FOUNDATION POROUS COATED FLARED RIM SHELL

K974095 · Encore Orthopedics, Inc. · LPH · Jan 28, 1998 · Orthopedic

Device Facts

Record IDK974095
Device NameFOUNDATION POROUS COATED FLARED RIM SHELL
ApplicantEncore Orthopedics, Inc.
Product CodeLPH · Orthopedic
Decision DateJan 28, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3358
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Foundation® Porous Coated Flared Rim Acetabular Cup is intended for treatment of patients who are candidates for total hip arthroplasty because the natural femoral head and neck and/or acetabulum have been affected by osteoarthritis, inflammatory arthritis, traumatic arthritis, rheumatoid arthritis, avascular necresis or family and fracture, and revision arthromasty where hone loss is minimal. This system includes a metal-backed sineli and modular polyethylene liner and is to be used press-fit. These devices are intended to aid the surgeon in relieving the patient of hip pain and restoring hip motion.

Device Story

Foundation® Porous Coated Flared Rim Acetabular Cup is a metal-backed acetabular component for total hip arthroplasty. Device is hemispherical, available in sizes 44mm-70mm, with or without screw holes. Used by orthopedic surgeons in clinical settings to replace natural femoral head/neck/acetabulum affected by arthritis, necrosis, or fracture. Implanted via press-fit technique to provide fixation, relieve pain, and restore hip motion. Output is a stable prosthetic hip joint interface. Benefits include improved joint function and pain relief.

Clinical Evidence

Bench testing only; specifically, attachment strength testing of liners to shell to ensure suitability for in-vivo loading.

Technological Characteristics

Fabricated from wrought/forged Ti-6Al-4V (ASTM F136). Outside surface coated with commercially pure titanium beads (ASTM F67) for porous fixation. Hemispherical form factor, 44mm-70mm sizes. Modular design with metal-backed shell and polyethylene liner.

Indications for Use

Indicated for patients undergoing total hip arthroplasty due to osteoarthritis, inflammatory arthritis, traumatic arthritis, rheumatoid arthritis, avascular necrosis, femoral neck fracture, or revision arthroplasty with minimal bone loss.

Regulatory Classification

Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

Submission Summary (Full Text)

{0}------------------------------------------------ ## Summary of Safety and Effectiveness K974095 Encore Orthopedics®, Inc. 9800 Metric Blvd. Austin, TX 7875 512-832-9500 : 14:4 Trade Name: Foundation® Porous Coated Flared Rim Acetabular Cup Common Name: Metal backed acetabular component Classification Name: Hip joint metal/polymer semi-constrained uncemented prosthesis per 21 CFR 888.3358 Description: The Foundation® Porqus Coated Flared Rim Acetabular Cup is fabricated from wrought/forged Ti-6Al-4V that conforms to ASTM F136. The outside surfaces are coated with commercially pure titanium beads (ASTM F67) for the purpose of providing a porous surface for enhanced press-fit fixation. The flared shells are available with and without screw holes. It is hemispherical in shape and is available in sizes 44mm - 70mm. Intended Use: The Foundation® Porous Coated Flared Rim Acetabular Cup is intended for treatment of patients who are candidates for total hip arthroplasty because the natural femoral head and neck have been affected by osteoarthritis, inflammatory arthritis, traumatic arthritis, rheumatoid arthritis, avascular necrosis or femoral neck fracture, and revision arthroplasty where bone loss is minimal. These devices are intended to aid the surgeon in relieving the patient of hip pain and restoring hip motion. The intended use is press-fit. Comparable Features to Predicate Device(s): The spherical shape, titanium substrate, and the indexability of liners in shell and bone screw holes are features that are comparable to other devices in commercial distribution. The outside surface of the shell is porous coated to provide a porous surface for enhanced fixation. Test Results: Testing on this device included attachment strength of liners to shell. All results are sufficient for in-vivo loading. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 IAN 2 8 1998 Ms. Debbie De Los Santos ·Requlatory Specialist Encore Orthopedics, Inc. 9800 Metric Boulevard Austin, Texas 78758 K974093, K974095, and K973119 Re: Trade Name: Foundation® Porous Coated Spiked Acetabular Cup Foundation® Porous Coated Hemispherical Acetabular Cup Foundation® Porous Coated Flared Rim Acetabular Cup Regulatory Class: II Product Code: LPH Dated: October 28 and November 19, 1997 October 30 and November 21, 1997 Received: Dear Ms. De Los Santos: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosures) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), jt may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially-equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical General requlation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. {2}------------------------------------------------ Page 2 - Ms. Debbie De Los Santos This letter will allow you to begin marketing your device as The FDA described in your 510(k) premarket notification. finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, elia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosures {3}------------------------------------------------ | 510(k) Number (if known): | K974095 | |---------------------------|----------------------------------------------------| | Device Name: | FOUNDATION POROUS COATED FZARED Rim ACETABULAR CUP | | Indications For Use: | | ## Foundation® Porous Coated Flared Rim Acetabular Cup Indications For Use The Foundation® Porous Coated Flared Rim Acetabular Cup is intended for treatment of patients who are candidates for total hip arthroplasty because the natural femoral head and neck and/or acetabulum have been affected by osteoarthritis, inflammatory arthritis, traumatic arthritis, rheumatoid arthritis, avascular necresis or family and fracture, and revision arthromasty where hone loss is minimal. This system includes a metal-backed sineli and modular polyethylene liner and is to be used press-fit. These devices are intended to aid the surgeon in relieving the patient of hip pain and restoring hip motion. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Alan Webb (Division Sign-Off) Division of General Restorative Devices 510(k) Number K974098 Prescription Use (per 21 CFR 801.109) という OR Over-The-Counter Use (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...