MODIFICATION OF THE ZPLATE-ATL ANTERIOR FIXATION SYSTEM

K990603 · Danek Medical, Inc. · KWQ · Mar 10, 1999 · Orthopedic

Device Facts

Record IDK990603
Device NameMODIFICATION OF THE ZPLATE-ATL ANTERIOR FIXATION SYSTEM
ApplicantDanek Medical, Inc.
Product CodeKWQ · Orthopedic
Decision DateMar 10, 1999
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3060
Device ClassClass 2
AttributesTherapeutic

Intended Use

The ZPLATE-ATL™ Anterior Fixation System is intended for screw/bolt fixation/ attachment to the anterolateral intervertebral bodies from T1 to L5 only. This system is to be used only on one side and placed in such a manner as to be as far away from blood vessels such as the aorta and nerve roots as possible. When properly used, this system will provide temporary stabilization until a solid spinal fusion develops. Specific indications include: - Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). - Pseudoarthrosis. - Spondylolysis. - Spondylolisthesis. - Fracture. - Neoplastic disease. - Unsuccessful previous fusion surgery. - Lordotic deformities of the spine. - Idiopathic thoracolumbar or lumbar scoliosis - Deformity (i.e., scoliosis, lordosis, and/or kyphosis) associated with deficient posterior elements such as that resulting from laminectomy, spina bifida, or myelomenigocele. - Neuromuscular deformity (i.e., scoliosis, lordosis, and / or kyphosis) associated with pelvic obliquity.

Device Story

ZPLATE-ATL™ Anterior Spinal Fixation System provides temporary spinal stabilization to facilitate fusion. System comprises various plates, bolts, screws, and nuts. Surgeon selects and configures components intraoperatively for individual patient anatomy. Device is implanted via anterior/anterolateral approach to T1-L5 vertebral bodies; unilateral placement required to avoid aorta and nerve roots. Provides mechanical support during bone healing. Benefits include stabilization of spinal segments in patients with degenerative, traumatic, or deformity-related conditions.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

System consists of metallic plates, bolts, screws, and nuts. Components are modular, allowing for custom construct configurations. Designed for unilateral anterolateral fixation. No specific materials or ASTM standards cited in the provided text.

Indications for Use

Indicated for patients requiring spinal stabilization from T1 to L5 for degenerative disc disease, pseudoarthrosis, spondylolysis, spondylolisthesis, fractures, neoplastic disease, failed fusion, lordotic deformities, idiopathic scoliosis, deformities due to deficient posterior elements, or neuromuscular deformity. Contraindicated for posterior element (pedicle) attachment in cervical, thoracic, or lumbar spine.

Regulatory Classification

Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Z-PLATE ATL™ Anterior Spinal Fixation System 510(k) Summary February 22, 1999 - Sofamor Danek USA I. Company: 1800 Pvramid Place Memphis. TN 38132 (901) 396-3133 ### Proprietary Trade Name: ZPLATE-ATL™ Anterior Spinal Fixation System II. #### Product Description III. The ZPLATE-ATL™ Anterior Spinal Fixation System consists of a variety of shapes and sizes of plates, bolts, screws and nuts, as well as ancillary products and instrument sets. The components can be locked into a variety of configurations, with each construct tailor-made for the individual case. #### IV. Indications The ZPLATE-ATL™ Anterior Fixation System is intended for screw/bolt fixation/attachment to the anterolateral intervertebral bodies from T1 to L5 only. This system is to be used only on one side and placed in such a manner as to be as far away from blood vessels such as the aorta and nerve roots as possible. When properly used, this system will provide temporary stabilization until a solid spinal fusion develops. Specific indications include: - Degenerative disc disease (as defined 1. by back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). - Pseudoarthrosis. 2. - 3. Spondylolysis. - Spondylolisthesis. 4. - న. Fracture. - Neoplastic disease. 6. - 7. Unsuccessful previous fusion surgery. - Lordotic deformities of the spine. 8. - 9. Idiopathic thoracolumbar or lumbar scoliosis X990603 - 10. Deformity (i.e., scoliosis, lordosis, and/or kyphosis) associated with deficient posterior elements such as that resulting from laminectomy, spina bifida, or myelomenigocele. - 11. Neuromuscular deformity (i.e., scoliosis, lordosis, and / or kyphosis) associated with pelvic obliquity. Warning: This device is not approved for screw attachment to the posterior elements (pedicle) of the cervical, thoracic, or lumbar spine. #### V. Substantial Equivalence Documentation was provided which demonstrated the ZPLATE-ATL™ Anterior Spinal Fixation System to be substantially equivalent to itself. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three abstract human figures or forms, arranged in a vertical stack. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR 1 0 1999 Richard W. Treharne, Ph.D. Vice President, Research and Regulatory Affairs Sofamor Danek 1800 Pyramid Place Memphis, Tennessee 38132 Re: K990603 > Trade Name: ZPLATE-ATL™ Anterior Fixation System Regulatory Class: II Product Code: KWO Dated: February 22, 1999 Received: February 23, 1999 Dear Dr. Treharne: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2}------------------------------------------------ Page 2 - Richard W. Treharne, Ph.D. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: ZPLATE-ATL™ Anterior Fixation System ### Indications for Use: The ZPLATE-ATL™ Anterior Fixation System is intended for screw/bolt fixation/ attachment to the anterolateral intervertebral bodies from T1 to L5 only. This system is to be used only on one side and placed in such a manner as to be as far away from blood vessels such as the aorta and nerve roots as possible. When properly used, this system will provide temporary stabilization until a solid spinal fusion develops. Specific indications include: - Degenerative disc disease (as defined 1. by back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). - 2. Pseudoarthrosis. - 3. Spondylolysis. - 4. Spondylolisthesis. - ડ. Fracture. - 6. Neoplastic disease. - Unsuccessful previous fusion surgery. 7. - Lordotic deformities of the spine. 8. - 9. Idiopathic thoracolumbar or lumbar scoliosis - 10. Deformity (i.e., scoliosis, lordosis, and/or kyphosis) associated with deficient posterior elements such as that resulting from laminectomy, spina bifida, or myelomenigocele. - 11. Neuromuscular deformity (i.e., scoliosis, lordosis, and / or kyphosis) associated with pelvic obliquity. Warning: This device is not approved for screw attachment to the posterior elements (pedicle) of the cervical, thoracic, or lumbar spine. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ## Concurrence of CDRH, Office of Evaluation (ODE) Prescription Use X (Per 21 CFR 801.109) (Optional 1-2-96) OR Over-the-counter Use Asevedo
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...