← Product Code [LPH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH) · K142119

# PROCOTYL L-O ACETABULAR SYSTEM (K142119)

_Microport Orthopedics, Inc. · LPH · Feb 27, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH/K142119

## Device Facts

- **Applicant:** Microport Orthopedics, Inc.
- **Product Code:** [LPH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH.md)
- **Decision Date:** Feb 27, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3358
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The PROCOTYL® L-O Acetabular System is intended for use in total hip arthroplasty for reduction or relief of pain and/ or improved hip function in skeletally mature patients with the following conditions: 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia: 2. inflammatory degenerative joint disease such as rheumatoid arthritis; 3. correction of functional deformity; and. 4. revision procedures where other treatments or devices have failed The PROCOTYL® L-O Acetabular System utilizes single use components. intended for use in conjunction with associated ceramic femoral heads as part of uncemented total hip arthroplasty.

## Device Story

PROCOTYL® L-O Acetabular System is an uncemented hip prosthesis consisting of acetabular cups, poly liners, and ceramic femoral heads. Used in total hip arthroplasty to relieve pain and improve function. Components are implanted by orthopedic surgeons in a clinical/hospital setting. System functions as a mechanical replacement for the hip joint; provides structural support and articulation. Benefits include restoration of joint mobility and pain reduction. Device is based on existing LINEAGE™ and DYNASTY® systems.

## Clinical Evidence

No clinical data provided. Safety and effectiveness supported by nonclinical bench testing, including static compression to failure (ISO 7206-10) and range of motion evaluation (ISO 21535).

## Technological Characteristics

Components made of Titanium Alloy (Ti6Al4V, ASTM F136/F620) with titanium plasma spray (ASTM F1580) or beaded coating (ASTM F1580/F67). Includes BIOLOX® delta Ceramic Femoral Head and RIM-LOCK A-CLASS® Poly Liner. Uncemented, semi-constrained prosthesis.

## Regulatory Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

## Predicate Devices

- LINEAGE™ Acetabular System ([K002149](/device/K002149.md) shells; [K052026](/device/K052026.md) liners)
- DYNASTY® Acetabular System ([K130376](/device/K130376.md) 28mm heads; [K140043](/device/K140043.md) 32-36mm heads)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of three human profiles facing right, arranged in a way that suggests unity and connection. The profiles are black and are set against a white background. Encircling the profiles are the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 27, 2015

MicroPort Orthopedics, Inc. Mr. Byron Ledbetter Regulatory Affairs Specialist II 5677 Airline Road Arlington, Tennessee 38002

Re: K142119 Trade/Device Name: PROCOTYL® L-O Acetabular System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, LZO Dated: January 20, 2015 Received: January 22, 2015

Dear Mr. Ledbetter:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

Page 2 - Mr. Byron Ledbetter

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

## K142119

Device Name

PROCOTYL® L-O Acetabular System

Indications for Use (Describe)

The PROCOTYL® L-O Acetabular System is intended for use in total hip arthroplasty for reduction or relief of pain and/ or improved hip function in skeletally mature patients with the following conditions:

1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia:

2. inflammatory degenerative joint disease such as rheumatoid arthritis;

3. correction of functional deformity; and.

4. revision procedures where other treatments or devices have failed

The PROCOTYL® L-O Acetabular System utilizes single use components. intended for use in conjunction with associated ceramic femoral heads as part of uncemented total hip arthroplasty.

## Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

## FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for MicroPort Orthopedics. The logo features a blue square with a white circle and a red diamond inside, followed by the text "MicroPort" in blue and "Orthopedics" in red. Below the logo is the text "MicroPort Orthopedics Inc." in black.

## 510(k) Summary

In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the use of the PROCOTYL® L-O Acetabular System.

| Submitted by:                        | MicroPort Orthopedics Inc.<br>5677 Airline Rd<br>Arlington, TN 38002<br>Phone: (866) 872-0211<br>Fax: (855) 446-2247            |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Date:                                | February 18, 2015                                                                                                               |
| Contact Person:                      | Byron Ledbetter<br>Regulatory Affairs Specialist II                                                                             |
| Proprietary Name of Modified Device: | PROCOTYL <sup>®</sup> L-O Acetabular System                                                                                     |
| Common Name:                         | acetabular shell, acetabular liner and femoral<br>head                                                                          |
| Classification Name and Reference:   | 888.3358 LPH<br>Hip joint metal/polymer/metal semi-<br>Constrained porous-coated<br>uncemented prosthesis<br>Class II           |
|                                      | 888.3353 LZO<br>Hip joint metal/ceramic/polymer semi<br>constrained cemented or nonporous,<br>uncemented prosthesis<br>Class II |
| Subject Product Code and Panel Code: | Orthopedics/87/LPH/LZO                                                                                                          |
| Predicate Devices:                   | LINEAGE™ Acetabular System (K002149<br>shells; K052026 liners)                                                                  |
|                                      | DYNASTY® Acetabular System (K130376<br>28mm heads; K140043 32-36mm heads)                                                       |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for MicroPort Orthopedics. The logo consists of a blue square with rounded corners, inside of which is a white circle containing a red diamond shape. To the right of the square is the text "MicroPort" in blue, with "Orthopedics" in red underneath. Below the logo is the text "MicroPort Orthopedics Inc." in black.

# Device Description

The PROCOTYL® L and O acetabular cup designs are based off the LINEAGE™ Acetabular System (K002149) which was designed from the TRANSCEND® product line. The subject PROCOTYL® acetabular system of this submission includes the PROCOTYL® L Acetabular Cup, the PROCOTYL® O Acetabular Cup, RIM-LOCK "A-CLASS®" Acetabular Poly Liner and BIOLOX® delta Ceramic Femoral Head.

# Intended Use

The PROCOTYL® L-O Acetabular System is intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:

- 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
- 2. inflammatory degenerative joint disease such as rheumatoid arthritis;
- 3. correction of functional deformity; and,
- 4. revision procedures where other treatments or devices have failed

The PROCOTYL® L-O Acetabular System utilizes single use components, intended for use in conjunction with associated ceramic femoral heads as part of uncemented total hip arthroplasty.

# Technological Characteristics of the Device

The indications for use of the PROCOTYL® L-O Acetabular System are identical to those for the predicate devices for the head (K130376 and K140043) and shells/liners (K002149). The subject devices are made from an identical Titanium Alloy, Ti6Al4V, wrought or forged (ASTM F136 or ASTM F620 respectively) and possess an identical titanium plasma spray coating (ASTM F1580) or beaded coating (ASTM F1580/F67) as the predicate device.

# Nonclinical Testing

The PROCOTYL® L-O Acetabular System ceramic heads were evaluated for static compression to failure and were deemed acceptable per ISO 7206-10. Additionally, the PROCOTYL® L-O Acetabular System was evaluated for range of motion and was deemed acceptable per ISO 21535.

# Clinical Testing

Clinical data was not provided for the subject devices.

# Conclusions

The indications for use and fundamental scientific technology of the PROCOTYL® L-O Acetabular System are identical to those of the predicate devices. The design features and materials of the subject devices are substantially equivalent to those of the predicate devices. The safety and effectiveness of the PROCOTYL® L-O Acetabular System are adequately supported by the substantial equivalence information, materials information, and nonclinical testing data provided within this premarket notification.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH/K142119](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH/K142119)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
