← Product Code [KWQ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ) · K190016

# Lateral Plate System (K190016)

_Corelink, LLC · KWQ · Jun 12, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K190016

## Device Facts

- **Applicant:** Corelink, LLC
- **Product Code:** [KWQ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ.md)
- **Decision Date:** Jun 12, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The CoreLink Lateral Plate System (LPS) is intended for use as a laterally placed supplemental fixation device via the lateral or anterolateral surgical approach above the bifurcation of the great vessel or via the anterior surgical approach, below the bifurcation of the great vessels. The CoreLink LPS is designed to provide temporary stability until fusion is achieved. It is intended for lateral or anterolateral lumbar (L 1-S1) fixation for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion. Alternatively, the CoreLink LPS may remain attached to CoreLink Lateral lumbar interbody devices after implantation. In this configuration the CoreLink LPS must only be used to treat patients with degenerative disk disease (DDD) at one or two contiquous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels.

## Device Story

Lateral Plate System (LPS) provides temporary spinal stability until fusion. System consists of lumbar plates and screws; features sliding, spring-loaded lock tabs to prevent screw back-out. Plates available in 2-screw or 4-screw configurations; multiple lengths for single-level fixation. Used by surgeons in lateral, anterolateral, or anterior surgical approaches to the lumbar spine. Device acts as supplemental fixation; may be used independently or attached to interbody devices. Provides mechanical support to stabilize spinal segments; aids in clinical decision-making by maintaining alignment during fusion process. Benefits patient by reducing motion at affected spinal levels.

## Clinical Evidence

Bench testing only. Testing included static axial compression bending, static torsion, dynamic axial compression bending fatigue (all per ASTM F1717), and static pull-apart testing. Results demonstrate sufficient strength for intended use.

## Technological Characteristics

Plates: Ti-6Al-4V ELI (ASTM F136). Lock springs: Nitinol (ASTM F2063). Mechanical fixation via screws and spring-loaded locking tabs. Dimensions: multiple plate lengths, two screw diameters. No software or energy source.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Plymouth Plate System ([K120092](/device/K120092.md))
- CAYMAN ([K131533](/device/K131533.md))
- LITe ([K150449](/device/K150449.md))
- Accufit ALP System ([K091044](/device/K091044.md))
- Valiant Anterior Lumbar Plate System ([K082187](/device/K082187.md))
- Halo II Anterior Lumbar Plate System ([K111866](/device/K111866.md))
- Terrace™ Anterior Cervical Plate System ([K163104](/device/K163104.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the letters "FDA" followed by "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font size.

June 12, 2019

CoreLink, LLC % Meredith L. May, MS, RAC Vice President Empirical Consulting LLC 4628 Northpark Drive Colorado Springs, Colorado 80918

Re: K190016

Trade/Device Name: Lateral Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWO Dated: May 7, 2019 Received: May 8, 2019

Dear Ms. May:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for CAPT Raquel Peat, PhD, MPH, USPHS Director OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement on last page.

510(k) Number (if known) K190016 Device Name Lateral Plate System

Indications for Use (Describe)

The CoreLink Lateral Plate System (LPS) is intended for use as a laterally placed supplemental fixation device via the lateral or anterolateral surgical approach above the bifurcation of the great vessel or via the anterior surgical approach, below the bifurcation of the great vessels. The CoreLink LPS is designed to provide temporary stability until fusion is achieved. It is intended for lateral or anterolateral lumbar (L 1-S1) fixation for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.

Alternatively, the CoreLink LPS may remain attached to CoreLink Lateral lumbar interbody devices after implantation. In this configuration the CoreLink LPS must only be used to treat patients with degenerative disk disease (DDD) at one or two contiquous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

☑Prescription Use (Part 21 CFR 801 Subpart D)
 ☐Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including
the time to review instructions, search existing data sources, gather and maintain the data needed and
complete and review the collection of information. Send comments regarding this burden estimate or any
other aspect of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services
   
Food and Drug Administration
   
Office of Chief Information Officer
   
Paperwork Reduction Act (PRA) Staff
   
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information
unless it displays a currently valid OMB number."*

FORM FDA 3881 (7/17)

{3}------------------------------------------------

# 5.510(K) SUMMARY

| Submitter's Name:          | CoreLink                                                                                         |
|----------------------------|--------------------------------------------------------------------------------------------------|
| Submitter's Address:       | 2072 Fenton Logistics Park Blvd.<br>St. Louis, Missouri 63026                                    |
| Submitter's Telephone:     | 888-349-7808                                                                                     |
| Contact Person:            | Meredith Lee May MS, RAC<br>Empirical Consulting<br>719.337.7579<br>Mmay@EmpiricalConsulting.com |
| Date Summary was Prepared: | 8 Jan 19                                                                                         |
| Trade or Proprietary Name: | Lateral Plate System                                                                             |
| Common or Usual Name:      | Appliance, Fixation, Spinal Intervertebral Body                                                  |
| Classification:            | Class II per 21 CFR §888.3060 Spinal Intervertebral Body<br>Fixation Orthosis                    |
| Product Code:              | KWQ                                                                                              |
| Classification Panel:      | Orthopedic Review Panel                                                                          |

## DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:

The CoreLink Lateral Plate System (LPS) is comprised of a lumbar plate and screws. The lumbar plate has a sliding, spring loaded lock tab for each screw position to prevent back-out of the screw. The plate is available in a 2-screw or 4-screw version and in multiple lengths for single level fixation. The screws are available in various lengths and 2 diameters.

## INDICATIONS FOR USE

The CoreLink Lateral Plate System (LPS) is intended for use as a laterally placed supplemental fixation device via the lateral or anterolateral surgical approach above the bifurcation of the great vessel or via the anterior surgical approach, below the bifurcation of the great vessels. The CoreLink LPS is designed to provide temporary stability until fusion is achieved. It is intended for lateral or anterolateral lumbar (L1-S1) fixation for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.

Alternatively, the CoreLink LPS may remain attached to CoreLink Lateral lumbar interbody devices after implantation. In this configuration the CoreLink LPS must only be used to treat patients with degenerative disk disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels.

## TECHNOLOGICAL CHARACTERISTICS

{4}------------------------------------------------

Lateral Plate System plates are made from Ti-6A1-4V ELI in accordance with ASTM F136, and the lock springs are manufactured from Nitinol per ASTM F2063. The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically the following characteristics are identical between the subject and predicates:

- Indications for Use
- . Intended Use
- Implant sizes ●
- Manufacturing ●

| 510k<br>Number | Trade or Proprietary or<br>Model Name      | Manufacturer                 | Predicate<br>Type |
|----------------|--------------------------------------------|------------------------------|-------------------|
| K120092        | Plymouth Plate System                      | Globus                       | Primary           |
| K131533        | CAYMAN                                     | K2M                          | Additional        |
| K150449        | LITe                                       | Stryker                      | Additional        |
| K091044        | Accufit ALP System                         | Spinal USA (Precision Spine) | Additional        |
| K082187        | Valiant Anterior Lumbar<br>Plate System    | Biomet Spine                 | Additional        |
| K111866        | Halo II Anterior Lumbar<br>Plate System    | NuVasive                     | Additional        |
| K163104        | Terrace™ Anterior<br>Cervical Plate System | CoreLink, LLC                | Additional        |

#### Table 5-1 Predicate Devices

## Performance Data

The Lateral Plate System has been tested in the following test modes:

- . Static axial compression bending per ASTM F1717
- . Static torsion per ASTM F1717
- Dynamic axial compression bending fatigue per ASTM F1717
- . Static pull apart testing

The results of this non-clinical testing show that the strength of the Lateral Plate System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.

#### CONCLUSION

The overall technology characteristics and mechanical performance data lead to the conclusion that the Lateral Plate System is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K190016](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K190016)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
