← Product Code [LPH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH) · K141676

# RENOVIS SURGICAL POROUS ACETABULAR CUP SYSTEM (K141676)

_Renovis Surgical Technologies, Inc. · LPH · Nov 4, 2014 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH/K141676

## Device Facts

- **Applicant:** Renovis Surgical Technologies, Inc.
- **Product Code:** [LPH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH.md)
- **Decision Date:** Nov 4, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3358
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Renovis Surgical Hip Replacement System is indicated for patients suffering from: 1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis; 2. Rheumatoid arthritis; 3. Correction of functional deformity; 4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement. unmanageable using other techniques; and 5. Revision procedures where other treatment or devices have failed.

## Device Story

Renovis Surgical Porous Acetabular Cup System consists of titanium, hemispherical acetabular shells; varying sizes (44-66 mm); options for no-hole or cluster-hole configurations. Used as part of total hip arthroplasty system; implanted by orthopedic surgeons in clinical/surgical settings. Device provides stable acetabular component for hip joint replacement; facilitates bone ingrowth via porous coating. Output is physical structural support for hip joint; aids in restoring joint function and mobility. Benefits include pain relief and improved joint mechanics for patients with degenerative or traumatic hip conditions.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included stereological evaluation of porous coatings (ASTM F1854-09), shear testing (ASTM F1044-05), tension testing (ASTM F1147-05), and disassembly force testing (ASTM F1820-97). Sterilization, shelf life, packaging, and shipping validation were performed per ISO 11137-2, ASTM F1980-07, ISO 11607-2, and ASTM D4169-09.

## Technological Characteristics

Hemispherical titanium alloy (ASTM F136-13, F620-11, F67-13) acetabular cups. Features porous coating for bone apposition. Dimensions: 44-66 mm outer diameter. Gamma sterilized. Mechanical fixation via press-fit/screws. Complies with ASTM F1854, F1044, F1147, F1820, and F983.

## Regulatory Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

## Predicate Devices

- Renovis A400 Surgical Hip Joint Replacement System (acetabular shells) ([K112897](/device/K112897.md))
- Renovis Tesera Trabecular Technology (T3) Acetabular Shell System ([K132312](/device/K132312.md))
- Encore Orthopedics Foundation Porous Coated Hemispherical Acetabular Cup ([K974093](/device/K974093.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure, possibly representing the department's focus on health and human well-being. The figure is composed of three overlapping profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 4, 2014

Renovis Surgical Technologies, Incorporated % Sharyn Orton, Ph.D. Senior Consultant MEDIcept, Incorporated 200 Homer Avenue Ashland, Massachusetts 01721

Re: K141676

Trade/Device Name: Renovis Surgical Porous Acetabular Cup System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH Dated: September 30, 2014 Received: October 3, 2014

Dear Dr. Orton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Lori A. Wiggins -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K141676

Device Name

Renovis Surgical Porous Acetabular Cup System

Indications for Use (Describe)

The Renovis Surgical Hip Replacement System is indicated for patients suffering from:

1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis;

2. Rheumatoid arthritis;

3. Correction of functional deformity;

4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement. unmanageable using other techniques; and

5. Revision procedures where other treatment or devices have failed.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### Traditional 510(k) Summary as required by 21 CFR 807.92(a) K141676

- A ) Submitted by: Renovis Surgical Technologies Inc. 1901 W. Lugonia Ave, Ste 340 Redlands, CA 92374 Phone: 909-557-2360 Fax: 909-307-8571
- Official Contact: Anthony DeBenedictis Vice President of Quality Assurance
- Consultant: Sharyn Orton, Ph.D. MEDIcept, Inc. 200 Homer Ave Ashland, MA 01721
- Date prepared: September 25, 2014
- B) Device Name: Prosthesis, hip, semi-constrained, metal/polymer, porous uncemented
	- Common Name: Total Hip Arthroplasty - Acetabular Components
	- Proprietary Name: Renovis Surgical Porous Acetabular Cup System
	- Device Class: Class II
	- Regulation number: 21 CFR 888.3358
	- Regulation name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis
	- Product code: LPH
	- Classification panel: Orthopedic
- C) Predicates:

| K112897 | Renovis A400 Surgical Hip Joint Replacement System (acetabular shells)      |
|---------|-----------------------------------------------------------------------------|
| K132312 | Renovis Tesera Trabecular Technology (T3) Acetabular Shell System           |
| K974093 | Encore Orthopedics Foundation Porous Coated Hemispherical<br>Acetabular Cup |

{4}------------------------------------------------

#### D) Device Description:

The Renovis Porous Acetabular Cup System ("Porous shell") are titanium, hemispherical cups consisting of varying sizes (outer diameter of 44 - 66 mm) and screw hole options (no hole shell and cluster hole shells) to accommodate patient anatomy and/or surgical technique. Instrumentation is provided with the system.

The Renovis Porous Acetabular Cup System is part of the Renovis Surgical Hip Replacement System and is to be used with Renovis Surgical Hip Replacement System acetabular and femoral components.

The Renovis Surgical Porous Cup System complies with the following material standards:

- o ASTM F136-13 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401) or
- ASTM F620-11 Standard Specification for Titanium Alloy Forgings for Surgical ● Implants in the Alpha Plus Beta Condition
- ASTM F67-13 Standard Specification for Unalloyed Titanium, for Surgical Implant ● Applications (UNS R50250, UNS R50400, UNS R50550 and UNS R50700)
- ASTM A564-13 Standard Specification for Hot-Rolled and Cold-Finished Age-● Hardening Stainless Steel Bars and Shapes
- E) Indications For Use:

The Renovis Surgical Hip Replacement System is indicated for patients suffering from:

- 1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
- 2. Rheumatoid arthritis;
- 3. Correction of functional deformity;
- 4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques; and
- 5. Revision procedures where other treatment or devices have failed.
- F) Substantial Equivalence Comparison and Discussion

The Renovis Surgical Porous Acetabular Cup System has similar Indications for Use, operating principle and basic design, sizes and geometries, biocompatible materials, and is offered gamma sterilized, like the predicate devices. The manufacturing processes used for the outer coating differ between the Renovis Surgical Porous Acetabular Cup System and two of the predicate devices. The Renovis Surgical Porous Acetabular Cup System meets the coating characterization criteria in 21 CFR 888.3358, and expected performance testing. The Renovis Surgical Porous Acetabular Cup System is substantially equivalent to the predicate devices.

{5}------------------------------------------------

- G) Performance Testing Bench
Performance testing was successfully completed.

The Renovis Surgical Porous Acetabular Cup System complies with:

- 21 CFR 888.3358 ●
- 0 FDA "Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement", April 8, 1994
- ASTM F1854-09 Standard Test Method for Stereological Evaluation of Porous Coatings ● on Medical Implants
- o ASTM F1472-08e1 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium Alloy for Surgical Implant Applications (UNS R56400)
- ASTM F1044-05 (2011) Standard Test Method for Shear Testing of Calcium Phosphate o Coatings and Metallic Coatings
- ASTM F1147-05 (2011) Standard Test Method for Tension Testing of Calcium Phosphate and Metallic Coatings
- ASTM F1820-97:2009 Standard Test Method for Determining the Forces for ● Disassembly of Modular Acetabular Devices

Sterilization, shelf life, packaging and shipping validation studies have been conducted. The Renovis Surgical Porous Acetabular Cup System complies with:

- o ISO 11137-2:2013 Sterilization of health care products -- Radiation -- Part 2: Establishing the sterilization dose
- ASTM F1980-07 (reapproved 2011) Standard Guide for Accelerated Aging of Sterile ● Barrier Systems for Medical Devices
- ISO 11607-2:2009: Packaging for terminally sterilized medical devices Part 2: ● Validation requirements for forming, sealing and assembly processes
- ISTA 2A Partial Simulation Performance Tests; Packaged-Products weighing 150 lb (68 kg) or Less
- ASTM D4169-09: Standard Practice for Performance Testing of Shipping Containers and Systems

## Conclusion

Renovis Surgical Porous Acetabular Cup System is substantially equivalent to the predicate devices, and does not raise new issues of safety or effectiveness.

- H) Other Standards
	- o ASTM F983-86 (Reapproved 2013) Standard Practice for Permanent Marking of Orthopaedic Implant Components
	- ASTM F565-04 (Reapproved 2013) Standard Practice for Care and Handling of ● Orthopedic Implants and Instruments

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH/K141676](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH/K141676)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
