← Product Code [KWQ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ) · K231839

# CastleLoc-P Anterior Cervical Plate System, AccelFix Lumbar Plate System (K231839)

_L&K BIOMED Co., Ltd. · KWQ · Jul 19, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K231839

## Device Facts

- **Applicant:** L&K BIOMED Co., Ltd.
- **Product Code:** [KWQ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ.md)
- **Decision Date:** Jul 19, 2023
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The CastleLoc-P Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine (C2-C7). The System is indicated for use in the immobilization of the spine as an adjunct to fusions in patients with: - degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), - spondylolisthesis, - trauma (i.e. fractures or dislocations), - tumors, - deformity (defined as kyphosis, lordosis, or scoliosis), - pseudoarthrosis. - failed previous fusion, - spinal stenosis. The AccelFix Lumbar Plate System is indicated for use via the anterolateral surgical approach above the bifurcation of the great vessels or via the anterior surgical approach, below the bifurcation of the great vessels as an adjunct to fusion. This system is indicated in the treatment of lumbosacral (L1-S1) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), pseudoarthrosis, spondylolysis, scoliosis, spondylolisthesis, lordotic deformities of the spinal stenosis, or a failed previous spine surgery.

## Device Story

CastleLoc-P and AccelFix are spinal fixation systems consisting of plates, screws, and lockers. Used by surgeons in clinical settings to immobilize and stabilize the spine as an adjunct to fusion. CastleLoc-P (cervical) attaches to the anterior cervical spine; AccelFix (lumbar) uses anterior, lateral, or anterolateral approaches. Implants provide mechanical stability to the spinal column during the fusion process, potentially reducing pain and correcting deformities. Devices are non-sterile.

## Clinical Evidence

No clinical data. Bench testing only; mechanical performance data from predicate devices was leveraged for the additional components.

## Technological Characteristics

Materials: Ti-6Al-4V ELI titanium alloy (ASTM F136). Cervical system: 1-4 level fusion configurations (13-97mm), 3.5mm/4.0mm self-tapping/self-drilling screws. Lumbar system: plates with spikes for stability, 5.5-6.0mm diameter screws (20-55mm length). Non-sterile.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- CastleLoc-P Anterior Cervical Plate System ([K143271](/device/K143271.md))
- LnK Anterior Cervical Plate System ([K113509](/device/K113509.md), [K143279](/device/K143279.md))
- Paeon Anterior Plate System ([K210035](/device/K210035.md), [K223719](/device/K223719.md))
- AccelFix Lumbar Plate System ([K192481](/device/K192481.md))
- Elatus Lumbar Plate System ([K210035](/device/K210035.md), [K223719](/device/K223719.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

July 19, 2023

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

L & K Biomed Co., Ltd. Katherine Kim RA #101, 201, 202 16-25, Dongbaekjungang-ro 16 beon-gil. Giheung-gu Yongin-si, Gyeonggi-do 17015 South Korea

Re: K231839

Trade/Device Name: CastleLoc-P Anterior Cervical Plate System, AccelFix Lumbar Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWQ Dated: June 20, 2023 Received: June 22, 2023

Dear Katherine Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O'neill -S

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

K231839

Device Name

CastleLoc-P Anterior Cervical Plate System

Indications for Use (Describe)

The CastleLoc-P Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine (C2-C7). The System is indicated for use in the immobilization of the spine as an adjunct to fusions in patients with:

- · degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies),
- · spondylolisthesis,
- · trauma (i.e. fractures or dislocations),
- · tumors,
- · deformity (defined as kyphosis, lordosis, or scoliosis),
- · pseudoarthrosis.
- · failed previous fusion,
- · spinal stenosis.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

|X | Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known)

K231839

Device Name

AccelFix Lumbar Plate System

### Indications for Use (Describe)

The AccelFix Lumbar Plate System is indicated for use via the anterolateral surgical approach above the bifurcation of the great vessels or via the anterior surgical approach, below the bifurcation of the great vessels as an adjunct to fusion. This system is indicated in the treatment of lumbosacral (L1-S1) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), pseudoarthrosis, spondylolysis, scoliosis, spondylolisthesis, lordotic deformities of the spinal stenosis, or a failed previous spine surgery.

| Type of Use (Select one or both, as applicable)                                                                 |                                                                                          |
|-----------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)             </span> | <span>                 ☐ Over-The-Counter Use (21 CFR 801 Subpart C)             </span> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services
   

 Food and Drug Administration
   

 Office of Chief Information Officer
   

 Paperwork Reduction Act (PRA) Staff
   

 PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the text "K231839" and "Page 1 of 3" in the upper right corner. In the lower left corner is the logo for L&K BIOMED. The logo includes the text "LEARNING & KNOWLEDGE BIOMED" above the letters "L&K BIOMED" in a larger font. To the right of the text is a circular graphic made of blue dots.

# 510(k) SUMMARY

The following 510(k) summary is being submitted as required by 21 CFR 807.92(a):

### 1. MANUFACTURER

| Submitter's Name:      | L&K Biomed Co., Ltd.                                                                                      |
|------------------------|-----------------------------------------------------------------------------------------------------------|
| Submitter's Address:   | #101, 201, 202 16-25, Dongbaekjungang-ro 16 beon-gil<br>Giheung-gu, Yongin-si, Gyeonggi-do, 17015, Korea. |
| Submitter's Telephone: | 82-2-6717-1983                                                                                            |
| Contact Person:        | Katherine Kim/82-2-6717-1983<br>khkim@lnkbiomed.com                                                       |
| Prepared Date:         | June 20, 2023                                                                                             |

#### 2. DEVICE IDENTIFICATION

| Device Trade Name        | CastleLoc-P Anterior Cervical Plate System      |
|--------------------------|-------------------------------------------------|
| Common/Usual Name        | Appliance, Fixation, Spinal Intervertebral Body |
| Regulation Class /Number | Class II / 21 CFR 888.3060                      |
| Regulation Name          | Spinal Intervertebral Body Fixation Orthosis    |
| Product Code             | KWQ                                             |
| Classification Panel     | Spinal Devices (DHT6B)                          |

| Device Trade Name        | AccelFix Lumbar Plate System                 |
|--------------------------|----------------------------------------------|
| Common/Usual Name        | Lumbar Spinal Plate                          |
| Regulation Class /Number | Class II/ 21 CFR 888.3060                    |
| Regulation Name          | Spinal Intervertebral Body Fixation Orthosis |
| Product Code             | KWQ                                          |
| Classification Panel     | Spinal Devices (DHT6B)                       |

#### 3. PREDICATE OR LEGALLY MARKETED DEVICES WHICH ARE SUBSTANTIALLY EQUIVALENT.

The subject devices are identical to the predicate devices in all characteristics.

| Subject Device Name                              | 510K NO.             | Trade or Proprietary or Model Name         | Predicate Type              |
|--------------------------------------------------|----------------------|--------------------------------------------|-----------------------------|
| CASTLELOC-P<br>Anterior Cervical<br>Plate System | K 143271             | CastleLoc-P Anterior Cervical Plate System | Primary                     |
|                                                  | K113509<br>K143279   | LnK Anterior Cervical Plate System         | Additional                  |
|                                                  |                      |                                            |                             |
|                                                  |                      | K 210035<br>K 223719                       | Paeon Anterior Plate System |
| AccelFix Lumbar<br>Plate System                  | K 192481             | AccelFix Lumbar Plate System               | Primary                     |
|                                                  | K 210035<br>K 223719 | Elatus Lumbar Plate System                 | Additional                  |

The design feature, indications for use, material and manufacturing process for the subject devices are substantially equivalent to the predicate devices (CastleLoc-P Anterior Cervical Plate System K113509, K143279, K143271, K190425/ AccelFix Lumbar Plate System K192481/200794).

{5}------------------------------------------------

- MATERIALS- And the additional components material is the same material used in the predicate devices. 4.

| CastleLoc-P Anterior Cervical Plate System | Ti-6Al-4V ELI titanium alloy (ASTM F136) |
|--------------------------------------------|------------------------------------------|
| AccelFix Lumbar Plate System               | Ti-6Al-4V ELI titanium alloy (ASTM F136) |

#### ર. DESCRIPTION OF THE DEVICE

The CastleLoc-P Anterior Cervical Plate System is composed of plates, screws and lockers which are made from titanium alloy per ASTM F136. These plates attach to the anterior cervical spine with a minimum of four screws per plate. The plates are offered in one-level, two-level, four-level fusion configurations (13~97mm). The plate screws are 3.5mm and 4.0mm diameter head screws. They are self-tapping and self-drilling threaded. This device can be provided both as non-sterile.

The AccelFix Lumbar Plate System's implants are Lumbar Plate System intended for use as an aid in spinal fixation. They are made from titanium alloy per ASTM F136. The AccelFix Lumbar Plate System consists of a variety of shapes and sizes of plates and screws. The plate has been designed to include spikes for added stability and alignment during screw insertion. The plates feature a curvature for anatomic fit. The diameter of screw is available from 5.5 to 6.0 mm and the length from 20 to 55 mm.

#### 6. INDICATION FOR USE

The CastleLoc-P Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine (C2-C7). The System is indicated for use in the immobilization and stabilization of the spine as an adjunct to fusions in patients with:

- degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies),
- . spondylolisthesis,
- trauma (i.e., fractures or dislocations), "
- . tumors.
- . deformity (defined as kyphosis, lordosis, or scoliosis),
- . pseudoarthrosis,
- . failed previous fusion,
- spinal stenosis.

The AccelFix Lumbar Plate System is indicated for use via the anterior, lateral, or anterolateral surgical approach above the bifurcation of the great vessels or via the anterior surgical approach, below the bifurcation of the great vessels as an adjunct to fusion. This system is indicated in the treatment of lumbar and lumbosacral (L1-S1) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), pseudoarthrosis, spondylolysis, scoliosis, spondylolisthesis, lordotic deformities of the spine, spinal stenosis, or a failed previous spine surgery.

#### 7. PERFORMANCE TESTING

### CastleLoc-P Anterior Cervical Plate System

The additional components to be added through this submission do not require additional mechanical testing. None of the additional components is the worst case of the CastleLoc-P Posterior Anterior Cervical Plate System.

### AccelFix Lumbar Plate System

The additional components to be added through this submission do not require additional mechanical testing. None of the additional components is the worst case of the AccelFix Lumbar Plate System.

Therefore, we substitute mechanical test data of CastleLoc-P Anterior Cervical Plate System (predicate devices K113509, K143279, K190425) and AccelFix Lumbar Plate System (predicate device K 192481) for additional components with the predicate device data.

{6}------------------------------------------------

#### SUMMARY OF TECHNOLOGY CHARACTERISTICS 8.

Subject devices are identical to the predicate devices in all (design feature, indications for use, surgical approach and manufacturing process) characteristics.

|                                                                                        | 510K no.                      | Indication<br>for use | Design  | Surgical<br>approach | Material                     |
|----------------------------------------------------------------------------------------|-------------------------------|-----------------------|---------|----------------------|------------------------------|
| Subject Device                                                                         | -                             | Same                  | Similar | Same                 | Ti-6Al-4V ELI<br>(ASTM F136) |
| CastleLoc-P Anterior Cervical<br>Plate System                                          | Primary<br>K143271            | Same                  | Similar | Same                 | Same                         |
| LnK Anterior Cervical Plate<br>System<br>CastleLoc-P Anterior Cervical<br>Plate System | K113509<br>K143279<br>K190425 | Same                  | Similar | Same                 | Same                         |
| Paeon Anterior Plate System                                                            | K210035<br>K223719            | Same                  | Similar | Same                 | Same                         |

CastleLoc-P CastleLoc-P Anterior Cervical Plate System

### AccelFix Lumbar Plate System

|                              | 510K no.           | Indication<br>for use | Design  | Surgical<br>approach | Material                     |
|------------------------------|--------------------|-----------------------|---------|----------------------|------------------------------|
| Subject Device               | -                  | Same                  | Similar | Same                 | Ti-6Al-4V ELI<br>(ASTM F136) |
| AccelFix Lumbar Plate System | Primary<br>K192481 | Same                  | Similar | Same                 | Same                         |
| Elatus Lumbar Plate System   | K210035<br>K223719 | Same                  | Similar | Same                 | Same                         |

#### 9. SUBSTANTIAL EQUIVALENCE

Subject devices are shown to be substantially equivalent to the predicate devices in indications for use, design, function, and materials used.

### 10. CONCLUSION

The overall technology characteristics lead to the conclusion that the CastleLoc-P Anterior Cervical Plate System, and AccelFix Lumbar Plate System is substantially equivalent to the predicate devices (CastleLoc-P Anterior Cervical Plate System K113509, K143271, K190425/ AccelFix Lumbar Plate System K192481/200794).

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K231839](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K231839)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
