Tornier Humeral Reconstruction System Max (Tornier HRS Max)

K251686 · Tornier, Inc. · KWT · Oct 30, 2025 · Orthopedic

Device Facts

Record IDK251686
Device NameTornier Humeral Reconstruction System Max (Tornier HRS Max)
ApplicantTornier, Inc.
Product CodeKWT · Orthopedic
Decision DateOct 30, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3650
Device ClassClass 2
AttributesTherapeutic

Intended Use

In Anatomic: Tornier HRS Max humeral assembly can be used for anatomic arthroplasty when used with a humeral head. It can be used as a hemiarthroplasty, if the natural glenoid provides a sufficient bearing surface, or in conjunction with a glenoid, as a total replacement. Tornier HRS Max is indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and with an intact or reconstructable rotator cuff with pain disabled by: • Rheumatoid arthritis • Non-inflammatory degenerative joint disease (i.e., osteoarthritis and avascular necrosis) • Fractures of the humeral head • Post-Traumatic arthritis • Oncology applications including bone loss due to tumor resection • Significant humeral resection where adequate fixation can be achieved In Reverse: Tornier HRS Max is indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and with massive and non-repairable rotator cuff-tear with pain disabled by: • Rheumatoid arthritis • Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis) • Correction of functional deformity • Fractures of the humeral head • Post-Traumatic arthritis • Massive and non-repairable rotator cuff tear • Oncology applications including bone loss due to tumor resection. • Significant humeral resection where adequate fixation can be achieved The reversed tray and polyethylene insert are indicated for use in the conversion from an anatomic to reversed shoulder arthroplasty without the removal of a well-fixed humeral assembly during revision surgery for patients with a functional deltoid muscle.

Device Story

Modular shoulder prosthesis system; addresses massive proximal humeral bone loss; provides deltoid wrapping. Configurations: anatomic, hemiarthroplasty, or reversed. Components: humeral stem (cemented/cementless), humeral heads (titanium/cobalt alloy), glenoid components (cemented), glenoid sphere (screw-fixed). Used in orthopedic surgery; implanted by surgeons. Titanium heads indicated for patients with cobalt alloy sensitivity. System allows conversion from anatomic to reverse arthroplasty during revision without removing well-fixed humeral stem. Benefits: restoration of joint function, pain relief, and structural support in complex oncology or revision cases.

Clinical Evidence

No clinical studies were performed. Substantial equivalence supported by non-clinical bench testing, including fatigue testing with corrosion evaluation, pull-out and torque-out testing, range of motion testing, biocompatibility, packaging/shelf-life, sterilization, endotoxin, and MRI compatibility evaluations.

Technological Characteristics

Modular shoulder prosthesis; materials include metal and polymer. Humeral stem (cemented/cementless), all-poly glenoid (cemented), glenoid sphere (screw-fixed). Titanium humeral heads available for cobalt-sensitive patients. Single-use components. Sterilization required.

Indications for Use

Indicated for shoulder joint replacement in patients with functional deltoid muscle. Anatomic configuration: intact/reconstructable rotator cuff; indicated for RA, non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), humeral head fractures, post-traumatic arthritis, oncology/tumor resection, or significant humeral resection. Reverse configuration: massive/non-repairable rotator cuff tear; indicated for RA, non-inflammatory degenerative joint disease, functional deformity correction, humeral head fractures, post-traumatic arthritis, oncology/tumor resection, or significant humeral resection. Reversed tray/insert indicated for conversion from anatomic to reverse arthroplasty without removing well-fixed humeral assembly.

Regulatory Classification

Identification

A shoulder joint metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).

Special Controls

*Classification.* Class II. The special controls for this device are:(1) FDA's: (i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ” (ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),” (iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,” (iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and (v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,” (2) International Organization for Standardization's (ISO): (i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-Aluminum 4-Vandium Alloy,” (ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-Chromium-Molybdenum Casting Alloy,” (iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-Chromium-Molybdenum Alloy,” (iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,” (v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,” (vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and (vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and (3) American Society for Testing and Materials': (i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,” (ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,” (iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,” (iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,” (v) F 1108-97 “Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,” (vi) F 1147-95 “Test Method for Tension Testing of Porous Metal Coatings,” (vii) F 1378-97 “Specification for Shoulder Prosthesis,” and (viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION October 30, 2025 Tornier, Inc. Stefanie Tarara Principal Regulatory Affairs Specialist 10801 Nesbitt Avenue South Bloomington, Minnesota 55441 Re: K251686 Trade/Device Name: Tornier Humeral Reconstruction System Max (Tornier HRS Max) Regulation Number: 21 CFR 888.3650 Regulation Name: Shoulder joint metal/polymer non-constrained cemented prosthesis Regulatory Class: Class II Product Code: KWT, KWS, HSD, PHX Dated: October 2, 2025 Received: October 3, 2025 Dear Stefanie Tarara: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251686 - Stefanie Tarara Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K251686 - Stefanie Tarara Page 3 Sincerely, **Farzana Sharmin -S** Digitally signed by Farzana Sharmin -S Date: 2025.10.30 18:07:56 -04'00' Farzana Sharmin, PhD Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} Tornier Humeral Reconstruction System Max (Tornier HRS Max) Page 11 of 43 | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251686 | ? | | Please provide the device trade name(s). | | ? | | Tornier Humeral Reconstruction System Max (Tornier HRS Max) | | | | Please provide your Indications for Use below. | | ? | | In Anatomic: Tornier HRS Max humeral assembly can be used for anatomic arthroplasty when used with a humeral head. It can be used as a hemiarthroplasty, if the natural glenoid provides a sufficient bearing surface, or in conjunction with a glenoid, as a total replacement. Tornier HRS Max is indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and with an intact or reconstructable rotator cuff with pain disabled by: • Rheumatoid arthritis • Non-inflammatory degenerative joint disease (i.e., osteoarthritis and avascular necrosis) • Fractures of the humeral head • Post-Traumatic arthritis • Oncology applications including bone loss due to tumor resection • Significant humeral resection where adequate fixation can be achieved In Reverse: Tornier HRS Max is indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and with massive and non-repairable rotator cuff-tear with pain disabled by: • Rheumatoid arthritis • Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis) • Correction of functional deformity • Fractures of the humeral head • Post-Traumatic arthritis • Massive and non-repairable rotator cuff tear • Oncology applications including bone loss due to tumor resection. • Significant humeral resection where adequate fixation can be achieved The reversed tray and polyethylene insert are indicated for use in the conversion from an anatomic to reversed shoulder arthroplasty without the removal of a well-fixed humeral assembly during revision surgery for patients with a functional deltoid muscle. Notice: • All components are single-use. • The coated humeral stem is intended for cemented or cementless use • The all-poly glenoid components are intended for cemented use only. • The glenoid sphere implant is anchored to the bone with screws and is for non-cemented fixation. • Titanium humeral heads are intended for patients with suspected cobalt alloy material sensitivity. The wear properties of titanium and titanium alloys are inferior to that of cobalt alloy. A titanium humeral head is not recommended for patients who lack a suspected material sensitivity to cobalt alloy. Please select the types of uses (select one or both, as applicable). ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | {4} K251686-Page 1 | 510(k) #: | 510(k) Summary | Prepared on: 2025-10-30 | | --- | --- | --- | | Contact Details | | 21 CFR 807.92(a)(1) | | Applicant Name | Tornier, Inc. | | | Applicant Address | 10801 Nesbitt Avenue South Bloomington MN 55441 United States | | | Applicant Contact Telephone | 1-269-800-2754 | | | Applicant Contact | Mrs. Stefanie Tarara | | | Applicant Contact Email | stefanie.tarara@stryker.com | | | Device Name | | 21 CFR 807.92(a)(2) | | Device Trade Name | Tornier Humeral Reconstruction System Max (Tornier HRS Max) | | | Common Name | Shoulder joint metal/polymer non-constrained cemented prosthesis | | | Classification Name | Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented | | | Regulation Number | 888.3650 | | | Product Code(s) | KWT, KWS, HSD, PHX | | | Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) | | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | K223631 | Comprehensive Segmental Revision System (SRS) | KWT | | K241878 | Tornier Humeral Reconstruction System (Tornier HRS); Tornier Perform Humeral System -Stem (Tornier PHS - Stem) | KWS | | Device Description Summary | | 21 CFR 807.92(a)(4) | | The HRS Max System is designed to address massive proximal humeral bone loss and provide for deltoid wrapping. The modular functionality of the HRS Max Shoulder System allows clinicians to create constructs that address unique anatomies which include proximal humeral bone loss, while also allowing for use in anatomic, hemiarthroplasty or reversed configuration. | | | | Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) | | In Anatomic: Tornier HRS Max humeral assembly can be used for anatomic arthroplasty when used with a humeral head. It can be used as a hemiarthroplasty, if the natural glenoid provides a sufficient bearing surface, or in conjunction with a glenoid, as a total replacement. | | | | Tornier HRS Max is indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and with an intact or reconstructable rotator cuff with pain disabled by: • Rheumatoid arthritis • Non-inflammatory degenerative joint disease (i.e., osteoarthritis and avascular necrosis) • Fractures of the humeral head • Post-Traumatic arthritis • Oncology applications including bone loss due to tumor resection • Significant humeral resection where adequate fixation can be achieved | | | {5} K251686-Page 2 In Reverse: Tornier HRS Max is indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and with massive and non-repairable rotator cuff-tear with pain disabled by: - Rheumatoid arthritis - Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis) - Correction of functional deformity - Fractures of the humeral head - Post-Traumatic arthritis - Massive and non-repairable rotator cuff tear - Oncology applications including bone loss due to tumor resection. - Significant humeral resection where adequate fixation can be achieved The reversed tray and polyethylene insert are indicated for use in the conversion from an anatomic to reversed shoulder arthroplasty without the removal of a well-fixed humeral assembly during revision surgery for patients with a functional deltoid muscle. Notice: - All components are single-use. - The coated humeral stem is intended for cemented or cementless use - The all-poly glenoid components are intended for cemented use only. - The glenoid sphere implant is anchored to the bone with screws and is for non-cemented fixation. - Titanium humeral heads are intended for patients with suspected cobalt alloy material sensitivity. The wear properties of titanium and titanium alloys are inferior to that of cobalt alloy. A titanium humeral head is not recommended for patients who lack a suspected material sensitivity to cobalt alloy. Indications for Use Comparison 21 CFR 807.92(a)(5) The Tornier HRS Max System indications for use align with those of the predicate device. Technological Comparison 21 CFR 807.92(a)(6) Tornier HRS Max and the predicate Comprehensive Segmental Revision System (SRS) have the same intended use, similar principle of operation and technological features. The differences for the subject Tornier HRS Max include multiple sized proximal bodies for deltoid wrapping. Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) Non-clinical testing was performed to demonstrate substantial equivalence to the predicate device. - Fatigue testing with corrosion evaluation - Pull out and torque out testing - Range of motion testing - Biocompatibility evaluation - Packaging and shelf-life evaluations - Sterilization evaluation - Endotoxin testing - MRI compatibility evaluation No clinical studies were performed. The Tornier HRS Max does not raise new questions of safety or effectiveness and has been shown to be substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...