← Product Code [LPH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH) · K132158

# NOVOHIP TOTAL HIP SYSTEM (K132158)

_Novosource, Inc. · LPH · Feb 21, 2014 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH/K132158

## Device Facts

- **Applicant:** Novosource, Inc.
- **Product Code:** [LPH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH.md)
- **Decision Date:** Feb 21, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3358
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

Hip joint arthroplasty

## Device Story

Non-cemented 4-part total hip replacement system; includes femoral stem, femoral head, acetabular poly liner, and acetabular metal shell. Femoral head articulates within poly acetabular component; poly component snaps into metal shell. Design restores natural hip rotation, extension, and flexion. Used by orthopedic surgeons in clinical settings for hip arthroplasty. Provides mechanical replacement of diseased or damaged hip joint surfaces to alleviate pain and restore function.

## Clinical Evidence

No clinical testing was required. Substantial equivalence supported by non-clinical bench testing, including fatigue performance (stem and neck), disassembly force, range of motion, burst strength (ceramic heads), rotational stability, torsional properties, pull-out strength, and assembly lever-out/push-out/torque-out testing.

## Technological Characteristics

Materials: Forged Ti 6AL 4V with Ti plasma spray/porous coating, CoCr, BIOLOX delta ceramic, and standard UHMWPE. Components: Femoral stem, femoral head (28, 32, 36 mm), acetabular poly liner, and acetabular shell (no-hole, cluster-hole, revision multi-hole). Mechanical fixation: Cementless. Design: Semi-constrained.

## Regulatory Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

## Predicate Devices

- Stelkast Provident Hip System ([K935484](/device/K935484.md), [K033944](/device/K033944.md))
- Smith & Nephew Biolox Delta Ceramic Femoral Head ([K083762](/device/K083762.md))
- Wright Medical Lineage Acetabular System ([K002149](/device/K002149.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) Summary

FEB 2 1 2014

# Applicant/Sponsor:

NovoSource, Inc. 714 East Monument Ave. Suite 220 Dayton, OH, 45402 937-531-6591

#### Contact Person:

Sharon Kvistad Navigant Consulting, LLC 9001 Wesleyan Road Suite 200 Indianapolis, IN 46268 317-228-8715 sharon.kvistad@navigant.com

Date of Preparation: February 18, 2014 .

### NAME AND CLASSIFICATION

| Proposed Trade Name:   | NovoHip Total Hip System                                                                                                                                             |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:           | Hip prosthesis                                                                                                                                                       |
| Classification Name:   | Hip joint metal/polymer/metal semi-constrained porous-coated<br>uncemented prosthesis per 21 CFR 888.3358                                                            |
|                        | Hip joint metal/ceramic/polymer semi-constrained cemented or<br>nonporous uncemented prosthesis per 21 CFR 888.3353                                                  |
| Device Product Code:   | LPH, LZO                                                                                                                                                             |
| Predicate Devices:     | Stelkast Provident Hip System (K935484, K033944)<br>Smith & Nephew Biolox Delta Ceramic Femoral Head (K083762)<br>Wright Medical Lineage Acetabular System (K002149) |
| Device Classification: | Class II                                                                                                                                                             |

#### DEVICE DESCRIPTION

The NovoHip Total Hip System is a non-cemented hip prosthesis that consists of a 4-part total hip replacement system including femoral stem, femoral head, acetabular poly liner, and acetabular metal (or shell) components. The femoral head component articulates within the poly acetabular component. The poly acetabular component snaps into the metal acetabular component. The design and sizing of the components correspond to natural hip anatomy to restore normal rotation, extension, and flexion.

{1}------------------------------------------------

# K132158 page 2/3

The femoral stem component is made from forged Ti 6AL 4V, with a Ti plasma spray coating. The unipolar femoral head component is made from CoCr, or BIOLOX® delta ceramic, in 28, 32, and 36 mm sizes.

The acetabular poly liner component is made of standard UHMWPE polyethylene in both hooded and non-hooded options. The acetabular poly liner component is offered with different inner and outer diameter combinations to accept various size uni-polar femoral heads and acetabular metal components.

The acetabular metal (or shell) component is made from forged Ti 6AL 4V with a Ti porous coating. It is available in no-hole, cluster-hole, and revision multi-hole styles.

# INTENDED USE

Hip joint arthroplasty

### INDICATIONS FOR USE

NovoSource hip implant components are indicated for use in cementless reconstruction of the articulating surface of femoral and/or acetabular portions of the hip that are severely disabled and/or very painful as a result of:

- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis
- Rheumatoid arthritis or traumatic arthritis
- · Correction of functional deformity
- · Non-union femoral neck fracture
- Trochanteric fractures of the proximal femur with head involvement which is unmanageable using other techniques.

The components can be used for primary hip arthroplasty or for revision of a failed total hip arthroplasty.

#### TECHNOLOGICAL CHARACTERISTICS

The NovoHip Total Hip System has the same intended use as the predicate devices. The NovoHip Total Hip System has similar indications for use as the as the predicate devices. The NovoHip Total Hip System is manufactured from the same materials as the predicate devices. The range of sizes of the NovoHip Total Hip System is similar to the predicate devices.

#### SUMMARY OF STUDIES

# Non-Clinical Testing

The NovoHip Total Hip System underwent the following testing:

- Fatigue Performance Test for NovoHip Stem .
- Fatigue Performance Test for the Neck Portion of the NovoHip Stem ●
- . Disassembly Force Test for the NovoHip Neck Taper/Femoral Head Interface
- . Range of Motion Test for the NovoHip Total Hip System

{2}------------------------------------------------

- . Burst Strength Test for NovoHip Ceramic Femoral Heads (Static Compression)
- . Cyclic Fatigue Test for NovoHip Ceramic Femoral Heads (Cyclic Compression)
- Post-Cyclic Fatigue Burst Test for NovoHip Ceramic Femoral Heads (Static Compression) .
- . Pull-Off Test for NovoHip Ceramic Femoral Heads
- . Rotational Stability Test for NovoHip Ceramic Femoral Heads
- . Torsional Properties Test for NovoHip Bone Screws
- . Driving Torque Test for NovoHip Bone Screws
- . Axial Pull-Out Strength Test for NovoHip Bone Screws
- . Lever-Out Test for NovoHip Acetabular Shell/Liner Assembly
- Torque-Out Test for NovoHip Acetabular Shell/Liner Assembly .
- . Push-In Test for NovoHip Acetabular Shell/Liner Assembly
- . Push-Out Test for NovoHip Acetabular Shell/Liner Assembly

All devices met the required performance specifications for testing and are considered equivalent to the predicate devices.

#### Clinical Testing

No clinical testing was required.

#### CONCLUSION

Based on testing results and the comparisons provided, the NovoHip Total Hip System is considered substantially equivalent to the Stelkast Provident Hip System and the Wright Medical Lineage in material, construction and performance characteristics.

{3}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002

February 21, 2014

Novosource, Incorporated % Ms. Sharon Kvistad Associate Director Navigant Consulting, LLC 9001 Wesleyan Road, Suite 200 Indianapolis, Indiana 46268

Re: K132158

Trade/Device Name: NovoHip Total Hip System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, LZO Dated: January 13, 2014 Received: January 14, 2014

Dear Ms. Kvistad:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{4}------------------------------------------------

# Page 2 - Ms. Sharon Kvistad

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Vincent Dogylin -S

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

#### Indications for Use

510(k) Number (if known): K132158

Device Name: NovoHip Total Hip System

:

Indications for Use:

NovoSource hip implant components are indicated for use in cementless reconstruction of the articulating surface of femoral and/or acetabular portions of the hip that are severely disabled and/or very painful as a result of:

- Non-inflammatory degenerative joint disease including osteoarthritis and avascular . necrosis
- Rheumatoid arthritis or traumatic arthritis .
- Correction of functional deformity .
- . Non-union femoral neck fracture
- Trochanteric fractures of the proximal femur with head involvement which is . unmanageable using other techniques.

The components can be used for primary hip arthroplasty or for revision of a falled total hip arthroplasty.

Prescription Use _____X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of

# Elizabet标办公Frank -S

Division of Orthopedic Devices

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH/K132158](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH/K132158)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
