K223753 · Double Medical Technology, Inc. · KWQ · Mar 14, 2023 · Orthopedic
Device Facts
Record ID
K223753
Device Name
Cervical Plate System
Applicant
Double Medical Technology, Inc.
Product Code
KWQ · Orthopedic
Decision Date
Mar 14, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3060
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Cervical Plate System is intended for anterior interbody screw fixation from C2 to T1 and is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusions in patients with: 1) degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies); 2)trauma (including fractures); 3)tumors: 4)deformity(defined as kyphosis, lordosis,or scoliosis); 5)pseudoarthrosis; 6)failed previous fusions
Device Story
Cervical Plate System comprises anterior cervical spine plates and screws used for anterior interbody fixation. Plates are fixed to the anterior cervical vertebrae (C2-T1) using fixed or variable angle screws to maintain surgical segment stability and prevent fusion implant migration. Used by surgeons in clinical settings to provide temporary stabilization during spinal fusion development. Implants are non-sterile, single-use devices.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical bench testing (biocompatibility per ISO 10993-1 and mechanical testing per ASTM F1717).
Technological Characteristics
Materials: Unalloyed titanium (ASTM F67), titanium alloy (ASTM F1472, ASTM F1295) for plates; titanium alloy (ASTM F136) for screws. Mechanical testing per ASTM F1717. Non-sterile, single-use implants; reusable instruments.
Indications for Use
Indicated for patients requiring temporary anterior cervical spine stabilization (C2-T1) during fusion for degenerative disc disease, trauma, tumors, deformity (kyphosis, lordosis, scoliosis), pseudoarthrosis, or failed previous fusions.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
K012184 — BLACKSTONE III ANTERIOR CERVICAL PLATING SYSTEM · Blackstone Medical, Inc. · Jan 30, 2002
K240484 — MSFX Mikron Cervical Anterior Plate System · Mikron Makina Sanayi VE Ticaret Ltd. Sti. · Apr 17, 2024
K022965 — BLACKSTONE CLASSIC ANTERIOR CERVICAL PLATE · Blackstone Medical, Inc. · Mar 28, 2003
K152623 — ATLANTIS Anterior Cervical Plate System · Medtronic Sofamor Danek USA, Inc. · Nov 12, 2015
K193494 — Origin Anterior Cervical Plate System · Orthopaedic & Spine Development · Apr 6, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). The logo features the acronym 'FDA' in bold, capitalized letters, followed by the full name 'U.S. FOOD & DRUG ADMINISTRATION' in a smaller font size. A blue square is positioned to the left of the text, serving as a visual element of the logo.
March 14, 2023
Double Medical Technology Inc. Yan Zuo International RA Supervisor No. 18, Shanbianhong East Road, Haicang District Xiamen. Fujian 361026 China
Re: K223753
Trade/Device Name: Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWQ Dated: December 8, 2022 Received: December 14, 2022
Dear Yan Zuo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin O'neill -S
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K223753
Device Name Cervical Plate System
#### Indications for Use (Describe)
Cervical Plate System is intended for anterior interbody screw fixation from C2 to T1 and is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusions in patients with: 1) degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies);
2)trauma (including fractures);
3)tumors:
4)deformity(defined as kyphosis, lordosis,or scoliosis);
5)pseudoarthrosis;
6)failed previous fusions
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K223753- 510(k) Summary
| 5.1 Submitter | |
|------------------------------------------------------------------|----------------------------------------------------------------------------------|
| Name | Double Medical Technology Inc. |
| Address | No. 18, Shanbianhong East Road, Haicang District, Xiamen,<br>361026, P. R. China |
| Phone | +86 592 6885079 |
| Contact person | Yan Zuo |
| Date prepared | March 10th, 2023 |
| 5.2 Proposed Device | |
| Trade/proprietary name | Cervical Plate System |
| Common or usual name | Cervical Plate System |
| Classification name | Spinal Intervertebral Body Fixation Orthosis |
| Regulation number | 21 CFR 888.3060 |
| Product code | KQW |
| Regulatory class | II |
| Classification panel | Orthopedic |
| 5.3 Predicate Device | |
| Legally marketed<br>device(s) to which<br>equivalence is claimed | Primary predicate device:<br>K141632 ZEVO™ Anterior Cervical Plate System |
| Reason for 510(k)<br>submission | New device(Implant) |
# K223753- 510(k) Summary
## 5.4 Device Description
Cervical Plate System consist of anterior cervical spine plate II, anterior cervical spine plate IV and screws. The anterior cervical spine plate is fixed to the front of the cervical vertebrae by screws to play an auxiliary role in maintaining the stability of the surgical segment and preventing the fusion implant from coming out. Screws are available for fixed angle or variable angle implantation.
Bone plates in Cervical Plate System are made of unalloyed titanium following ASTM F67 and titanium alloy following ASTM F1472 as well as ASTM F1295, and bone screws in Cervical Plate System are made of titanium alloy following ASTM F136.
Cervical Plate System is provided as non-sterile. The implants are intended for single-use only, while the instruments are reusable.
{4}------------------------------------------------
## 5.5 Indication for Use
Cervical Plate System is intended for anterior interbody screw fixation from C2 to T1 and is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusions in patients with:
(1) degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies);
- (2) trauma (including fractures);
- (3) tumors;
- (4) deformity(definedaskyphosis, lordosis,or scoliosis);
- (5) pseudoarthrosis;
- (6) failed previous fusions.
#### 5.6 Comparison of Technological Characteristics with the Predicate Device
The rationale for substantial equivalence is based on consideration of the following characteristics:
Regulatory Classification: Same as the predicate devices
Indications for Use: Substantially equivalent (SE) to the predicate devices
Materials: Substantially equivalent (SE) to the predicate devices
Design Features: Substantially equivalent (SE) to the predicate devices
## 5.7 Non-Clinical Performance Data
#### 5.7.1 Biocompatibility testing
The biocompatibility evaluation for the Cervical Plate System was conducted in accordance with the FDA Guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
#### 5.7.2 Mechanical testing
The following tests were performed (per ASTM F1717 Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model on Anterior Cervical Spine Plate System to demonstrate substantially equivalent of safety and efficacy with the predicate device.
- Static compression bending test
- Static torsion test
- Dynamic compression bending test
## 5.8 Clinical Data
No clinical performance data was provided to demonstrate substantially equivalence.
{5}------------------------------------------------
## 5.9 Conclusion
The information provided within this premarket notification demonstrates that proposed device is determined to be substantially equivalent (SE) to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.