EMPOWR™ blade stem

K223779 · Encore Medical, LLC · LPH · Feb 13, 2023 · Orthopedic

Device Facts

Record IDK223779
Device NameEMPOWR™ blade stem
ApplicantEncore Medical, LLC
Product CodeLPH · Orthopedic
Decision DateFeb 13, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3358
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The hip joint metal uncemented prosthesis is intended to replace a hip joint. The device is intended for primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intraoperative instability and for whom all other options to constrained acetabular components have been considered. Joint replacement is indicated for patients suffering from disability due to: - noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the . natural femoral head; - . rheumatoid arthritis; - . correction of functional deformity; - femoral fracture; and . - . revision procedures where other treatments or devices have failed. This stem is to be press-fit. This stem is intended for cementless use.

Device Story

EMPOWR™ blade stem is a press-fit, porous-coated, wedge-style femoral stem for total hip arthroplasty. It serves as a prosthetic replacement for the femoral head. Designed for cementless implantation, it features enhancements over the predicate Linear® Hip Stem, including coxa vara offset, updated neck angle, and reduced distal angle. The device is implanted by surgeons in a clinical setting using a reproducible surgical technique compatible with various surgical approaches. It is compatible with existing Enovis™ femoral heads and acetabular systems. By replacing the diseased or damaged hip joint, the device aims to restore patient mobility and reduce pain. It is intended for permanent implantation.

Clinical Evidence

No clinical data submitted. Substantial equivalence is supported by bench testing, including CAD-based range of motion analysis (ISO 21535:2007), distal stem fatigue testing (ISO 7206-4:2010), and neck stem fatigue testing (ISO 7206-6:2013). Biocompatibility was evaluated per ISO 10993-1.

Technological Characteristics

Press-fit, wedge-style femoral stem; porous titanium coating; permanent implant; cementless use. Materials are qualified for permanent tissue/bone contact. Testing standards: ISO 7206-4 (distal fatigue), ISO 7206-6 (neck fatigue), ISO 21535 (ROM).

Indications for Use

Indicated for primary or revision total hip arthroplasty patients, including those with high risk of dislocation (prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, intraoperative instability), noninflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, or femoral fracture.

Regulatory Classification

Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. February 13, 2023 Encore Medical LLC Trev Thorsen Regulatory Affairs 9800 Metric Blvd Ste 200 Austin, Texas 78758 Re: K223779 Trade/Device Name: EMPOWR™ blade stem Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH, LZO, KWY, KWZ Dated: December 15, 2022 Received: December 16, 2022 Dear Trey Thorsen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Christopher Ferreira -S for Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. 510(k) Number (if known) K223779 Device Name EMPOWR™ blade stem #### Indications for Use (Describe) The hip joint metal uncemented prosthesis is intended to replace a hip joint. The device is intended for primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intraoperative instability and for whom all other options to constrained acetabular components have been considered. Joint replacement is indicated for patients suffering from disability due to: - noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the . natural femoral head; - . rheumatoid arthritis; - . correction of functional deformity; - femoral fracture; and . - . revision procedures where other treatments or devices have failed. This stem is to be press-fit. This stem is intended for cementless use. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------| | <div style="display:flex; align-items:center;"><div><span style="font-size:20px">☒</span></div><div>Prescription Use (Part 21 CFR 801 Subpart D)</div></div> | <div style="display:flex; align-items:center;"><div><span style="font-size:20px">☐</span></div><div>Over-The-Counter Use (21 CFR 801 Subpart C)</div></div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ | Page 1 of 3 Image /page/3/Picture/1 description: The image shows the logo for "enovis". The word "enovis" is written in a stylized font, with the "e", "n", and the first part of the "o" in red. The second part of the "o" is a black circle with a gradient effect, and the "vis" is in black. There is a trademark symbol to the bottom right of the logo. # 510(k) Summary 14DEC2022 #### I. SUBMITTER Encore® Medical, L.P. (dba Enovis™) 9800 Metric Blvd. Austin, TX 78758 Contact Person: Trey Thorsen Email: trey.thorsen@enovis.com Phone: 850-450-3932 ### II. DEVICE Name of Device: EMPOWR™ blade stem Common or Usual Name: Total Hip Implant Classification Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis – 21 CFR 888.3358 Regulatory Class: II Primary Product Code: LPH Secondary Product Code: LZO, KWY, KWY, KWZ #### III. PREDICATE DEVICE Primary Predicate: Enovis™ Linear® Hip Stem, K120241 Reference Predicate: Zimmer Biomet™ TaperLoc® Complete Stem, K120030 ### IV. DEVICE DESCRIPTION The EMPOWR™ blade stem represents the latest generation of a press-fit, porous coated, wedge style femoral stem. Building on the design principles of its predicate, Linear® Hip Stem, the EMPOWR™ blade stem system features enhancements such as addition of coxa vara offset, neck angle update, and distal angle reduction for wider application. The EMPOWR™ blade stem is applicable for use in total hip arthroplasty and the prosthetic replacement of femoral neck fractures in all age groups. It offers a stem design with a reproducible surgical technique carefully developed to enable consistent and accurate implantation. The EMPOWR™ blade stem features streamlined and accurate instrumentation that is highly adaptable to any surgical approach. ### V. INTENDED USE/INDICATIONS FOR USE Intended Use: Enovis™ hip devices are intended for treatment of patients who are candidates for total hip arthroplasty per the indications for use. While hip replacements are not intended to withstand activity levels and loads of normal healthy bone, they are a means of restoring mobility and reducing pain for many patients. {5}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for "enovis". The "eno" portion of the logo is red, while the "o" is stylized as a black and grey ring. The "vis" portion of the logo is black, and the trademark symbol is in the lower right corner. | Page 2 of 3 Indications for Use: The hip joint metal uncemented prosthesis is intended to replace a hip joint. The device is intended for primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intraoperative instability and for whom all other options to constrained acetabular components have been considered. Joint replacement is indicated for patients suffering from disability due to: - noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the ● natural femoral head; - rheumatoid arthritis; - correction of functional deformity; - femoral fracture; and - revision procedures where other treatments or devices have failed. This stem is to be press-fit. This stem is intended for cementless use. #### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The EMPOWR™ blade stem and accessories are technologically the same when compared to the predicate Linear Hip Stem (cleared via K120241). Comparative testing demonstrates substantial equivalence between the subject and predicate device. The stem trunnions and the porous titanium coating are identical between the subject and predicate device. There is a slight difference between the subject device ("EMPOWR™ blade stem") indications for use and the predicate device, Linear® stem ("Linear®") (K120241) indications for use statement. The added statements are for clarity and do not change the overall intent of the indications. The subject EMPOWR™ blade stem has one size smaller and one size larger than the predicate Enovis™ Linear® stem. The subject EMPOWR™ blade stem has two fewer sizes than the predicate Biomet® TaperLoc® Complete stem. The smallest size is equivalent between the two systems. The largest subject stem is not larger than the largest Biomet® TaperLoc® Complete stem. The size offering for the subject device is within the size offering of the predicate Biomet® TaperLoc® Complete stem offering. The component compatibility of the EMPOWR™ blade stem is identical to the Linear® stem (K120241) as femoral stems are compatible with all Enovis™ femoral heads and acetabular systems, including Encore® previously cleared FMP® and EMPOWR™ semi constrained acetabular shells, previously cleared femoral heads, previously cleared unipolar heads/sleeves, EMPOWR™ dual mobility and mobility system, bipolar systems, and FMP® constrained system. #### Biocompatibility testing The biocompatibility evaluation for the EMPOWR™ blade stem was conducted in accordance with the FDA guidance, Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, May 1, 1995, and International Standard ISO 10993-1 Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process, as recognized by FDA. {6}------------------------------------------------ The EMPOWR™ blade stem is intended for permanent implantation, contacting tissue/bone. Biocompatibility was ensured through the use of qualified materials and contact agents. #### Performance Testing Performance testing was completed on the subject EMPOWR™ blade stem and is detailed in the Performance Testing section. The following testing was performed to FDA recognized standards and internal protocols that is specific to the EMPOWR™ blade stem: - CAD Based Range of Motion (ROM) ISO 21535:2007 ● - Distal Stem Fatigue Testing – ISO 7206-4:2010 - Neck Stem Fatigue Testing ISO 7206-6:2013 ● #### Animal Studies No animal data submitted. Clinical Studies No clinical data submitted. ### VIII. CONCLUSIONS All testing and evaluations demonstrate that the subject device is substantially equivalent to the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...