METAL BACKED ACETABULAR COMPONENT

K963998 · Encore Orthopedics, Inc. · LPH · Jan 27, 1997 · Orthopedic

Device Facts

Record IDK963998
Device NameMETAL BACKED ACETABULAR COMPONENT
ApplicantEncore Orthopedics, Inc.
Product CodeLPH · Orthopedic
Decision DateJan 27, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3358
Device ClassClass 2
AttributesTherapeutic

Intended Use

The metal backed acetabular component is intended for treatment of patients who are candidates for total hip arthroplasty because the natural femoral head and neck and/or acetabulum have been affected by osteoarthritis, inflammatory arthritis, traumatic arthritis, rheumatoid arthritis, avascular necrosis or femoral neck fracture, and revision arthroplasty where bone loss is minimal. These devices are intended to aid the surgeon in relieving the patient of hip pain and restoring hip motion.

Device Story

Metal backed acetabular component for total hip arthroplasty; hemispherical Ti-6Al-4V shell with plasma-sprayed exterior for press-fit fixation; includes optional screw holes for secondary fixation; ultra-high molecular weight polyethylene (UHMWPE) liners available in neutral or 10° hooded designs; hooded design addresses subluxation risk from soft tissue laxity; used by orthopedic surgeons in clinical settings; device relieves hip pain and restores motion; provides stable prosthetic interface for femoral head.

Clinical Evidence

Bench testing only. Attachment strength of liners to shell tested; results sufficient for in-vivo loading.

Technological Characteristics

Materials: Ti-6Al-4V (wrought/forged or cast) shell, UHMWPE liners. Surface: Plasma-sprayed roughened exterior. Geometry: Hemispherical shell with rim flare. Fixation: Press-fit with optional cancellous bone screws. Sterilization: Not specified.

Indications for Use

Indicated for patients requiring total hip arthroplasty due to osteoarthritis, inflammatory arthritis, traumatic arthritis, rheumatoid arthritis, avascular necrosis, femoral neck fracture, or revision arthroplasty with minimal bone loss.

Regulatory Classification

Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

Related Devices

Submission Summary (Full Text)

{0} K96 3998 # Summary of Safety and Effectiveness Encore Orthopedics®, Inc. 8900 Shoal Creek Blvd. Suite 300 Austin, TX 78759 (512) 206-1437 Ashley M. Bock JAN 27 1997 **Trade Name:** Metal Backed Acetabular Component **Common Name:** Metal backed acetabular component **Classification Name:** Hip joint metal/polymer semi-constrained uncemented prosthesis per 21 CFR 888.3358 **Description:** The metal shell is fabricated from wrought/forged or cast Ti-6Al-4V. The outside surface of the shell is plasma sprayed to provide a roughened surface for enhanced press-fit fixation. The metal shell is hemispherical in shape and is available with screw holes. The screw holes are for the use of cancellous bone screws for secondary fixation. The poly liners are manufactured from ultra high molecular weight polyethylene and are available in a neutral and 10° hooded design. The hooded design is intended for those patients prone to subluxation because of soft tissue laxity. **Intended Use:** The metal backed acetabular component is intended for treatment of patients who are candidates for total hip arthroplasty because the natural femoral head and neck and/or acetabulum have been affected by osteoarthritis, inflammatory arthritis, traumatic arthritis, rheumatoid arthritis, avascular necrosis or femoral neck fracture, and revision arthroplasty where bone loss is minimal. These devices are intended to aid the surgeon in relieving the patient of hip pain and restoring hip motion. **Comparable Features to Predicate Device(s):** The spherical shape, titanium substrate, indexability of liners in shell and bone screw holes are features that are comparable to other devices in commercial distribution. The outside surface of the shell is plasma sprayed to provide a roughened surface for enhanced fixation. The Encore® metal backed acetabular component flares at the rim to enhance the press-fit fixation, with a straight line rather than a change in radius. **Test Results:** Testing on this device included attachment strength of liners to shell. All results are sufficient for in-vivo loading.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...