← Product Code [LPH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH) · K183129

# Medtronic Total Hip Arthroplasty System (K183129)

_Medtronic Advanced Energy · LPH · Jul 26, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH/K183129

## Device Facts

- **Applicant:** Medtronic Advanced Energy
- **Product Code:** [LPH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH.md)
- **Decision Date:** Jul 26, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3358
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Medtronic Total Hip Arthroplasty System is indicated for cement-less use only in the following cases: . Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis; Rheumatoid arthritis; . Correction of functional deformity; Revision procedures where other treatments or devices have failed; and Nonunions, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques The Medtronic Total Hip Arthroplasty System Acetabular Screws are intended for supplemental fixation of the associated Medtronic Acetabular Cup.

## Device Story

Total hip arthroplasty system designed to restore hip joint alignment, stability, and range of motion while alleviating pain. System comprises femoral and acetabular components, including highly crosslinked UHMWPE acetabular liners and porous-coated cups. Used by orthopedic surgeons in clinical settings to replace articulating surfaces of the hip joint. Acetabular screws provide supplemental fixation for the acetabular cup. Device benefits patients by providing a durable, cementless prosthetic solution for degenerative or traumatic hip conditions.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including wear testing per ISO 14242-1:2014 and ISO 14242-2:2016, and material characterization per ASTM F2565, ISO 10993-17, and ISO 10993-18.

## Technological Characteristics

System components include femoral and acetabular implants. Acetabular liners utilize highly crosslinked UHMWPE (ASTM F2565). Acetabular cups feature porous coating. Testing standards: ISO 14242-1:2014, ISO 14242-2:2016, ISO 10993-17, ISO 10993-18.

## Regulatory Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

## Predicate Devices

- Responsive Orthopedics Total Hip Arthroplasty System ([K163585](/device/K163585.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Medtronic Advanced Energy Gina Cunsolo Regulatory Affairs Specialist 180 International Drive Portsmouth, New Hampshire 03801

Re: K183129

Trade/Device Name: Medtronic Total Hip Arthroplasty System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH, LZO Dated: June 25, 2019 Received: June 26, 2019

Dear Gina Cunsolo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

July 26, 2019

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for
Ve

for Vesa Vuniqi Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

The following is the Indications for Use Statement. Refer to Attachment 1 for FDA Form 3881.

#### Device Name:

Medtronic Total Hip Arthroplasty System

#### Indications:

The Medtronic Total Hip Arthroplasty System is indicated for cement-less use only in the following cases:

- . Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
- Rheumatoid arthritis;
- . Correction of functional deformity;
- Revision procedures where other treatments or devices have failed; and
- Nonunions, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques

The Medtronic Total Hip Arthroplasty System Acetabular Screws are intended for supplemental fixation of the associated Medtronic Acetabular Cup.

{3}------------------------------------------------

## 510(k) Summary

| Submitter:                          | Medtronic Advanced Energy<br>180 International Drive<br>Portsmouth, NH 03801                                                                                                                                                                                                                                                                                                                                                                                                                        |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                     | Gina Cunsolo<br>Regulatory Affairs Specialist<br>Phone: (603) 842-6210<br>Fax: (603) 742-1488<br>E-mail: Gina.M.Cunsolo@medtronic.com                                                                                                                                                                                                                                                                                                                                                               |
| Date Summary Prepared:              | November 9, 2018                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Device Trade Name:                  | Medtronic Total Hip Arthroplasty System                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Common Name:                        | Hip Prosthesis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Classification Name:                | 21 CFR 888.3358 - Hip joint metal/polymer/metal semi-<br>constrained porous-coated uncemented prosthesis                                                                                                                                                                                                                                                                                                                                                                                            |
|                                     | 21 CFR 888.3353 - Hip joint metal/ceramic/polymer semi-<br>constrained cemented or nonporous uncemented prosthesis                                                                                                                                                                                                                                                                                                                                                                                  |
| Product Code:                       | LPH, LZO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Predicate Devices:                  | Responsive Orthopedics Total Hip Arthroplasty System<br>(K163585)                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Device Description:                 | The Medtronic Total Hip Arthroplasty System is a total hip<br>system for the restoration of alignment, stability range of<br>motion, and alleviation of pain, by replacing the articulating<br>surfaces of the hip joint.  The system includes femoral and<br>acetabular components. The implants are available in a variety<br>of sizes to accommodate varying patient anatomy.                                                                                                                    |
|                                     | New acetabular components including highly crosslinked<br>UHMWPE acetabular liners and porous coated cups are the<br>subject of this 510(k).                                                                                                                                                                                                                                                                                                                                                        |
| Indications for Use:                | The Medtronic Total Hip Arthroplasty System is indicated for<br>cementless use only in the following cases:<br>Noninflammatory degenerative joint disease including<br>osteoarthritis and avascular necrosis; Rheumatoid arthritis; Correction of functional deformity; Revision procedures where other treatments or<br>devices have failed; and Nonunions, femoral neck and trochanteric fractures of<br>the proximal femur with head involvement that are<br>unmanageable using other techniques |
|                                     | The Medtronic Total Hip Arthroplasty System Acetabular<br>Screws are intended for supplemental fixation of the<br>associated Medtronic Acetabular Cup.                                                                                                                                                                                                                                                                                                                                              |
| Technological Characteristics:      | The Medtronic Total Hip Arthroplasty System is similar to that<br>of the predicate devices, as it still has the same indications for<br>use and intended use of the predicate. The difference between<br>the two is the material change of the acetabular liner from a<br>conventional polyethylene to a highly crosslinked UHMWPE<br>liner.                                                                                                                                                        |
| Summary of Non-Clinical<br>Testing: | The following tests were completed to support the addition of<br>a highly crosslinked UHMWPE acetabular liner and cup:<br>Wear Testing per:  ISO 14242-1:2014 Implants for Surgery -<br>Wear of total hip-joint prostheses - Part 1:<br>Loading and displacement parameters for<br>wear-testing machines and corresponding<br>environmental conditions for test ISO 14242-2: 2016 Implants For Surgery -<br>                                                                                        |
| ο                                   | ISO 10993-17 Biological evaluation of medica<br>devices - Part 17: Establishment of allowable<br>limits for leachable substances                                                                                                                                                                                                                                                                                                                                                                    |
| ο                                   | ISO 10993-18, 2005, Biological Evaluation of<br>Medical Devices - Part 18: Chemical<br>Characterization of Materials                                                                                                                                                                                                                                                                                                                                                                                |
| -                                   | Material Characterization per:                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| ο                                   | ASTM F2565 Standard Guide For Extensively<br>Irradiation-Crosslinked Ultra-High Molecular<br>Weight Polyethylene Fabricated Forms For<br>Surgical Implant Applications                                                                                                                                                                                                                                                                                                                              |
| Summary of Clinical Tests:          | Clinical Testing was not required for these products.                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Conclusion:                         | Based on the indications for use, performance characteristics,<br>and bench testing completed on the Medtronic Total Hip<br>Arthroplasty System, the subject device is equivalent to the<br>predicate device.                                                                                                                                                                                                                                                                                       |

{4}------------------------------------------------

{5}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH/K183129](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH/K183129)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
