K061956 · Artimplant AB · KYI · Jun 7, 2007 · Orthopedic
Device Facts
Record ID
K061956
Device Name
ARTELON STT SPACER
Applicant
Artimplant AB
Product Code
KYI · Orthopedic
Decision Date
Jun 7, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3770
Device Class
Class 2
Attributes
Therapeutic
Intended Use
ARTELON STT Spacer is intended to be implanted into the scaphotrapeziotrapezioidal (STT) joint as an interpositional spacer between the scaphoid bone and the trapezialtrapezioid bones. The device is intended to be used in thumb disabilities caused by osteoarthritis.
Device Story
ARTELON STT Spacer is a woven, L-shaped, interpositional implant; designed for surgical implantation into the scaphotrapeziotrapezioidal (STT) joint. Acts as a space-filling device between scaphoid and trapezial/trapezioid bones to treat thumb disabilities resulting from osteoarthritis. Implanted by surgeons in a clinical/OR setting. Provides mechanical spacing to alleviate joint dysfunction. Supplied sterile for single-patient use.
Indicated for patients with thumb disabilities caused by osteoarthritis requiring interpositional spacing in the scaphotrapeziotrapezioidal (STT) joint.
Regulatory Classification
Identification
A wrist joint carpal trapezium polymer prosthesis is a one-piece device made of silicone elastomer or silicone elastomer/polyester material intended to be implanted to replace the carpal trapezium bone of the wrist.
K964381 — AVANTA ORTHOPAEDICS TRAPEZIAL IMPLANT · Avanta Orthopaedics, Inc. · Jan 22, 1997
K965204 — AVANTA ORTHOPAEDICS TRAPEZIUM IMPLANT · Avanta Orthopaedics, Inc. · Jun 3, 1997
K092548 — EXTREMITY MEDICAL TRAPEZIUM PROSTHESIS · Extremity Medical, LLC · Dec 15, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
Kobl956 page 14
# SECTION 5. 510(K) SUMMARY
### Summary of Safety and Effectiveness
JUN 2 7 2007
In accordance with 21 CFR 807.92, the following information constitutes Artimplant AB's summary for the Artelon® STT Spacer.
| SUBMITTER'S NAME: | Artimplant AB |
|---------------------|---------------------------------------------------|
| ADRESS: | Hulda Mellgrens gata 5, SE-421 32 Västra Frölunda |
| | Sweden |
| CONTACT PERSON: | Ajrulla Zuta |
| TELEPHONE NUMBER: | +46 31 7465600 |
| FAX NUMBER: | +46 31 7465660 |
| DATE OF SUBMISSION: | 07/10/2006 |
#### 1. Identification of device
Proprietary Name: Artelon® STT Spacer Common Name: Trapezium Polymer Prosthesis Classification Status: 888.3770 Product Code: KYI
#### 2. Equivalent devices
We believe that the ARTELON STT Spacer is substantially equivalent to: K040070 Artelon® CMC Spacer Artimplant AB Pre-amendment device, Swanson Trapezium Implant, Dow Corning Wright
#### 3. Description of the Device
ARTELON STT Spacer is a woven one-piece, L-shaped implant made of ARTELON, a polycaprolactone based polyurethaneurea.
#### Indications for use 4.
ARTELON STT Spacer is intended to be implanted into the scaphotrapeziotrapezioidal (STT) joint as an interpositional spacer between the scaphoid bone and the trapezioid bones. The device is intended to be used in thumb disabilities caused by osteoarthritis.
#### ડ. Comparison to predicate device.
The predicate device, Swanson Trapezium Implant (pre-amendment device) is similar to ARTELON STT Spacer in that it is used as a space filling device for treatment of osteoarthritis and articulates against the STT joint. The ARTELON CMC Spacer is also a spacing device, but intended for implantation in the CMC-1 joint as an interpositional spacer.
All three devices are made of elastomers. The ARTELON STT Spacer is made of the same material, ARTELON, and is of the same basic design as the ARTELON CMC Spacer (K040070).
All devices are supplied sterile and for single patient use.
{1}------------------------------------------------
page 2 of 2
#### 6. Discussion of performance testing.
A collection of tests have been successfully completed. The results from the performance testing demonstrate that ARTELON STT Spacer provides appropriate assurance of safety and effectiveness.
#### 7. Conclusion
Based on comparison to the predicate devices, ARTELON STT Spacer is substantially equivalent to the legally marketed devices and presents no new concerns of safety and effectiveness.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with its wings spread, symbolizing protection and care. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA", which is arranged in a circular pattern around the left side of the logo.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Artimplant AB c/o M Squared Associates, Inc. Ms. Terry Sheridan Powell Senior Project Manager 719 A Street NE Washington DC 20002
JUN 2 7 2007
Re: K061956 Trade/Device Name: Artelon® STT Spacer Regulation Number: 21 CFR 888.3770 Regulation Name: Wrist joint carpal trapezium polymer prosthesis Regulatory Class: Class II Product Code: KYI Dated: May 15, 2007 Received: May 17, 2007
Dear Ms. Powell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Terry Sheridan Powell
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or 240-276-3150 or on the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark A. Mcllarian
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indication for Use
510(k) Number (if known): K061956
Device Name: ARTELON STT Spacer
Indication For Use:
ARTELON STT Spacer is intended to be implanted into the scaphotrapeziotrapezioidal (STT) joint as an interpositional spacer between the scaphoid bone and the trapezialtrapezioid bones. The device is intended to be used in thumb disabilities caused by osteoarthritis.
Prescription Use X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use _ (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Mahal Millers
K061956
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.