hip replacement due to severe pain/disability from rheumatoid arthritis, osteoarthritis, posttraumatic arthritis, collagen disorders, avascular necrosis, nonunion of femoral fractures, congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis, or previous fusion with inadequate bone stock.

Regulatory Classification

Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## Exhibit I # 510(k) Summary UNI-ROM™ Femoral Hip Stem # FEB - 6 1998 1974351 Johnson & Johnson Professional, Inc. 325 Paramount Drive Raynham, Massachusetts 02767 #### 1. Contact Person: Janet G. Johnson Associate Regulatory Affairs Specialist (508) 828-3466 #### Device Information: 2. | Proprietary Name: | UNI-ROM™ Total Hip System Femoral<br>Stem | |----------------------|-------------------------------------------------------------------------------------------| | Common Name: | Femoral Stem Prostheses | | Classification Name: | Hip joint metal/polymer/metal semi-<br>constrained porous-coated uncemented<br>prosthesis | | Regulatory Class: | Class II, per 21 §CFR 888.3358 | | Product Code: | 87 LPH | #### 3. Substantial Equivalence: The UNI-ROM™ Femoral Hip Stem is substantially equivalent in terms of intended use, materials, design, manufacturing processes, sterilization method, packaging and performance to the S-ROM Femoral Hip Stem (K961939, K954935 and K851422) and P.F.C. 2 Femoral Stem (K945518 and K935452). #### 4. Indications for Use: The UNI-ROM™ Total Hip Stem is indicated for use in total hip replacement procedures for patients suffering severe pain and disability due to structural damage in the hip joint from rheumatoid arthritis, osteoarthritis, posttraumatic arthritis, collagen disorders, avascular necrosis, and nonunion of femoral fractures. Use of the prosthesis is also indicated for patients with congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis, and disability due to previous fusion, where bone stock is inadequate for other reconstructive techniques. {1}------------------------------------------------ #### 5. Physical Description: The UNI-ROM™ Femoral Hip Stem is a one-piece porous coated hip stem that manufactured from titanium alloy (Ti-6Al-4V). The stem is designed with steps on the outer surface of the porous-coated area to encourage compressive loading of the bone there by allowing for uncemented fixation. The proximal circumferential area is porous coated with commercially pure titanium. The UNI-ROM™ Femoral Hip Stem is anatomically shaped with flutes and a coronal slot in the distal portion of the stem. The UNI-ROM™ Femoral hip stem can be used in either the right or left hip and mates with the femoral hip head using an 11/13 Morse-type taper. Further, the femoral stem is designed to accept an S-ROM Co-Cr Femoral head or an S-ROM Ceramic Femoral Head. The determination of substantial equivalence for this device was based in part on bench top performance data. {2}------------------------------------------------ - ## Similarities and Differences Matrix | | UNI-ROM™<br>Femoral Hip<br>Stem | S-ROM®<br>Femoral Hip<br>Stem (K961939,<br>K954935 and<br>K851422) | P.F.C. 2 Femoral<br>Stem - Size<br>Small (K945518<br>and K935452) | |------------------------------------------------------------------------|------------------------------------------------------|--------------------------------------------------------------------|-------------------------------------------------------------------| | <b>DESIGN</b> | | | | | Femoral Head<br>locks onto the stem<br>with a Morse-type<br>taper lock | Yes | Yes | Yes | | Intended for non-<br>cemented fixation | Yes | Yes | Yes | | Distal Slot | Yes | Yes | No | | Distal Flutes | Yes | Yes | Yes | | Adjunct to a Total<br>Hip System | Yes | Yes | Yes | | Anatomically<br>Shaped | Yes | Yes | Yes | | Press Fit | Yes | Yes | Yes | | Porous Coating | Proximal | Proximal Sleeve | Proximal | | Recommended for<br>use with: | Femoral Heads<br>with a 11/13<br>Morse Type<br>Taper | Femoral Heads<br>with a 11/13<br>Morse Type<br>Taper | Femoral Heads<br>with a 10/12<br>Morse Type<br>Taper | | <b>INTENDED</b> | | | | | Indicated for use in<br>primary hip<br>replacement<br>procedures. | Yes | Yes | Yes | | <b>MATERIALS</b> | | | | | Ti-6Al-4V Substrate | Yes | Yes | Yes | | Commercially Pure<br>Titanium Porous<br>Coating | Yes | Yes | No | | Ti-6Al-4V Porous<br>Coating | No | No | Yes | {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with three overlapping wings. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 FEB - 6 1998 Ms. Janet G. Johnson · Associate Regulatory Affairs Specialist Johnson and Johnson Professional, Inc. 325 Paramount Drive Raynham, Massachusetts 02767-0350 Re: K974331 UNI-ROM™ Femoral Hip Stem Trade Name: Regulatory Class: II Product Code: LPH Dated: November 17, 1997 Received: November 18, 1997 Dear Ms. Johnson: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic ...... Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ ### Page 2 - Ms. Janet G. Johnson This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D. Celta M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 510(k) Number (if known) Device Name # UNI-ROM™ Femoral Hip Stem ~ Indications For Use: The UNI-ROM™ Total Hip System Femoral Stem is indicated for use in total hip replacement procedures for patients suffering severe pain and disability due to structural damage in the hip joint from rheumatoid arthritis, osteoarthritis, posttraumatic arthritis, collagen disorders, avascular necrosis, and nonunion of femoral fractures. Use of the prosthesis is also indicated for patients with congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis, and disability due to previous fusion, where bone stock is inadequate for other reconstructive techniques. (Please do not write below this line - Continue on another page if necessary) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General Restorative Devices K974331 510(k) Number . Prescription Use X (Per 21 CFR §801.109) OR Over-the-Counter Use (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...