← Product Code [KWQ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ) · K180022

# Athena Cervical Plate System (K180022)

_Medyssey USA, Inc. · KWQ · Jul 20, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K180022

## Device Facts

- **Applicant:** Medyssey USA, Inc.
- **Product Code:** [KWQ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ.md)
- **Decision Date:** Jul 20, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Athena Cervical Plate System is intended for anterior interbody screw fixation of the cervical spine. The system is indicated for use in the temporary stabilization of the anterior spine during the development of a solid spinal fusion in patients with degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudoarthrosis, and/or failed previous fusions. The Athena Cervical Plate System can be implanted in the sub-axial cervical spine from C3 through C7 levels.

## Device Story

Athena Cervical Plate System provides mechanical support for anterior cervical spine stabilization during fusion. System comprises various plate sizes and screws; available in fixed and variable configurations for different implantation angles. Used by surgeons in clinical settings to stabilize spine segments C3-C7. Device acts as internal fixation hardware; supports bone healing by maintaining alignment until solid fusion occurs. Benefits patient by providing structural stability, reducing motion at fusion site, and facilitating recovery from degenerative conditions, trauma, or deformity.

## Clinical Evidence

Bench testing only. Mechanical performance evaluated per ASTM F1717-13, including static torsion, static compression, and dynamic compression bending (to 5 million cycles). Results demonstrate performance equivalent to predicate devices.

## Technological Characteristics

Spinal intervertebral body fixation orthosis. Components include plates and screws in fixed and variable configurations. Materials and design features are substantially equivalent to predicate systems. Testing conducted per ASTM F1717-13 standard.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Vertebron Cervical Plate ([K081567](/device/K081567.md))
- Vectra Cervical Plate System ([K050451](/device/K050451.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Medyssey USA, Inc. % Rich Jansen Consultant Silver Pine Consulting, LLC. 3851 Mossv Oak Drive Ft. Myers, Florida 33905

Re: K180022

Trade/Device Name: Athena Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWQ Dated: June 26, 2018 Received: June 27, 2018

Dear Mr. Jansen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov

July 20, 2018

{1}------------------------------------------------

## Page 2 - Mr. Rich Jansen

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Melissa Hall-S

For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES                                                                                                                                                                                                                                                                                                                     | Form Approved: OMB No. 0910-0120                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Food and Drug Administration                                                                                                                                                                                                                                                                                                                                | Expiration Date: January 31, 2017                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                |
| Indications for Use                                                                                                                                                                                                                                                                                                                                         | See PRA Statement below.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                |
| 510(k) Number (if known)                                                                                                                                                                                                                                                                                                                                    | K180022                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                |
| Device Name                                                                                                                                                                                                                                                                                                                                                 | Athena Cervical Plate System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                |
| Indications for Use (Describe)                                                                                                                                                                                                                                                                                                                              | The Athena Cervical Plate System is intended for anterior interbody screw fixation of the cervical spine. The system is indicated for use in the temporary stabilization of the anterior spine during the development of a solid spinal fusion in patients with degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudoarthrosis, and/or failed previous fusions. The Athena Cervical Plate System can be implanted in the sub-axial cervical spine from C3 through C7 levels. |                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                |
| Type of Use (Select one or both, as applicable)                                                                                                                                                                                                                                                                                                             | <table><tr><td><div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16"></rect> <path d="M2,2 L14,14 M2,14 L14,2" stroke="black" stroke-width="1"></path> </svg> </span>                             Prescription Use (Part 21 CFR 801 Subpart D)                             C)                         </div></td><td><div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16"></rect> </svg> </span>                             Over-The-Counter Use (21 CFR 801 Subpart                         </div></td></tr></table>                                        | <div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16"></rect> <path d="M2,2 L14,14 M2,14 L14,2" stroke="black" stroke-width="1"></path> </svg> </span>                             Prescription Use (Part 21 CFR 801 Subpart D)                             C)                         </div> | <div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16"></rect> </svg> </span>                             Over-The-Counter Use (21 CFR 801 Subpart                         </div> |
| <div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16"></rect> <path d="M2,2 L14,14 M2,14 L14,2" stroke="black" stroke-width="1"></path> </svg> </span>                             Prescription Use (Part 21 CFR 801 Subpart D)                             C)                         </div> | <div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16"></rect> </svg> </span>                             Over-The-Counter Use (21 CFR 801 Subpart                         </div>                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response,
including the time to review instructions, search existing data sources, gather and maintain the
data needed and complete and review the collection of information. Send comments regarding
this burden estimate or any other aspect of this information collection, including suggestions for
reducing this burden, to:

 Department of Health and
 Human Services Food and Drug
 Administration
 Office of Chief Information
 Officer Paperwork Reduction
 Act (PRA) Staff
 PRAStaff@fda.hhs.gov

*"An agency may not conduct or sponsor, and a person is not required to respond to, a
collection of information unless it displays a currently valid OMB number."*

FORM FDA 3881 (8/14) (301) 443-6740

Page 1 of 1

36& 3XEOLVKLQJ 6HUYLFHV

{3}------------------------------------------------

# 510(k) Summary

| Date Prepared:      | June 26, 2018                                                 |
|---------------------|---------------------------------------------------------------|
| Submitter:          | Shawn Kim                                                     |
|                     | Medyssey USA Inc.                                             |
|                     | 1550 E. Higgins Road                                          |
|                     | Suite 123                                                     |
|                     | Skokie, IL 60077                                              |
|                     | 847-427-0200                                                  |
|                     | 888-518-9070 (fax)                                            |
| Regulatory Contact: | Rich Jansen, Pharm. D.<br>Silver Pine Consulting              |
| Product             |                                                               |
| Trade Names:        | Athena Cervical Plate System                                  |
| Product Class:      | Class II                                                      |
| Classification:     | 21 CFR §888.3060 Spinal Intervertebral Body Fixation Orthosis |
| Name:               | Cervical Plate System                                         |
| Product Codes:      | KWQ                                                           |
| Panel Code:         | 87                                                            |

#### Device Descriptions:

The Medyssey Athena Cervical Plate System is designed to provide mechanical support while biologic fusion takes place. The system consists of multiple plate sizes and screws to accommodate various patients' anatomy. The plates are available in fixed and variable forms to allow for fixed attachment or implantation at variable angles.

#### Indications for Use:

The Athena Cervical Plate System is interbody screw fixation of the cervical spine. The system is indicated for use in the temporary stabilization of the anterior spine during the development of a solid spinal fusion in patients with degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, or scoliosis), pseudoarthrosis, and/or failed previous fusions. The Athena Cervical Plate System can be implanted in the sub-axial cervical spine from C3 through C7 levels.

#### Predicate Device(s):

The primary predicate device is the Vertebron Cervical Plate (K081567). An additional predicate device is the Vectra Cervical Plate System (K050451).

The Athena Cervical Plate was compared to the predicated devices in regards to:

{4}------------------------------------------------

- Indications for Use ●
- Materials
- Number of levels covered
- Plate Dimensions
- Screw Dimensions
- Design features and technological characteristics ●

The Athena Plate is considered to be substantially equivalent to one or both of the predicate devices in all elements of comparison

#### Performance Standards:

The pre-clinical testing was performed by an independent laboratory. Testing was conducted per ASTM F1717-13.

Testing included:

| Test                                      | Result                                 |
|-------------------------------------------|----------------------------------------|
| Static Torsion Bending                    | Substantially equivalent to predicates |
| Static Compression Bending                | Substantially equivalent to predicates |
| Dynamic Compression Bending to 5MM cycles | Substantially equivalent to predicates |

### Conclusion:

Medyssey concludes that the Athena Cervical Plate system is as safe, as effective, and performs as well as or better than the legally marketed predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K180022](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K180022)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
