Fuji Cervical Plate System

K183084 · Altus Partners, LLC · KWQ · May 29, 2019 · Orthopedic

Device Facts

Record IDK183084
Device NameFuji Cervical Plate System
ApplicantAltus Partners, LLC
Product CodeKWQ · Orthopedic
Decision DateMay 29, 2019
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3060
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Fuji Cervical Plate System is intended for anterior interbody fixation of the cervical spine. The system is indicated for use in temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies); trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudosarthrosis and/or failed previous fusions. The Fuji Cervical Plate System can be implanted in the sub-axial cervical spine from the C3 through C7 levels.

Device Story

Fuji Cervical Plate System; anterior interbody fixation device for cervical spine. Consists of titanium alloy plates and bone screws (self-tapping/self-drilling). Plates feature graft windows for intraoperative visualization and bone graft fixation. Implanted by surgeons in clinical settings to provide temporary stabilization during fusion development. Device extension adds 4.5mm screw diameter to existing system. No software or AI components.

Clinical Evidence

Bench testing only; engineering analysis provided to demonstrate performance equivalence.

Technological Characteristics

Titanium alloy (ASTM F136). System includes cervical plates (various lengths for 1-4 levels) and bone screws (fixed/variable, self-tapping/self-drilling). Features longitudinal graft windows. No software or electronic components.

Indications for Use

Indicated for patients requiring anterior cervical spine stabilization during fusion. Applicable for degenerative disc disease, trauma (fractures), tumors, deformity (kyphosis, lordosis, scoliosis), pseudosarthrosis, and failed previous fusions. Implantable in sub-axial cervical spine (C3-C7).

Regulatory Classification

Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" on the bottom. May 29, 2019 Altus Partners, LLC Mr. Mark Melton Regulatory Affairs & Quality Assurance 1340 Enterprise Drive West Chester, Pennsylvania 19380 Re: K183084 Trade/Device Name: Fuji Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: October 10, 2018 Received: November 6, 2018 Dear Mr. Melton: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for CAPT Raquel Peat, PhD, MPH, USPHS Director OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) #### K183084 Device Name Fuji Cervical Plate System #### Indications for Use (Describe) The Fuji Cervical Plate System is intended for anterior interbody fixation of the cervical spine. The system is indicated for use in temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies); trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudosarthrosis and/or failed previous fusions. The Fuji Cervical Plate System can be implanted in the sub-axial cervical spine from the C3 through C7 levels. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary | SUBMITTER: | Altus Partners, LLC<br>1340 Enterprise Drive<br>West Chester, PA 19380<br>Phone: 610-355-4156<br>Fax: 610-300-3049 | |-------------------------------|--------------------------------------------------------------------------------------------------------------------| | CONTACT PERSON: | Mark Melton<br>RA&QA<br>mmelton@altus-spine.com | | DATE PREPARED: | May 29, 2019 | | COMMON NAME: | Cervical Plate System | | PROPRIETARY NAME: | Fuji Cervical Plate System | | PRIMARY PREDICATE DEVICES: | Altus Spine Cervical Plate System (K163061) | | ADDITIONAL PREDICATE DEVICES: | N/A | | ADDITIONAL PREDICATE DEVICES: | N/A | |-------------------------------|-------------------------------------------| | CLASSIFICATION NAME: | 21 CFR §888.3060 | | PRODUCT CODES: | KWQ | | DEVICE CLASS: | Class II | | MATERIAL: | Titanium Alloy that conforms to ASTM F136 | ## DEVICE DESCRIPTION: The purpose of this submission is to add the addition of a 4.5mm screw offering for the current FDA cleared Fuji Cervical plate System. It is intended for anterior fixation of the cervical spine. The fixation construct consists of a cervical plate that is attached to the vertebral body of the cervical spine with both self-tapping and self-drilling bone screws. The plates are available in a variety of lengths addressing multiple levels of fixation (one to four). The Fuji Cervical Plate System includes graft windows on the longitudinal center line for intraoperative visualization and for screw fixation of bone graft. Fixed or variable bone screws are available in two diameters and a variety of lengths, with the options of either self-tapping or self-drilling threads. ## INDICATIONS FOR USE: The Fuji Cervical Plate System is intended for anterior interbody fixation of the cervical spine. The system is indicated for use in temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies); trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudosarthrosis and/or failed previous fusions. The Fuji Cervical Plate System can be implanted in the sub-axial cervical spine from the C3 through C7 levels. {4}------------------------------------------------ ### SUMMARY OF TECHNOLOGICAL CHARACTERISTICS: The technological characteristics of the Fuji Cervical Plate System is equivalent to the predicate device, except for the addition of the larger diameter screw. ## SUMMARY OF NON-CLINAL TESTS SUBMITTED: Engineering analysis was presented to demonstrate that the Fuji Cervical Plate System line extension does not present a new worst case in performance and is substantially equivalent to the predicates. ### SUBSTANTIAL EQUIVALENCE CONCLUSION: The Fuji Cervical Plate System is substantially equivalent to the predicate devices in terms of intended use, material used, and performance. The device has a similar design, dimensions and instrumentation to the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...