The DVO™ Total and Hemi Shoulder System is intended for use in total or hemi-arthroplasty.
Device Story
Modular total and hemi shoulder prosthesis system; includes humeral stems, humeral heads (standard/eccentric), and glenoids (pegged/keeled). Humeral heads mate to stems via locking taper. Stems available in standard/long lengths, fluted, with fins/suture holes. Used by orthopedic surgeons in clinical/surgical settings for shoulder arthroplasty. Provides mechanical replacement of the glenohumeral joint to restore function and reduce pain in patients with degenerative or traumatic joint disease. Output is a physical implant system; clinical decision-making relies on surgeon assessment of joint pathology and bone quality. Benefits include joint pain relief and improved mobility.
Clinical Evidence
Bench testing only. Evidence includes dimensional analysis and mechanical strength testing (fatigue load-carrying capability) to demonstrate substantial equivalence to predicate devices.
Indicated for patients with severely painful/disabled joints due to osteoarthritis, traumatic arthritis, or rheumatoid arthritis; proximal humerus fracture/dislocations with severe comminution or vascular compromise; or difficult clinical scenarios where arthrodesis/resection are unacceptable (e.g., revision). Hemi-shoulder replacement is also indicated for ununited humeral head fractures, avascular necrosis, and rotator cuff tear arthropathy. Stems/glenoids labeled 'for cemented use only' require bone cement; other stems allow press-fit or cemented use. Single use only.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
DePuy Global shoulder system (Total Shoulder, K905786, K914000)
Advantage Shoulder Humeral Stem and Eccentric Head (K992065)
Submission Summary (Full Text)
{0}------------------------------------------------
K060988 page 143
Image /page/0/Picture/1 description: The image shows the letters "DVO" in white font on a black background. The letters are large and bold, and they are the main focus of the image. The letters are evenly spaced and aligned horizontally. The image is simple and minimalist, with a focus on the letters themselves.
DEC - 6 2006
| Extremity Solutions<br>720 E. Winona Ave., Warsaw, IN 46580 | 510(k) Summary of Safety and Effectiveness |
|-------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| SUMMARY PREPARED: | 28 NOVEMBER 2006 |
| 510(k) SPONSOR/APPLICANT: | DVO TM Extremity Solutions, LLC<br>720 E. Winona Ave., Warsaw IN 46580 |
| 510(k) PREPARER and<br>CONTACT PERSON: | Dina L. Weissman, J.D.<br>P.O. Box 205, Derby CT 06418<br>Tel: (203) 736-8631, Email: dina.weissman@sbcglobal.net |
| TRADE NAME: | Total and Hemi Shoulder System |
| COMMON NAMES: | Shoulder prosthesis, humeral head |
| CLASSIFICATION, CLASS and<br>DEVICE PRODUCT CODE: | 21 CFR 888.3660 Class II, 87 KWS<br>21 CFR 888.3690, Class II, 87 HSD |
| PREDICATE DEVICES: | The DePuy Global shoulder system cleared in:<br>• Total Shoulder, K905786, K914000<br>• Advantage Shoulder Humeral Stem and Eccentric Head, K992065 |
#### DEVICE DESCRIPTION:
This sterile modular total and hemi shoulder system is comprised of a humeral stem, two styles of humeral heads (standard and exentric) and two styles of glenoids (pegged or keeled). The heads mate on the stems through a locking taper. The heads are highly polished and articulate with the glenoids.
The humeral stems are offered in two lengths: standard (110mm with diameters of 6 to 16mm) or long (200 to 220mm in length with diameters of 8 to 14mm). The stems are fluted, except for the 6mm diameter stem. The stems also have fins with suture holes. The head is available in 15 standard sizes (40 to 56mm in heights of 15, 18 and 21 mm) and 10 eccentric sizes (40 to 56mm in heights of 18mm or 21mm).
Materials include titanium humeral stems, cobalt chrome humeral heads and ultrahigh molecular weight polyethylene glenoids.
(Continued on next page)
Page 1 of 2
{1}------------------------------------------------
page 2/3
Image /page/1/Picture/1 description: The image shows the logo for DVO Extremity Solutions. The address 720 E. Winona Ave., Warsaw, IN 46580 is printed below the company name. The logo is in black and white.
# 510(k) Summary of Safety and Effectiveness
### INTENDED USE:
The DVOTM Total and Hemi Shoulder System is intended for use in total or hemi-arthroplasty.
### INDICATIONS FOR USE:
The DVO™ Total and Hemi Shoulder System is indicated for:
- A severely painful and/or disabled joint resulting from osteoarthritis, traumatic arthritis or: 1. rheumatoid arthritis:
- 2. Fracture/dislocations of the proximal humeru:) where the articular surface is severely comminuted. separated from its blood supply or where the surgeon's experience indicates that alternative methods of treatment are unsatisfactory;
- 3. Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g., revision of a failed primary : omponent).
Hemi-shoulder replacement is also indicated for:
- Ununited humeral head fractures; 1.
- 2. Avascular necrosis of the humeral head.
- 3. Rotator cuff tear arthropathy.
Humeral stems and glenoids labeled "for cemented use only" are indicated only for use with bone cement. Humeral stems are also indicated for press-fit uncemented use or for use with bone cement.
This is a single use device.
(Continued on next page)
Page 2 of 3
{2}------------------------------------------------
page 13
Image /page/2/Picture/1 description: The image shows the logo for "DVO Extremity Solutions" along with their address. The logo is in bold, white letters against a black background. The address is "720 E. Winona Ave., Warsaw, IN 46580".
# 510(k) Summary of Safety and Effectiveness
### COMPARISON TO PREDICATES:
The DVO Total and Hemi Shoulder System is similar to the listed predicate devices in intended use for total or hemi shoulder replacement, performance characteristics, materials of construction (titanum alloy, cobalt chrome alloy and UHMWPE), methods of sterilization (gamma irradiation), packaging, manufacturing methods and design. This is evidenced by comparison of technological characteristics, dimensional analysis and nonclinical testing.
Specifically, both the DVO subject devices and the Dr.Puy predicate devices are comprised of a humeral stem and a humeral head mating together with an articulating UHMWPE glenoved. The locking tapers for the head and stem are similar in both the subject and predicate systems,
The stems of both systems are offered in similar lengths (standard and long), with similar diameters. The stems are also similar in being tapered, fluted and having fins with suture holes,
The heads of both systems are of similar size and shape and are offered in standard or eccentric styles. Curvature sizes range from 40mm to 56mm with head heights of 15mm, 18mm and 21 coven
The UHMWPE glenoids of both systems are offer in both pegged and keeled configurations.
The systems also share equal indications for use.
From the standpoint of strength analysis, the subject @rvice is substantially equivalent to the predicate stems because its fatigue load carrying capability exceeds the predicate devices.
Page 3 of 3
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circle around the eagle. The eagle is black, and the text is also black.
DEC - 6 2006
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DVO Extremity Solutions, LLC % Ms. Dina L. Weissman, J.D. P.O. Box 205 Derby. Connecticut 06418
Re: K060988
Trade/Device Name: DVOTM Total and Hemi Shoulder System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained, cemented prosthesis Regulatory Class: Class II Product Code: KWS, HSD Dated: November 29, 2006 Received: December 1, 2006
Dear Ms. Weissman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commercc prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 – Ms. Dina L. Weissman, J.D.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Barbara Buehler
Mark N. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K060988
#### Indications for Use
510(k) Number (if known): K060988
Device Name: DVOTM Total and Hemi Shoulder System
Indications for Use: The DVO™ Total and Hemi Sl»vulder System is indicated for:
- 1. A severely painful and/or disabled joint resulting from osteoarthritis, traumatic arthritis or rheumatoid arthritis:
- Fracture/dislocations of the proximal humerus where the articular surface is severely comminuted, 2. separated from its blood supply or where the surgeon's experience indicates that alternative methods of treatment are unsatisfactory;
- నే. Other difficult clinical problems where shoulder anthrodesis or resection arthroplasty are not acceptable (c.g., revision of a failed primary component).
Hemi-shoulder replacement is also indicated for:
- 1. Ununited humeral head fractures;
- 2. Avascular necrosis of the humeral head.
- 3. Rotator cuff tear arthropathy.
Humeral stems and glenoids labeled "for cemented use only" are indicated only for use with bone cement. Humeral stems are indicated for press-fit uncemented use or for use with bone cement.
This is a single use device.
Prescription Use XXXXXX Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
ーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーーー
Sarbaue Bosens Armson
Division Sign. Off.
Division of Ger Compt 11:10:00 and Neurological Device
Page 1 of 1
510(k) Number Kolo0988
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.