← Product Code [LPH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH) · K023375

# MODIFICATION TO ESOP CO-CR FEMORAL HEADS (K023375)

_Pro-Active Healthcare · LPH · Oct 25, 2002 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH/K023375

## Device Facts

- **Applicant:** Pro-Active Healthcare
- **Product Code:** [LPH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH.md)
- **Decision Date:** Oct 25, 2002
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3358
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

Prosthetic replacement of the femoral head associated with hip prosthesis, which cone is compatible with the head's cone and with a prosthetic acetabular cup, whose diameter is compatible with the head.

## Device Story

ESOP® Co-Cr Femoral Head is a metallic prosthetic component for hip arthroplasty. Device attaches to femoral stems via machined cones (12/14 or 10/12 tapers). Available in multiple diameters and insertion depths (short, medium, long, extra-long collars) to adjust neck length. Used by orthopedic surgeons in clinical settings for full or intermediary hip replacement. Provides structural replacement of the femoral head to restore joint function in patients with arthritis, trauma, or necrosis. Benefits include improved mobility and pain relief through joint reconstruction.

## Clinical Evidence

Bench testing only. Verification and validation activities were performed to demonstrate safety and effectiveness.

## Technological Characteristics

Material: Chrome cobalt. Diameters: 22, 28, 32 mm. Taper/Cone dimensions: 12/14 (5°43') and 10/12 (6°). Configurations: Short, medium, long, and extra-long collars. Intended for cementless application.

## Regulatory Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

## Predicate Devices

- ESOP® Co-Cr Femoral Heads ([K021109](/device/K021109.md))
- Total Hip Femoral Head - 12/14 Taper ([K021673](/device/K021673.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Applicant: Pro-Active

## KO23375

Page 1 of 2

## 510 (K) SUMMARY – ESOP® Co-Cr Femoral Head

OCT 2 5 2002

| Submitter name:                                                                    | Fournitures Hospitalières Industrie                                                                                                                                                                                                                                                                                                                                                                                   |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter address:                                                                 | 6 Rue Nobel, Z.I. de Kernevez<br>QUIMPER, France 29000                                                                                                                                                                                                                                                                                                                                                                |
| Contact person:                                                                    | Mrs Christine QUENDEZ                                                                                                                                                                                                                                                                                                                                                                                                 |
| Phone Number:<br>Fax Number:                                                       | +33.2.98.55.68.95<br>+33.2.98.53.42.13                                                                                                                                                                                                                                                                                                                                                                                |
| Date prepared:                                                                     | October 4, 2002                                                                                                                                                                                                                                                                                                                                                                                                       |
| Device Trade Name:                                                                 | ESOP® Co-Cr Femoral Head                                                                                                                                                                                                                                                                                                                                                                                              |
| Device common name:                                                                | Femoral head                                                                                                                                                                                                                                                                                                                                                                                                          |
| Classification name:                                                               | Hip joint metal/polymer/metal semi-constrained porous-<br>coated uncemented prosthesis                                                                                                                                                                                                                                                                                                                                |
| Predicate Device:                                                                  | ESOP® Co-Cr Femoral Heads (K021109), Fournitures<br>Hospitalières Industrie<br>Total Hip Femoral Head - 12/14 Taper (K021673), Smith &<br>Nephew, Inc.                                                                                                                                                                                                                                                                |
| Device description:                                                                | The Metallic Femoral Heads are intended to be adapted to<br>special cones machined for this purpose. Femoral head<br>are available in several diameters for use in full hip or<br>intermediary replacement. The length of the neck on the<br>femoral stem can be adjusted by using the heads available<br>in different insertion depths (short, medium, long and extra-<br>long collars), and of different diameters. |
| Intended use:                                                                      | Prosthetic replacement of the femoral head associated<br>with hip prosthesis, which cone is compatible with the<br>head's cone and with a prosthetic acetabular cup,<br>whose diameter is compatible with the head.                                                                                                                                                                                                   |
| Device Technological<br>Characteristics and<br>Comparison to Predicate<br>Devices: | Femoral heads made of chrome cobalt (Ø 22, 28 & 32<br>mm), short collar, medium collar, long collar & extra-long<br>collar with cone dimensions 12/14, 5°43' & 10/12, 6°. The<br>identified predicate devices have the same intended use,<br>are made of the same material, have the same size, and<br>collar + cone dimensions are of the same order of<br>magnitude.                                                |
| Performance Data:                                                                  | Verification/validation and design control activities<br>demonstrate the safety and effectiveness of the ESOP®<br>Co-Cr Femoral Heads.                                                                                                                                                                                                                                                                                |
| Conclusion:                                                                        | The ESOP® Co-Cr Femoral Heads are substantially<br>equivalent to predicate devices in terms of intended use,<br>safety, and effectiveness.                                                                                                                                                                                                                                                                            |

{1}------------------------------------------------

KOZ3375

Page 2 of 2

.

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 00 Corporate Boulevard ockville MD 20850

Dr. Andre Weith Director, Pro-Active Healthcare c/o Pharmanet, Inc. 815 Connecticut Ave., NW Suite 800 Washington, DC 20006

Re: K023375

Trade/Device Name: ESOP® Co-Cr Femoral Heads Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH Dated: October 4, 2002 Received: October 8, 2002

Dear Dr. Weith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Dr. Andre Weith -

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Mark A. Millman

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known):

K0233345

Device Name:

ESOP® Co-Cr Femoral Heads

Indications for Use:

The ESOP® Co-Cr Femoral Heads are intended for use in degenerative and inflammatory arthritis of the hip joint, trauma, non-acute femoral neck fracture, revision of previously failed hip arthroplasties, and idiopathic avascular (osteo) necrosis where radiographic evidence shows there is sufficient sound bone to seat the prosthesis. This device is intended for cementless application.

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH; Office of Device Evaluation (ODE)

| Prescription Use     | OR | Over-The-Counter Use |
|----------------------|----|----------------------|
| (Per 21 CFR 801.109) |    |                      |

(Division Sign-Off)

Division of General, Restorative

and Neurological Devices

(Optional Format 1-2-96)

| 510(k) Number | K023373 |
|---------------|---------|
|---------------|---------|

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH/K023375](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/LPH/K023375)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
