← Product Code [KWT](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWT) · K182500

# Catalyst CSR Press-Fit Humeral Components (K182500)

_Catalyst Orthoscience, Inc. · KWT · Jan 11, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWT/K182500

## Device Facts

- **Applicant:** Catalyst Orthoscience, Inc.
- **Product Code:** [KWT](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWT.md)
- **Decision Date:** Jan 11, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3650
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Catalyst CSR Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint where hemi- or total shoulder arthroplasty is determined by the surgeon to be the preferred method of treatment. The Catalyst CSR Shoulder System is intended for use in patients with the following conditions where the humeral head, neck and glenoid vault are of sufficient bone stock and the rotator cuff is intact or reconstructable. - Osteoarthritis - Avascular Necrosis - Rheumatoid Arthritis - Post-traumatic Arthritis - Correction of functional deformity The Catalyst CSR humeral and glenoid implants are intended for cemented use. The Catalyst CSR Press-Fit humeral implants are intended for uncemented or cemented use.

## Device Story

Bone-preserving total shoulder prosthesis; provides alternative to traditional designs requiring extensive bone resection. System includes humeral and glenoid components; new press-fit humeral components added for uncemented or cemented use. Humeral component features non-spherical polished articulating surface and 4 pegs for rotational stability/alignment; back-side geometry matches 4 cut surfaces on humeral head to restore anatomy. Used by orthopedic surgeons in clinical settings for shoulder arthroplasty. Benefits patient by preserving bone stock while addressing degenerative joint conditions.

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including static shear testing for fixation strength and engineering analysis of torque-out. Bacterial endotoxin testing confirmed compliance with limits (<10EU/component or <20EU/device construct).

## Technological Characteristics

Humeral components manufactured from cast Co-Cr-Mo alloy (ASTM F75). Undersurface and proximal alignment pegs feature plasma-sprayed CP Ti coating (ASTM F1580). Design utilizes 4-peg geometry for rotational stability. Compatible with previously cleared 3-peg glenoid components.

## Regulatory Identification

A shoulder joint metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Special Controls

*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-Aluminum 4-Vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-Chromium-Molybdenum Casting Alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-Chromium-Molybdenum Alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal Coatings,”
(vii) F 1378-97 “Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”

## Predicate Devices

- Catalyst OrthoScience CSR Shoulder System ([K152825](/device/K152825.md))
- Catalyst OrthoScience CSR Shoulder System ([K181287](/device/K181287.md))
- Exactech Equinoxe Resurfacing System ([K131298](/device/K131298.md))
- Biomet Copeland Resurfacing Heads ([K003044](/device/K003044.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 11, 2019

Catalyst OrthoScience, Inc. Dale Davison Sr. VP of Manufacturing & Product Development 14710 Tamiami Trail North Naples, Florida 34110

### Re: K182500

Trade/Device Name: Catalyst CSR Press-Fit Humeral Components Regulation Number: 21 CFR 888.3650 Regulation Name: Shoulder Joint Metal/Polymer Non-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: KWT, HSD Dated: December 5, 2018 Received: December 7, 2018

Dear Dale Davison:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Vesa
Digitally signed
by Vesa Vuniqi -
S
Vuniqi -S Date: 2019.01.11
14:05:05 -05'00'

 Mark N. Melkerson For: Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K182500

Device Name

Catalyst OrthoScience CSR Shoulder System

Indications for Use (Describe)

The Catalyst CSR Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint where hemi- or total shoulder arthroplasty is determined by the surgeon to be the preferred method of treatment. The Catalyst CSR Shoulder System is intended for use in patients with the following conditions where the humeral head, neck and glenoid vault are of sufficient bone stock and the rotator cuff is intact or reconstructable.

- Osteoarthritis
- Avascular Necrosis ●
- Rheumatoid Arthritis ●
- Post-traumatic Arthritis
- Correction of functional deformity

The Catalyst CSR humeral and glenoid implants are intended for cemented use.

The Catalyst CSR Press-Fit humeral implants are intended for uncemented or cemented use.

Type of Use (Select one or both, as applicable)

区Prescription Use (Part 21 CFR 801 Subpart D)

□Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED

# FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

| K182500                          | 510(k) Summary                                                                                                                                                                                          |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prepared:                        | December 4, 2018                                                                                                                                                                                        |
| Submitter:                       | Catalyst OrthoScience, Inc.<br>14710 Tamiami Trail North<br>Naples, FL 34110                                                                                                                            |
| Contact:                         | Dale Davison<br>Sr. VP of Manufacturing & Product Development<br>Catalyst OrthoScience, Inc.<br>1-239-325-9976 ext 102<br>ddavison@catalystortho.com                                                    |
| Proprietary Name:                | Catalyst CSR Shoulder System                                                                                                                                                                            |
| Common Name:                     | Shoulder Prosthesis                                                                                                                                                                                     |
| Classification Names:            | 21 CFR 888.3650: Shoulder joint metal/polymer non-constrained<br>cemented prosthesis; Class II<br>21 CFR 888.3690: Shoulder joint humeral (hemi-shoulder) metallic<br>uncemented prosthesis; Class II   |
| Product Codes:                   | KWT, HSD                                                                                                                                                                                                |
| Substantially Equivalent Devices | Catalyst OrthoScience CSR Shoulder System, K152825<br>Catalyst OrthoScience CSR Shoulder System, K181287<br>Exactech Equinoxe Resurfacing System, K131298<br>Biomet Copeland Resurfacing Heads, K003044 |

### Device Description:

The Catalyst CSR Shoulder System is a bone preserving total shoulder prosthesis designed for use in patients where the humeral head, neck and glenoid vault are of sufficient bone stock and there is an intact or reconstructable rotator cuff. The design requires minimal bone resection, thus giving the patient an alternative to other total shoulder designs where more bone is removed.

{4}------------------------------------------------

This submission adds Press-Fit Humeral Components to the CSR Shoulder System. The press-fit humeral components have a non-spherical polished surface for articulation with the glenoid component or the glenoid cavity of the scapula. The humeral components incorporate 4 pegs which assist with alignment and provide rotational stability. The four plane geometry of the back side of the humeral component matches four cut surfaces on the humeral head to recreate the geometry and thickness of the removed bone. The CSR Press-Fit Humeral Components are manufactured from cast Co-Cr-Mo alloy conforming to ASTM F75. The undersurface of the head and the proximal portion of the alignment pegs are coated with a plasma sprayed coating of CP Ti conforming to ASTM F1580.

The CSR humeral components are compatible with previously cleared CSR 3 Peg glenoid components.

# Intended Use / Indications:

The Catalyst CSR Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint where hemi- or total shoulder arthroplasty is determined by the surgeon to be the preferred method of treatment. The Catalyst CSR Shoulder System is intended for use in patients with the following conditions where the humeral head, neck and glenoid vault are of sufficient bone stock and the rotator cuff is intact or reconstructable.

- Osteoarthritis ●
- Avascular Necrosis
- Rheumatoid Arthritis ●
- . Post-traumatic Arthritis
- Correction of functional deformity

The Catalyst CSR humeral and glenoid implants are intended for cemented use.

The Catalyst CSR Press-Fit humeral implants are intended for uncemented or cemented use.

# Summary of Technologies/Substantial Equivalence:

The Catalyst CSR Press-Fit Humeral Components are substantially equivalent to the predicate devices in regards to intended use and indications, materials, design and sizes. The press-fit indication and the addition of a plasma sprayed coating are similar to other predicate devices and do not rase new types of safety and effectiveness questions, nor are there new technological issues.

{5}------------------------------------------------

#### Non-Clinical Testing:

Static shear testing was conducted and demonstrated that the fixation strength of the Catalyst CSR Press-Fit Humeral Components, tested in shear, met the pre-determined acceptance criterion. An engineering analysis of torque-out of the predicate Catalyst CSR Humeral Components was determined to be applicable to the Catalyst CSR Press-Fit Humeral Components as well. Bacterial Endotoxin Testing was performed. The Catalyst CSR Press-Fit Humeral Components met the acceptable endotoxin limit of <10EU/component or <20EU/device construct.

# Clinical Testing:

Clinical testing was not necessary to demonstrate substantial equivalence of the Catalyst CSR Press-Fit Humeral Components to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWT/K182500](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWT/K182500)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
