← Product Code [KWQ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ) · K192494

# NexGen Anterior Cervical Plate System (K192494)

_Precision Spine · KWQ · Oct 29, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K192494

## Device Facts

- **Applicant:** Precision Spine
- **Product Code:** [KWQ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ.md)
- **Decision Date:** Oct 29, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The NexGen Anterior Cervical Plate System is intended for anterior interbody screw fixation of the cervical spine during the development of a cervical spinal fusion.

## Device Story

NexGen Anterior Cervical Plate System consists of bone plates and screws (self-drilling or self-tapping) used for temporary anterior interbody screw fixation of the cervical spine (C2-T1). Implants are surgically placed by a surgeon via an anterior approach to provide stabilization during cervical spinal fusion. The system provides mechanical support to the vertebral bodies to facilitate fusion. Benefits include stabilization of the cervical spine in patients with degenerative, traumatic, or structural spinal conditions.

## Clinical Evidence

No clinical data. Bench testing only: static and dynamic compression and torsion testing performed per ASTM F1717.

## Technological Characteristics

Implants constructed from Ti-6Al-4V ELI (ASTM F136). System includes various shapes/sizes of bone plates and screws (self-drilling/self-tapping). Mechanical fixation via anterior approach. No software or energy source.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Zion Anterior Cervical Plating System, Astura Medical ([K160702](/device/K160702.md))
- C-Tek® MaxAn™Anterior Cervical Plate System, Biomet ([K080646](/device/K080646.md))
- CSLP Cervical Plate, Synthes ([K000536](/device/K000536.md))
- UNIPLATE, DePuy ([K042544](/device/K042544.md) / [K082273](/device/K082273.md) / [K100070](/device/K100070.md))

## Reference Devices

- AccuFit™ Lateral Plate System, Precision Spine ([K162211](/device/K162211.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

October 29, 2019

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".

Precision Spine % Mr. J.D. Webb The OrthoMedix Group, Inc. 4313 W. 3800 S. West Haven, Utah 84401

Re: K192494

Trade/Device Name: NexGen Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWQ Dated: September 10, 2019 Received: September 11, 2019

Dear Mr. Webb:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

(Ronald Jean) Vacant Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K192494

Device Name NexGen Anterior Cervical Plate System

#### Indications for Use (Describe)

The NexGen Anterior Cervical Plate System is indicated for use in temporary stabilization of the anterior spine from C2 to T1 during the development of cervical spinal fusions in patients with: degenerative disc disease (DDD) (as defined by neck pain of discogenic origin with degeneration of disc confirmed by patient history and radiographic studies); spondylolisthesis; trauma (including fractures or dislocations); spinal tumors; deformity (defined as kyphosis, or scoliosis); spinal stenosis; pseudoarthrosis; and failed previous fusions.

| Type of Use (Select one or both, as applicable)                                                |                                                                                               |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <div> <span> </span> <span>     Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span> </span> <span>     Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary:

| Date Prepared                                                                  | September 10, 2019                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|--------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted By                                                                   | Precision Spine<br>2050 Executive Drive<br>Pearl, MS 39208                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Primary Contact                                                                | J.D. Webb<br>4313 W. 3800 S<br>West Haven, UT 84401<br>512-590-5810<br>e-mail: jdwebb@orthomedix.net                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Secondary Contact                                                              | Michael C. Dawson<br>Precision Spine<br>2050 Executive Drive<br>Pearl, MS 39208<br>973-455-7150 ext. 128                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Trade Name                                                                     | NexGen Anterior Cervical Plate System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Common Name                                                                    | Anterior Cervical Plate System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Classification Name                                                            | Spinal intervertebral body fixation orthosis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Class                                                                          | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Product Code                                                                   | KWQ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| CFR Section                                                                    | 21 CFR section 888.3060                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Device Panel                                                                   | Orthopedic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Primary Predicate<br>Device                                                    | Zion Anterior Cervical Plating System, Astura Medical (K160702)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Secondary Predicate<br>Devices                                                 | C-Tek® MaxAn™Anterior Cervical Plate System, Biomet (K080646)<br>CSLP Cervical Plate, Synthes (K000536)<br>UNIPLATE, DePuy (K042544 / K082273 / K100070)                                                                                                                                                                                                                                                                                                                                                                                                |
| Reference Device                                                               | AccuFit™ Lateral Plate System, Precision Spine (K162211)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Device Description                                                             | The NexGen Anterior Cervical Plate System components are temporary implants that are<br>intended for anterior interbody screw fixation of the cervical spine during the development<br>of a cervical spinal fusion. The NexGen Anterior Cervical Plate System consists of a variety<br>of shapes and sizes of bone plates, screws (available in self-drilling or self-tapping<br>configurations), and associated instruments. Fixation is provided by bone screws inserted<br>into the vertebral body of the cervical spine using an anterior approach. |
| Materials                                                                      | Implants are made from Ti-6Al-4V ELI conforming to ASTM F136.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Intended Use                                                                   | The NexGen Anterior Cervical Plate System is intended for anterior interbody screw<br>fixation of the cervical spine during the development of a cervical spinal fusion.                                                                                                                                                                                                                                                                                                                                                                                |
| Substantial<br>Equivalence Claimed<br>to Predicate Devices                     | The NexGen Anterior Cervical Plate System is substantially equivalent to the<br>predicate devices in terms of<br>intended use, design, materials used, and<br>mechanical performance.                                                                                                                                                                                                                                                                                                                                                                   |
| Indications for Use                                                            | The NexGen Anterior Cervical Plate System is indicated for use in temporary stabilization<br>of the anterior spine from C2 to T1 during the development of cervical spinal fusions in                                                                                                                                                                                                                                                                                                                                                                   |
|                                                                                | patients with: degenerative disc disease (DDD) (as defined by neck pain of discogenic<br>origin with degeneration of disc confirmed by patient history and radiographic studies);<br>spondylolisthesis; trauma (including fractures or dislocations); spinal tumors; deformity<br>(defined as kyphosis, lordosis, or scoliosis); spinal stenosis; pseudoarthrosis; and failed<br>previous fusions.                                                                                                                                                      |
| Summary of the<br>technological<br>characteristics<br>compared to<br>predicate | When compared to the predicate devices, NexGen Anterior Cervical Plate System has the<br>same intended use and similar technological characteristics, including:<br>• Design<br>• Function/Performance<br>• Principle of Operation<br>• Materials of Construction<br>• Fundamental Scientific Technology                                                                                                                                                                                                                                                |
| Non-clinical Test<br>Summary                                                   | The following analyses were conducted:<br>• Static and dynamic compression and torsion per ASTM F1717<br>The results of these evaluations show that the NexGen Anterior Cervical Plate System is<br>as strong or stronger than the predicate devices.                                                                                                                                                                                                                                                                                                   |
| Clinical Test<br>Summary                                                       | No clinical studies were performed                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Conclusions: Non-<br>clinical and Clinical                                     | Precision Spine considers the NexGen Anterior Cervical Plate System to be equivalent to<br>the predicate devices listed above. This conclusion is based upon the devices' similarities<br>in principles of operation, technology, materials and indications for use                                                                                                                                                                                                                                                                                     |

In accordance with 21 CFR 807.92 of the Federal Code of Regulations

{4}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K192494](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K192494)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
