K033580 · Nexmed, Inc. · LPH · Aug 5, 2004 · Orthopedic
Device Facts
Record ID
K033580
Device Name
NEXFLEX TOTAL HIP SYSTEM
Applicant
Nexmed, Inc.
Product Code
LPH · Orthopedic
Decision Date
Aug 5, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
1) When used as a hemi-hip replacement system, it is intended for osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal involvement of the corresponding acetabulum, femoral head or neck fractures, aseptic necrosis of the femoral head, previous failed hip arthroplasty where there is evidence of sufficient bone quality to adequately set the implant. 2) When used as a total hip replacement system, it is intended for osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal involvement of the corresponding acetabulum, femoral head or neck fractures, aseptic necrosis of the femoral head, previous failed hip arthroplasty where there is evidence of sufficient bone quality to adequately set the implant. 3) In addition, the NexFlex Total Hip System is intended for cases where alternative modes of treatment appear less preferable and the associated risks of a total hip replacement are thought to be acceptable. It is intended for severely disabled joints, which could result from arthritis or late stages of avascular necrosis and revisions of unsuccessful acetabular cup arthroplasty and/or femoral procedure.
Device Story
NexFlex Total Hip System is a sterile orthopedic implant system comprising femoral and acetabular components. Designed for total or hemi-hip replacement surgery to restore range of motion and treat joint deformities. Features hydroxyapatite (HA) coating on stems and cups to facilitate fixation. Used by orthopedic surgeons in clinical/hospital settings. Implants replace damaged bone/cartilage, addressing arthritis, necrosis, or failed prior arthroplasty. Benefits include improved joint function and pain relief for patients with severe joint disability.
Clinical Evidence
Bench testing only. Fatigue testing and microstructure examination were performed to evaluate mechanical performance and biocompatibility. Results demonstrated that the device is at least comparable to predicate devices.
Technological Characteristics
Total or hemi-hip replacement system consisting of femoral and acetabular implants. Features hydroxyapatite (HA) coating. Materials and specific ASTM standards are not detailed in the provided text. Device is a sterile, semi-constrained or uncemented/cemented prosthesis. No software, electronics, or connectivity features.
Indications for Use
Indicated for patients with osteo-, rheumatoid, or post-traumatic arthritis of the hip, femoral head/neck fractures, aseptic necrosis of the femoral head, or failed prior hip arthroplasty requiring revision. Applicable for hemi-hip or total hip replacement in patients with sufficient bone quality. Also indicated for severely disabled joints due to arthritis or late-stage avascular necrosis where alternative treatments are less preferable.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Submission Summary (Full Text)
{0}------------------------------------------------
AUG - 5 2004
NexFlex™ Total Hip System 510(k) SUMMARY November 2003
K033580
page 1 of 7
- l. Company: Nexmed, Inc. 6110 Corte Del Cedro Carlsbad. CA 92009 USA (760) 431-9286
- II. Contact Person: Ellen Yarnall, Director of Regulatory Affairs
- III. Trade/Proprietary Name: NexFlex™ Total Hip System with HA Coating
# IV. Product Description:
The NexFlex™ Total Hip System is a sterile total or hemi-hip replacement system. It consists of a series of femoral and acetabular implants that are used to help restore patient range of motion and aid in the treatment of other deformities as histor in the Indications for Use.
The purpose of this 510(k) is to provide for HA coated hip sterns and acetabular cups. This submission also includes additions / modifications to the and accedural under formal change control procedures.
#### V. Indications for Use:
- 1) When used as a hemi-hip replacement system, it is intended for osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal involvement of the corresponding acetabulum, femoral head or neck fractures, aseptic necrosis of the femoral head, previous failed hip arthroplasty where there is evidence of sufficient bone quality to adequately set the implant.
- 2) When used as a total hip replacement system, it is intended for osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal involvement of the corresponding acetabulum, femoral head or neck fractures, aseptic necrosis of the femoral head, previous failed hip arthroplasty where there is evidence of sufficient bone quality to adequately set the implant.
- 3) In addition, the NexFlex Total Hip System is intended for cases where alternative modes of treatment appear less preferable and the associated risks of a total hip replacement are thought to be acceptable. It is intended for severely disabled joints, which could result from arthritis or late stages of avascular necrosis and revisions of unsuccessful acetabular cup arthroplasty and/or femoral procedure,
{1}------------------------------------------------
# NexFlex™ Total Hip System 510(k) SUMMARY November 2003
c 33586
page 2 of 2
### VI. Substantial Equivalence:
The NexFlex Total Hip System with HA coated components is substantially equivalent to various total hip and hemi-hip systems commercially available.
### VII. Performance Data:
Fatigue testing and microstructure examination the NexFlex Total Hip System was previously submitted. The test results demonstrated that the mechanical performance and biocompatibility characteristics are at least comparable to, if not better than, those of the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG - 5 2004
Ms. Ellen A. Yarnall Director of Regulatory Affairs Nexmed, Inc. 6110 Corte Del Cedro Carlsbad, California 92009
Re: K033580
Trade/Device Name: Nexflex Total Hip System Trade Dovice Name: 21 CFR 888.3358; 21 CFR 888.3350; 21 CFR 888.3390 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis; Hip joint metal/polymer semi-constrained cemented prosthesis; Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis
Regulatory Class: II Product Code: LPH, JDI, KWY Dated: July 5, 2004 Received: July 7, 2004
Dear Ms. Yarnall:
We have reviewed your Section 510(k) premarket notification of intent to market the device we have reviewed your becatermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Incrisiate commerce prior to has ray, been reclassified in accordance with the provisions of Amendinents, or to do rough that nortic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include Controls provisions of the 1100 Justing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (111), it hay of bund in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not r lease be actised that the a determination that your device complies with other requirements mean that 127 was Federal statutes and regulations administered by other Federal agencies.
{3}------------------------------------------------
Page 2 - Ms. Ellen A. Yarnall
You must comply with all the Act's requirements, including, but not limited to: registration 1 ou intest compry with and 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if requirelate as electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) I rais reter notification. The FDA finding of substantial equivalence of your device to a premiurker notification - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), If you desire specific at not Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division may ordain onufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark N. Milkin
Celia Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
K033580 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: NexFlex™ Total Hip System
Indications for Use:
- 1) When used as a hemi-hip replacement system, it is intended for osteo-, rheumatoid, wnen ased as a nom hip of the hip with minimal involvement of the corresponding and post-tradmidtio arthatis or neck fractures, aseptic necrosis of the femoral head, acctabolumi, fornoral head his where there is evidence of sufficient bone quality to adequately set the implant.
- 2) When used as a total hip replacement system, it is intended for osteo-, rheumatoid, with as ou a tetal hittis of the hip with minimal involvement of the corresponding and post tradinatio arthatio a neck fractures, aseptic necrosis of the femoral head, abotabulari, to hip arthroplasty where there is evidence of sufficient bone quality to adequately set the implant.
- 3) In addition, the NexFlex Total Hip System is intended for cases where alternative in addition, the Noxt Topear less preferable and the associated risks of a total hip modes of treatment are thought to be acceptable. It is intended for severely disabled joints, ropidoomerit and the arthritis or late stages of avascular necrosis and revisions of unsuccessful acetabular cup arthroplasty and/or
plasty and/or femoral procedure.
Muhle n Millikun
ral. Restorative, and Neurological Devices
Prescription Use (Part 21 CFR 801 Subpart D)
**510(k) Number***_*_
AND/OR
Over-The Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.