← Product Code [KYI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KYI) · K061954

# ARTELON CMC SPACER ARTHRO (K061954)

_Artimplant AB · KYI · Jun 1, 2007 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KYI/K061954

## Device Facts

- **Applicant:** Artimplant AB
- **Product Code:** [KYI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KYI.md)
- **Decision Date:** Jun 1, 2007
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3770
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The ARTELON CMC Spacer Arthro is intended to be implanted into the first carpometacarpal joint (CMC-I) as an interpositional spacer between the trapezial bone and the first metacarpal bone. The device is intended to be used in thumb disabilities caused by osteoarthritis.

## Device Story

Woven textile interpositional spacer; made of degradable polycaprolactone-based polyurethaneurea fibers (ARTELON). Implanted into first carpometacarpal (CMC-I) joint; vertical portion separates arthritic joint surfaces; horizontal portion fixated using commercially available suture anchors and non-absorbable sutures. Single-use; sterile. Used by orthopedic surgeons to treat thumb osteoarthritis; provides mechanical separation of bones to reduce joint pain and improve function.

## Clinical Evidence

Bench testing only. Testing performed in accordance with ISO 10993 (biocompatibility), ISO 14971 (risk management), ISO 11137 (sterilization), ISO 11607, ISO 11737-1, AAMI TIR27:2001, and EN868-1. No clinical data provided.

## Technological Characteristics

Material: Woven polycaprolactone-based polyurethaneurea fibers. Form factor: Interpositional spacer with vertical and horizontal wings. Fixation: Suture anchors and USP size 1 non-absorbable sutures. Sterilization: Double sterile barrier. Standards: ISO 10993, ISO 14971, ISO 11137, ISO 11607, ISO 11737-1, AAMI TIR27:2001, EN868-1.

## Regulatory Identification

A wrist joint carpal trapezium polymer prosthesis is a one-piece device made of silicone elastomer or silicone elastomer/polyester material intended to be implanted to replace the carpal trapezium bone of the wrist.

## Predicate Devices

- ARTELON CMC Spacer ([K040070](/device/K040070.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## Artimplant AB: Artelon CMC Spacer Arthro- K061954

## 510(k) SUMMARY

## Summary of Safety and Effectiveness

In accordance with 21 CFR 807.92, the following information constitutes Artimplant AB's summary for the ARTELON CMC Spacer Arthro.

SUBMITTER'S NAME Artimplant AB ADDRESS: Hulda Mellgrens gata 5. SE-421 32 Västra Frölunda. Sweden CONTACT PERSON: Ajrulla Zuta TELEPHONE NUMBER: +46 31 7465600 FAX NUMBER: DATE OF SUBMISSION: July 10, 2006

#### 1. Identification of device

Proprietary Name: ARTELON CMC Spacer Arthro Common Name: Prosthesis, wrist, carpal trapezium Classification Status: Class II per Regulation §888.3770 Product Code: KYI, wrist joint carpal trapezium polymer prosthesis

## 2. Equivalent device

ARTELON CMC Spacer, K040070, 09/21/2004.

#### 3. Description of the Device

ARTELON CMC Spacer Arthro is a woven textile device made of ARTELON fibres. ARTELON is a biocompatible, degradable material made of polycaprolactone-based polyurethaneurea.

ARTELON CMC Spacer Arthro is designed to act as an interposition part in the joint. The purpose of the vertical portion is to separate the surfaces of the arthritic joint from each other. The purpose of the horizontal portion of ARTELON CMC Spacer Arthro is to fixate the device.

ARTELON CMC Spacer Arthro is a single-use device packaged in double sterile barriers. The ARTELON CMC Spacer Arthro wings are fixed in place using commercially available suture anchors (non-absorbable, diameter approximately 2 mm) in combination with two double-armed class I nonabsorbable USP size 1 sutures.

{1}------------------------------------------------

## 4. Intended Use

The ARTELON CMC Spacer Arthro is intended to be implanted into the first carpometacarpal joint (CMC-I) as an interpositional spacer between the trapezial bone and the first metacarpal bone.

The device is intended to be used in thumb disabilities caused by osteoarthritis. This is the same intended use as previously cleared for the ARTELON CMC Spacer, K040070.

## 5. Discussion of Performance Testing

Performance bench testing was utilized needed to demonstrate Substantial Equivalence. The new product met the same release specifications as the previously cleared device. This fact, in conjunction with a comparison of dimensions, materials, basic design, and intended use demonstrated Substantial Equivalence.

A collection of tests have been performed in accordance with:

- ISO 10993 standards
- ISO 14971
- ISO 11137 ●
- . ISO 11607
- ISO 11737-1 .
- AAMI TIR27:2001 .
- EN868-1 �

#### 6. Conclusion

The ARTELON CMC Spacer Arthro described in the submission is substantially equivalent to the predicate device, and presents no new concerns about safety and effectiveness. Additionally, the devices have identical indication to the predicate device, and the labeling is consistent both with FDA guidance as well as current medical practice.

{2}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Artimplant AB c/o Ms. Terry Sheridan Powell Senior Project Manager M Squared Associates, Inc. 719 A Street, NE Washington, DC 20002

JUN - 1 2007

Re: K061954

Trade Name: ARTELON CMC Spacer Arthro Regulation Number: 21 CFR 888.3770 Regulation Name: Wrist joint carpal trapezium polymer prosthesis Regulatory Class: Class II Product Code: KYI Dated: April 24, 2007 Received: May 4, 2007

Dear Ms. Powell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Terry Sherdan Powell

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark A. Milburn

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(K) Number (if known): K061954

Device Name: Artelon® CMC Spacer Arthro

Indications for Use:

Artelon® CMC Spacer Arthro intended to be implanted into first carpometacarpal joint (CMC-I) as an interpositional spacer between the trapezial bone and the first metacarpal bone. The device is intended to be used in thumb disabilities caused by osteoarthritis.

Prescription Use X AND/OR

(Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Q gvice Evaluation (ODE) ttice of sion Sign-Off) (Division of General, Restorative, and Neurological Devi-

Page 1 of 1

**510(k) Number**

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KYI/K061954](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KYI/K061954)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
