K974044 · DePuy Orthopaedics, Inc. · KWS · Jan 22, 1998 · Orthopedic
Device Facts
Record ID
K974044
Device Name
GLOBAL TOTAL SHOULDER ECCENTRIC HUMERAL HEAD
Applicant
DePuy Orthopaedics, Inc.
Product Code
KWS · Orthopedic
Decision Date
Jan 22, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3660
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The humeral components of the Global Total Shoulder are intended for cemented or cementless use as a total or hemi-shoulder replacement which is indicated for: 1. A severely painful and/or a severely disabled joint resulting from osteoarthritis, traumatic arthritis, or rheumatoid arthritis; 2. Fracture-dislocations of the proximal humerus where the articular surfaces are severely comminuted, separated from blood supply or where experience shows more conventional methods of treatment are unsatisfactory; 3. Other difficult clinical management problems where arthrodesis or resection are not acceptable (i.e., revision of a failed primary component). The glenoid components of the Global Total Shoulder are intended for cemented use only for the above indications. Hemi-shoulder replacement is also indicated for: 1. Ununited humeral head fractures of long duration; 2. Avascular necrosis of the humeral head.
Device Story
The Global Total Shoulder Eccentric Humeral Head is a prosthetic component for total or hemi-shoulder replacement. It features a male taper shifted 4.0mm off-center, allowing surgeons to 'dial' the head position for improved anatomical alignment on the proximal humerus. It is designed for use with existing DePuy Global Total Shoulder humeral bodies and glenoid components. The device is implanted by orthopedic surgeons in a clinical/surgical setting. By providing an offset option, the device enables better coverage of the cut surface of the proximal humerus, potentially improving joint stability and function for patients suffering from severe arthritis, fractures, or necrosis.
Clinical Evidence
No clinical data provided; substantial equivalence is based on design, material, and geometric similarity to the predicate device.
Technological Characteristics
Material: Co-Cr-Mo alloy. Design: Humeral head prosthesis with 4.0mm eccentric male taper offset. Connectivity: None (mechanical implant). Sterilization: Not specified. Software: None.
Indications for Use
Indicated for patients with severe joint pain/disability due to osteoarthritis, traumatic arthritis, or rheumatoid arthritis; proximal humerus fracture-dislocations with severe comminution or vascular compromise; revision of failed primary components; ununited humeral head fractures; or avascular necrosis of the humeral head.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
DePuy Global™ Total Shoulder (standard) Humeral Head
Submission Summary (Full Text)
{0}------------------------------------------------
CLINKEGLOS
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
EXHIBIT DePuy, Inc. NAME OF FIRM: 700 Orthopaedic Drive JAN 22 1998 Warsaw, Indiana 46581-0988 510(K) CONTACT: Arlene C. Saull, RAC Sr. Submissions Associate DePuy Orthopaedics, Inc. 700 Orthopaedic Drive Warsaw, IN 46581-0988 Global™ Total Shoulder Eccentric Humeral Head TRADE NAME: Total Shoulder Prosthesis, or Hemi-Shoulder Prosthesis COMMON NAME: Class III. when used as a total shoulder per 21 CFR, CLASSIFICATION: Shoulder joint metal/polymer semi-constrained 888.3660: cemented prosthesis. Class II, when used as a hemi-shoulder per 21 CFR, 888.3690. Shoulder joint humeral (hemi-shoulder) metallic uncemented prosthesis. Shoulder, Semi-Constrained KWS: Prosthesis, DEVICE PRODUCT CODE: 87 Metal/Polymer, Cemented (Class III) 87 HSD Prosthesis, Hemi-Shoulder, Humeral, Metallic Uncemented (Class II) SUBSTANTIALLY DePuy Global™ Total Shoulder EQUIVALENT DEVICES: DePuy Global™ Total Shoulder, Humeral Component, for
Cemented or Uncemented Use
#### DEVICE DESCRIPTION AND INTENDED USE:
The DePuy Global Total Shoulder Eccentric Humeral Head will be available in a variety of sizes which are designed with offset male tapers to be used with the currently marketed DePuy Global Total Shoulder Humeral Bodies and Glenoid Components. The male taper on the Global Total Shoulder Eccentric Humaral Head is shifted 4.0mm off the head center. This offset permits the head to be "dialed" to cover the cut surface of the proximal humerus. This allows placement of the head in a more anatornically aligned position.
The humeral components of the Global Total Shoulder are intended for cemented or cementless use as a total or hemi-shoulder replacement which is indicated for:
1 is ge 1 al 2
{1}------------------------------------------------
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
- A severely paintul and/or a severely disabled joint resulting from osteoarthritis, traumatic 】. arthritis, or rhewnatoid arthritis;
- Fracture-dislocations of the proximal humerus where the articular surfaces are severely 2. comminuted, separated from blood supply or where experience shows more conventional methods of treatment are unsatisfactory;
- 3. Other difficult :) incal management problems where arthrodesis or resection are not acceptable (i.e., revision of a failed primary component).
The glenoid components of the Global Total Shoulder are intended for cemented use only for the above indications.
Hemi-shoulder replacement is also indicated for:
- Ununited humeral head fractures of long duration; 1 -
- Avascular necrosis of the humeral head. 2.
## BASIS OF SUBSTANTIAL EQUIVALENCE:
The subject Global Total Shoulder Eccentric Humeral Heads are identical in design to the predicate Global Total Shoulder (:: andard) Humeral Head with the exception that the male taper is offsel 4.0mm.
The subject Global Total Shoulder Eccentric Humeral Heads are designed to be used with the same humeral bodies and glenvid components as the predicate standard humeral heads. The geometry of the articular surface of the subject eccentric humeral head is identical to that of the predicate standard concentric humeral head, thus the contact area will remain the same.
The subject Global Total Shoulder Eccentric Humeral Head is similar in material, sizes, intended use, and design to the predicate Global Total Shoulder (standard) Humeral Heads, with the exception of the eccentricity (offset) of the male taper of the subject device versus the concentricity of the male taper of the predicate device.
The indications for use are the same for both and they replace the same anatomic structures. Also, the Co-Cr-Mo alloy material of the subject eccentric humeral head is the same as that used for the predicate device.
Based on the information provided in this premarket notification DePuy considers the Global Total Shoulder Eccentric Humeral Head to be substantially equivalent to the predicate Global Total Shoulder (standard) Humoral Head currently marketed.
Page 2 of 2
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 2 1998
Arlene C. Saull, RAC Senior Submissions Associate DePuy Inc. P.O. Box 988 700 Orthopaedic Drive Warsaw, Indiana 46581-0988
K974044 Re: Global™ Total Shoulder Eccentric Humeral Head Requlatory Class: III Product Codes: KWS and HSD Dated: October-23, 1997 ---Received: October 24, 1997
Dear Ms. Saull:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). This decision is based on the humeral component being used as a cemented or uncemented humeral hemi-shoulder, or as the cemented or uncemented humeral component of a total shoulder system which has a cemented glenoid component. You may, therefore, market your device subject to the general controls provisions of the Act and the following limitations:
- The glenoid component of the Global Total Shoulder may " 1. not be labeled or promoted for non-cemented use.
- All labeling for this device, including package label 2. and labeling included within the package, must prominently state that the glenoid component is intended for cemented use only.
{3}------------------------------------------------
### Page 2 - Arlene C. Saull, RAC
- Any non-cemented fixation of the glenoid component of 3. this device is considered investigational and may only be investigated as a significant risk device in accordance with the investigational device exemption (IDE) regulation under 21 CFR, Part 812. All users of the glenoid component for non-cemented fixation must receive approval from their respective institutional review boards (IRBs) and the Food and Drug Administration (FDA) to conduct the investigation.
The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Failure to Administration (FDA) will verify such assumptions. comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Requlations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be
{4}------------------------------------------------
Page 3 - Arlene C. Saull, RAC
obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
~Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "EXHIBIT" in a handwritten, block-letter style. The letters are bold and black against a white background. The word appears to be written with a thick marker or pen, giving it a clear and legible appearance.
## INDICATIONS
510(k) Number (if known)
Device Name Global Total Shoulder
The subject device Global Shoulder Eccentric Humeral Head is part of the Global Total Shoulder System.
Indications for Use:
The humeral components of the Global Total Shoulder are intended for cementless use as a total or hemi-shoulder replacement which is indicated for:
- 1. A severely painful und/or a severely disabled joint resulting from osteoarthritis, traumatic arthritis, or rheumatoid arthritis:
- 2. Fracture-dislocations of the proximal humerus where the articular surfaces are severely comminuted, separated from blood supply or where experience shows more conventional methods of treatment are unsatisfactory:
- 3. Other difficult clinical management problems where arthrodesis or resection are not acceptable (i.e., revision of a failed primary component).
The glenoid components of the Global Total Shoulder are intended for cemented use only for the above indications.
Hemi-shoulder replacement is also indicated for:
1. Ununited humeral head fractures of long duration;
2. Avascular necrosis of the humeral head.
Concurrence of CDRH, Office of Device Evaluation:
OR
\$\infty\$
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number. K974044
Prescription Use
Over-The Counter Use __ (Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.